Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality

被引:0
作者
J. Geleijns
M. Salvadó Artells
W. J. H. Veldkamp
M. López Tortosa
A. Calzado Cantera
机构
[1] Leiden University Medical Center,Radiology Department
[2] Universitat Rovira i Virgili,Departament de Ciències Mèdiques Bàsiques
[3] Facultat de Medicina i Ciències de la Salut,Departamento de Radiología
[4] Universidad Complutense de Madrid,undefined
来源
European Radiology | 2006年 / 16卷
关键词
X-ray computed tomography; Radiation protection; Radiation effects;
D O I
暂无
中图分类号
学科分类号
摘要
This study aimed at assessment of efficacy of selective in-plane shielding in adults by quantitative evaluation of the achieved dose reduction and image quality. Commercially available accessories for in-plane shielding of the eye lens, thyroid and breast, and an anthropomorphic phantom were used for the evaluation of absorbed dose and image quality. Organ dose and total energy imparted were assessed by means of a Monte Carlo technique taking into account tube voltage, tube current, and scanner type. Image quality was quantified as noise in soft tissue. Application of the lens shield reduced dose to the lens by 27% and to the brain by 1%. The thyroid shield reduced thyroid dose by 26%; the breast shield reduced dose to the breasts by 30% and to the lungs by 15%. Total energy imparted (unshielded/shielded) was 88/86 mJ for computed tomography (CT) brain, 64/60 mJ for CT cervical spine, and 289/260 mJ for CT chest scanning. An increase in image noise could be observed in the ranges were bismuth shielding was applied. The observed reduction of organ dose and total energy imparted could be achieved more efficiently by a reduction of tube current. The application of in-plane selective shielding is therefore discouraged.
引用
收藏
页码:2334 / 2340
页数:6
相关论文
共 68 条
[11]  
Frush DP(2005)Monte Carlo calculation of radiation dose in CT examinations using phantom and patient tomographic models Radiat Prot Dosimetry 114 364-368
[12]  
Yoshizumi T(2004)Reproducibility of multi-slice spiral computed tomography scans: an experimental study Med Phys 31 2785-2786
[13]  
Varchena V(2005)The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations Med Phys 32 1621-1629
[14]  
Poe SA(2000)Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions Eur Radiol 10 391-394
[15]  
Lucaya J(2005)Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation Radiology 237 213-223
[16]  
Hein E(undefined)undefined undefined undefined undefined-undefined
[17]  
Rogalla P(undefined)undefined undefined undefined undefined-undefined
[18]  
Klingebiel R(undefined)undefined undefined undefined undefined-undefined
[19]  
Hamm B(undefined)undefined undefined undefined undefined-undefined
[20]  
Hopper KD(undefined)undefined undefined undefined undefined-undefined