Application of microcolumnion chromatography using octadecylsilica immobilized with bovine serum albumin as stationary phase for the determination of inorganic anions

被引:0
|
作者
Rahmiana Zein
Edison Munaf
Toyohide Takeuchi
Tomoo Miwa
机构
[1] Gifu University,Faculty of Engineering
[2] Andalas University,Laboratory of Analytical Environmental Chemistry, Faculty of Mathematics and Natural Sciences
关键词
Thiocyanate; Flow Cell; Tartaric Acid; Anal Chim; Inorganic Anion;
D O I
暂无
中图分类号
学科分类号
摘要
Microcolumn ion chromatography of inorganic anions has been studied using octadecylsilica immobilized with bovine serum albumin as a stationary phase. The analytes were monitored with a conductivity detector. The detection limits for chloride, nitrate and thiocyanate at S/N = 3 were 0.8, 1.0 and 3.0 μmol L−1, respectively. The present system was applied to the determination of inorganic anions in tap water and serum samples.