Concept of minimal heart rate for each pitch value to avoid interpolation artifact when using dual-source CT: a phantom study

被引:1
作者
Joon-Won Kang
Kyung-Hyun Do
Jae-Yeon Chung
Hyun Jong Cho
Joon Beom Seo
Tae-Hwan Lim
机构
[1] University of Ulsan College of Medicine,Department of Radiology and Research Institute of Radiology
[2] Asan Medical Center,Radiology team
[3] Asan Medical Center,undefined
来源
The International Journal of Cardiovascular Imaging | 2010年 / 26卷
关键词
Artifacts; Computed tomography; Multidetector computed tomography; Dual-source computed tomography; Image quality;
D O I
暂无
中图分类号
学科分类号
摘要
Interpolation artifact is known to occur when the heart rate is decreased lower than the critical value for the specific pitch. The purpose of our study is to determine the minimum heart rate (minHR) for the specific pitch that provides images without interpolation artifact when using dual-source computed tomography (DSCT). We scanned the ‘thin slice thickness block’ of the CT performance phantom provided by the American Association of Physicists in Medicine using DSCT for variable pitches. Change in heart rate was simulated through ECG editing by changing R–R interval. Axial, sagittal, and coronal image sets were reconstructed and assessed for the presence and extent of interpolation artifact. MinHR at which no interpolation artifact was detected for each pitch value was determined. Length of interpolation artifact (LOA) on sagittal view was also measured when the heart rate was simulated at 10 bpm lower than the minHR on each pitch setting. MinHRs for each pitch value were 9–10 bpm from the estimated heart rate. However, minHR for the lowest pitch value 0.2, estimated heart value 40 bpm was 37 bpm. LOA was larger in the low heart rate condition. Measured values of minHR were correlated exactly with the calculated values. MinHRs that provide images without interpolation artifact for each pitch value when using DSCT were determined. The concept of minHR is important for obtaining high quality images of cardiac CT angiography when using DSCT.
引用
收藏
页码:103 / 109
页数:6
相关论文
共 37 条
[1]  
Klingenbeck-Regn K(1999)Subsecond multi-slice computed tomography: basics and application Eur J Radiol 31 110-124
[2]  
Schaller S(2007)Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns Radiol Med 112 31-46
[3]  
Flohr T(2006)Relationship between noise, dose, and pitch in cardiac multi–detector row CT Radiographics 26 1785-1794
[4]  
Francone M(2000)Cardiac imaging by means of electrocardiographically gated multisection spiral CT: initial experience Radiology 217 564-571
[5]  
Napoli A(2006)Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality Radiology 241 378-385
[6]  
Carbone I(2006)First performance evaluation of a dual-source CT (DSCT) system Eur Radiol 16 256-268
[7]  
Primak AN(2006)Contrast-enhanced coronary artery visualization by dual-source computed tomography—initial experience Eur J Radiol 57 331-335
[8]  
McCollough CH(2006)Dual-source CT cardiac imaging: initial experience Eur Radiol 16 1409-1415
[9]  
Bruesewitz MR(2007)Strategies for reduction of radiation dose in cardiac multislice CT Eur Radiol 17 2028-2037
[10]  
Ohnesorge BM(2007)Physics of cardiac imaging with multiple-row detector CT Radiographics 27 1495-1509