MRI artifact masquerading as orbital disease

被引:6
作者
Brown B.A. [1 ,2 ,5 ]
Swallow C.E. [3 ,4 ]
Eiseman A.S. [1 ,2 ]
机构
[1] Ophthalmology Service, Walter Reed Army Medical Center, Washington, DC
[2] Department of Surgery, Uniformed Services, University of Health Sciences, Bethesda, MD
[3] Radiology Service, Walter Reed Army Medical Center, Washington, DC
[4] Dept. of Radiol. and Nucl. Medicine, Uniformed Services, University of Health Sciences, Bethesda, MD
[5] Ophthalmology Service, Moncrief Army Community Hospital, Fort Jackson, SC
关键词
Artifact; Magnetic resonance imaging; Orbit;
D O I
10.1023/B:INTE.0000006832.22117.f6
中图分类号
学科分类号
摘要
Objectives: To describe the diagnostic challenge associated with magnetic resonance imaging (MRI) artifact and to describe methods of minimizing this artifact. Methods: Gadolinium enhancement was demonstrated to be artifact by using an inversion recovery pulse sequence technique. Results: A patient with complaints of painless loss of vision also had radiologic abnormalities that led to subsequent referral to a tertiary care facility. Increased signals on the fat-suppressed contrast enhanced MRI was demonstrated to be artifact by using an inversion recovery technique which produced a normal MRI of the orbit. Conclusions: Clinical confusion leading to unnecessary and costly evaluation may be avoided if one considers the possibility of artifact when reviewing magnetic resonance studies. When magnetic susceptibility is suspected, the use of an inversion recovery sequence can better delineate the true nature of the abnormality.
引用
收藏
页码:343 / 347
页数:4
相关论文
共 9 条
  • [1] Delfort E.M., Beltran J., Johnson G., Rousseau J., Merchandise X., Cotton A., Fat suppression in MR imaging: Techniques and pitfalls, Radiographics, 19, 2, pp. 373-382, (1999)
  • [2] Joseph P.M., Atlas S.W., Artifacts, Magnetic Resonance Imaging of the Brain and Spine, 2nd edn., (1996)
  • [3] Borges A.R., Lufkin R.B., Huang A.Y., Farchani K., Arnold A.C., Frequency selective fat suppression MR imaging. Local asymmetric failure of fat mimicking orbital disease, J Neuroophthalmol, 17, pp. 12-17, (1997)
  • [4] Lissac M., Metrop D., Brugirard J., Et al., Dental materials in magnetic resonance imaging, Invest Radiol, 26, pp. 40-45, (1991)
  • [5] Smith F.W., Crosher G.A., Mascara - An unsuspected cause of magnetic imaging artifact, Magnet Reson Imag, 3, pp. 287-289, (1985)
  • [6] Pusey E., Lufkin R.B., Brown R.L.J., Et al., Magnetic resonance imaging artifacts: Mechanism and clinical significance, Radiographics, 6, pp. 891-911, (1986)
  • [7] Young I.R., Bydder G.M., Hajnal J.V., Contrast properties in the inversion recovery sequence, Advanced MR Imaging Techniques, (1997)
  • [8] Wehrli F.W., MacFall J.R., Shutts D., Breger R., Herfkens R.J., Mechanisms of contrast in NMR imaging, J Comput Assist Tomogr, 8, pp. 369-380, (1984)
  • [9] Anzai Y., Lufkin R.B., Et al., Fat suppression failure artifacts simulating pathology on frequency selective fat suppression MR images of the head and neck, AJNR, 13, pp. 879-884, (1992)