Carbon fiber–reinforced PEEK versus titanium implants: an in vitro comparison of susceptibility artifacts in CT and MR imaging

被引:0
|
作者
Theresa Krätzig
Klaus C. Mende
Malte Mohme
Helge Kniep
Marc Dreimann
Martin Stangenberg
Manfred Westphal
Tobias Gauer
Sven O. Eicker
机构
[1] University Medical Center Hamburg-Eppendorf,Department of Neurosurgery
[2] University Medical Center Hamburg-Eppendorf,Department of Neuroradiology
[3] University Medical Center Hamburg-Eppendorf,Division of Spine Surgery, Department of Trauma and Orthopedic Surgery
[4] University Medical Center Hamburg-Eppendorf,Department of Radiation Oncology
来源
Neurosurgical Review | 2021年 / 44卷
关键词
Susceptibility artifacts; Imaging; Pedicle screws; Titanium; Carbon fiber–reinforced PEEK; Spine surgery;
D O I
暂无
中图分类号
学科分类号
摘要
Artifacts in computed tomography (CT) and magnetic resonance imaging (MRI) due to titanium implants in spine surgery are known to cause difficulties in follow-up imaging, radiation planning, and precise dose delivery in patients with spinal tumors. Carbon fiber–reinforced polyetheretherketon (CFRP) implants aim to reduce these artifacts. Our aim was to analyze susceptibility artifacts of these implants using a standardized in vitro model. Titanium and CFRP screw-rod phantoms were embedded in 3% agarose gel. Phantoms were scanned with Siemens Somatom AS Open and 3.0-T Siemens Skyra scanners. Regions of interest (ROIs) were plotted and analyzed for CT and MRI at clinically relevant localizations. CT voxel–based imaging analysis showed a significant difference of artifact intensity and central overlay between titanium and CFRP phantoms. For the virtual regions of the spinal canal, titanium implants (ti) presented − 30.7 HU vs. 33.4 HU mean for CFRP (p < 0.001), at the posterior margin of the vertebral body 68.9 HU (ti) vs. 59.8 HU (CFRP) (p < 0.001) and at the anterior part of the vertebral body 201.2 HU (ti) vs. 70.4 HU (CFRP) (p < 0.001), respectively. MRI data was only visually interpreted due to the low sample size and lack of an objective measuring system as Hounsfield units in CT. CT imaging of the phantom with typical implant configuration for thoracic stabilization could demonstrate a significant artifact reduction in CFRP implants compared with titanium implants for evaluation of index structures. Radiolucency with less artifacts provides a better interpretation of follow-up imaging, radiation planning, and more precise dose delivery.
引用
收藏
页码:2163 / 2170
页数:7
相关论文
共 50 条
  • [31] A comparison of lesion mapping analyses based on CT versus MR imaging in stroke
    Moore, Margaret J.
    Jenkinson, Mark
    Griffanti, Ludovica
    Huygelier, Hanne
    Gillebert, Celine R.
    Demeyere, Nele
    NEUROPSYCHOLOGIA, 2023, 184
  • [32] Prospective Comparison of the Diagnostic Accuracy of MR Imaging versus CT for Acute Appendicitis
    Repplinger, Michael D.
    Pickhardt, Perry J.
    Robbins, Jessica B.
    Kitchin, Douglas R.
    Ziemlewicz, Tim J.
    Hetzel, Scott J.
    Golden, Sean K.
    Harringa, John B.
    Reeder, Scott B.
    RADIOLOGY, 2018, 288 (02) : 467 - 475
  • [33] Biomechanical effects of posterior pedicle screw-based instrumentation using titanium versus carbon fiber reinforced PEEK in an osteoporotic spine human cadaver model
    Oikonomidis, Stavros
    Greven, Johannes
    Bredow, Jan
    Eh, Madita
    Prescher, Andreas
    Fischer, Horst
    Thuering, Johannes
    Eysel, Peer
    Hildebrand, Frank
    Kobbe, Philipp
    Scheyerer, Max Joseph
    Herren, Christian
    CLINICAL BIOMECHANICS, 2020, 80
  • [34] Effect of different preparation methods on mechanical behaviors of carbon fiber-reinforced PEEK-Titanium hybrid laminates
    Ji, Chunming
    Wang, Bing
    Hu, Jiqiang
    Zhang, Chang'an
    Sun, Yuguo
    POLYMER TESTING, 2020, 85
  • [35] The dosimetric impact of stabilizing spinal implants in radiotherapy treatment planning with protons and photons: standard titanium alloy vs. radiolucent carbon-fiber-reinforced PEEK systems
    Mueller, Birgit S.
    Ryang, Yu-Mi
    Oechsner, Markus
    Duesberg, Mathias
    Meyer, Bernhard
    Combs, Stephanie E.
    Wilkens, Jan J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (08): : 6 - 14
  • [36] Navigation accuracy and assessability of carbon fiber-reinforced PEEK instrumentation with multimodal intraoperative imaging in spinal oncology
    Hubertus, Vanessa
    Wessels, Lars
    Frueh, Anton
    Tkatschenko, Dimitri
    Nulis, Irini
    Bohner, Georg
    Prinz, Vincent
    Onken, Julia
    Czabanka, Marcus
    Vajkoczy, Peter
    Hecht, Nils
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [37] Navigation accuracy and assessability of carbon fiber-reinforced PEEK instrumentation with multimodal intraoperative imaging in spinal oncology
    Vanessa Hubertus
    Lars Wessels
    Anton Früh
    Dimitri Tkatschenko
    Irini Nulis
    Georg Bohner
    Vincent Prinz
    Julia Onken
    Marcus Czabanka
    Peter Vajkoczy
    Nils Hecht
    Scientific Reports, 12
  • [38] Challenges Associated With Pencil Beam Scanning Proton Therapy for Spinal Tumors Following Surgical Stabilization: A Robustness Evaluation of Carbon Fiber Reinforced Polyetheretherketone (Carbon-PEEK) Versus Titanium
    Snider, J. W., III
    Mikroutsikos, L.
    Albertini, F.
    Bolsi, A.
    Walser, M.
    Weber, D. C.
    Schneider, R. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E699 - E700
  • [39] Intervertebral test spacers and postfusion MRI artifacting: A comparative in vitro study of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials
    Ernstberger, Thorsten
    Buchhorn, Gottfried
    Heidrich, Gabert
    CENTRAL EUROPEAN JOURNAL OF MEDICINE, 2009, 4 (04): : 496 - 500
  • [40] Anterior cervical discectomy and fusion with a dynamic translational plating versus a rigid carbon fiber reinforced PEEK plating system - a comparison study of radiographic parameters
    Burkhardt, Benedikt W.
    Kerolus, Mena G.
    Witiw, Christopher D.
    Oertel, Joachim M.
    Fessler, Richard G.
    BRITISH JOURNAL OF NEUROSURGERY, 2024, 38 (01) : 99 - 103