Use of electronic portal imaging devices for electron treatment verification

被引:0
作者
T. Kairn
T. Aland
S. B. Crowe
J. V. Trapp
机构
[1] Genesis Cancer Care Queensland,Science and Engineering Faculty
[2] Queensland University of Technology,Cancer Care Services
[3] Royal Brisbane and Women’s Hospital,undefined
来源
Australasian Physical & Engineering Sciences in Medicine | 2016年 / 39卷
关键词
Electrons; Portal imaging; Radiation therapy;
D O I
暂无
中图分类号
学科分类号
摘要
This study aims to help broaden the use of electronic portal imaging devices (EPIDs) for pre-treatment patient positioning verification, from photon-beam radiotherapy to photon- and electron-beam radiotherapy, by proposing and testing a method for acquiring clinically-useful EPID images of patient anatomy using electron beams, with a view to enabling and encouraging further research in this area. EPID images used in this study were acquired using all available beams from a linac configured to deliver electron beams with nominal energies of 6, 9, 12, 16 and 20 MeV, as well as photon beams with nominal energies of 6 and 10 MV. A widely-available heterogeneous, approximately-humanoid, thorax phantom was used, to provide an indication of the contrast and noise produced when imaging different types of tissue with comparatively realistic thicknesses. The acquired images were automatically calibrated, corrected for the effects of variations in the sensitivity of individual photodiodes, using a flood field image. For electron beam imaging, flood field EPID calibration images were acquired with and without the placement of blocks of water-equivalent plastic (with thicknesses approximately equal to the practical range of electrons in the plastic) placed upstream of the EPID, to filter out the primary electron beam, leaving only the bremsstrahlung photon signal. While the electron beam images acquired using a standard (unfiltered) flood field calibration were observed to be noisy and difficult to interpret, the electron beam images acquired using the filtered flood field calibration showed tissues and bony anatomy with levels of contrast and noise that were similar to the contrast and noise levels seen in the clinically acceptable photon beam EPID images. The best electron beam imaging results (highest contrast, signal-to-noise and contrast-to-noise ratios) were achieved when the images were acquired using the higher energy electron beams (16 and 20 MeV) when the EPID was calibrated using an intermediate (12 MeV) electron beam energy. These results demonstrate the feasibility of acquiring clinically-useful EPID images of patient anatomy using electron beams and suggest important avenues for future investigation, thus enabling and encouraging further research in this area. There is manifest potential for the EPID imaging method proposed in this work to lead to the clinical use of electron beam imaging for geometric verification of electron treatments in the future.
引用
收藏
页码:199 / 209
页数:10
相关论文
共 212 条
  • [1] El-Gayed AA(1993)Time trend of patient setup deviations during pelvic irradiation using electronic portal imaging Radiother Oncol 26 162-171
  • [2] Bel A(2013)A snapshot of radiation therapy techniques and technology in Queensland: an aid to mapping undergraduate curriculum J Med Radiat Sci 60 25-34
  • [3] Vijlbrief R(1992)A review of electronic portal imaging devices (EPIDs) Med Phys 19 1-16
  • [4] Bartelink H(2006)Comparison of ghosting effects for three commercial a-Si EPIDs Med Phys 33 2448-2451
  • [5] Lebesque JV(2003)Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy Med Phys 30 1618-1627
  • [6] Bridge P(2005)Initial patient imaging with an optimised radiotherapy beam for portal imaging Radiother Oncol 76 63-71
  • [7] Carmichael MA(2010)Contrast enhancement of EPID images via difference imaging: a feasibility study Phys Med Biol 55 N533-N544
  • [8] Brady C(2005)Low-dose megavoltage cone-beam CT for radiation therapy Int J Radiat Oncol Biol Phys 61 552-560
  • [9] Dry A(2006)Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of paraspinous tumors in the presence of orthopedic hardware Int J Radiat Oncol Biol Phys 66 323-326
  • [10] Boyer AL(2014)Evaluation of MegaVoltage Cone Beam CT image quality with an unmodified Elekta Precise Linac and EPID: a feasibility study Australas Phys Eng Sci Med 37 291-302