Use of electronic portal imaging devices for electron treatment verification

被引:0
作者
T. Kairn
T. Aland
S. B. Crowe
J. V. Trapp
机构
[1] Genesis Cancer Care Queensland,Science and Engineering Faculty
[2] Queensland University of Technology,Cancer Care Services
[3] Royal Brisbane and Women’s Hospital,undefined
来源
Australasian Physical & Engineering Sciences in Medicine | 2016年 / 39卷
关键词
Electrons; Portal imaging; Radiation therapy;
D O I
暂无
中图分类号
学科分类号
摘要
This study aims to help broaden the use of electronic portal imaging devices (EPIDs) for pre-treatment patient positioning verification, from photon-beam radiotherapy to photon- and electron-beam radiotherapy, by proposing and testing a method for acquiring clinically-useful EPID images of patient anatomy using electron beams, with a view to enabling and encouraging further research in this area. EPID images used in this study were acquired using all available beams from a linac configured to deliver electron beams with nominal energies of 6, 9, 12, 16 and 20 MeV, as well as photon beams with nominal energies of 6 and 10 MV. A widely-available heterogeneous, approximately-humanoid, thorax phantom was used, to provide an indication of the contrast and noise produced when imaging different types of tissue with comparatively realistic thicknesses. The acquired images were automatically calibrated, corrected for the effects of variations in the sensitivity of individual photodiodes, using a flood field image. For electron beam imaging, flood field EPID calibration images were acquired with and without the placement of blocks of water-equivalent plastic (with thicknesses approximately equal to the practical range of electrons in the plastic) placed upstream of the EPID, to filter out the primary electron beam, leaving only the bremsstrahlung photon signal. While the electron beam images acquired using a standard (unfiltered) flood field calibration were observed to be noisy and difficult to interpret, the electron beam images acquired using the filtered flood field calibration showed tissues and bony anatomy with levels of contrast and noise that were similar to the contrast and noise levels seen in the clinically acceptable photon beam EPID images. The best electron beam imaging results (highest contrast, signal-to-noise and contrast-to-noise ratios) were achieved when the images were acquired using the higher energy electron beams (16 and 20 MeV) when the EPID was calibrated using an intermediate (12 MeV) electron beam energy. These results demonstrate the feasibility of acquiring clinically-useful EPID images of patient anatomy using electron beams and suggest important avenues for future investigation, thus enabling and encouraging further research in this area. There is manifest potential for the EPID imaging method proposed in this work to lead to the clinical use of electron beam imaging for geometric verification of electron treatments in the future.
引用
收藏
页码:199 / 209
页数:10
相关论文
共 212 条
  • [11] Antonuk L(2015)Use of a megavoltage electronic portal imaging device to identify prosthetic materials Australas Phys Eng Sci Med 38 93-100
  • [12] Fenster A(2002)Assessment of flatness and symmetry of megavoltage X-ray beam with an electronic portal imaging device (EPID) Australas Phys Eng Sci Med 25 58-66
  • [13] Van Herk M(2007)Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID Phys Med Biol 52 1721-1733
  • [14] Meertens H(2015)Evaluation of the machine performance check application for TrueBeam Linac Radiat Oncol 10 97-273
  • [15] Munro P(2006)Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method Med Phys 33 259-3155
  • [16] Reinstein LE(2003)Dosimetric IMRT verification with a flat-panel EPID Med Phys 30 3143-234
  • [17] Wong J(2004)The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields Radiother Oncol 71 223-2851
  • [18] McDermott LN(2006)GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications Med Phys 33 2839-992
  • [19] Nijsten SMJJG(2008)The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy Radiat Oncol 3 24-2644
  • [20] Sonke JJ(2009)A simple approach to using an amorphous silicon EPID to verify IMRT planar dose maps Med Phys 36 984-309