Thermoluminescence dosimetry for in-vivo verification of high dose rate brachytherapy for prostate cancer

被引:0
作者
R. Das
W. Toye
T. Kron
S. Williams
G. Duchesne
机构
[1] RMIT University,Department of Applied Physics
[2] Radiation Oncology Victoria,School of Medicine and Health Sciences
[3] University of Melbourne,Department of Physical Sciences
[4] Peter MacCallum Cancer Centre,undefined
来源
Australasian Physics & Engineering Sciences in Medicine | 2007年 / 30卷
关键词
HDR brachytherapy; prostate cancer; in-vivo dosimetry; TLD; treatment planning; rectum dose; urethra dose;
D O I
暂无
中图分类号
学科分类号
摘要
It was the aim of the study to verify dose delivered in urethra and rectum during High Dose Rate brachytherapy boost (HDRBB) of prostate cancer patients. During the first fraction of HDRBB measurement catheters were placed in the urethra and rectum of prostate cancer patients. These contained LiF:Mg,Ti Thermoluminescence Dosimetry (TLD) rods of 1mm diameter, with up to 11 detectors positioned every 16 mm separated by radio-opaque markers. A Lorentzian peak function was used to fit the data. Measurements from 50 patients were evaluated and measured doses were compared with predictions from the treatment planning system (Plato Vs 13.5 to 14.1). Prospective urinary and rectal toxicity scores were collected following treatment. In more than 90% of cases, the Lorentzian peak function provided a good fit to both experimental and planning urethral data (r2 > 0.9). In general there was good agreement between measured and predicted doses with the average difference between measured and planned maximum dose being 0.1Gy. No significant association between dose and any clinical endpoints was observed in 43 patients available for clinical evaluation. An average inferior shift of 2 mm between the plan and the measurement performed approximately 1 hour after the planning CT scan was found for the dose distribution in the cohort of patients for the urethra measurements. Rectal measurements proved to be more difficult to interpret as there is more variability of TLD position between planning and treatment. TLD in-vivo measurements are easily performed in urethra and rectum during HDR brachytherapy of prostate patients. They verify the delivery and provide information about the dose delivered to critical structures. The latter may be of particular interest if higher doses are to be given per fraction such as in HDR monotherapy.
引用
收藏
页码:178 / 184
页数:6
相关论文
共 50 条
  • [21] Prostate cancer boost using high-dose-rate brachytherapy: Impact of the learning curve on the dosimetry
    Benhaim, C.
    Chand, M-E
    Gal, J.
    Hijazi, H.
    Gautier, M.
    Hannoun-Levi, J-M
    CANCER RADIOTHERAPIE, 2014, 18 (07): : 659 - 665
  • [22] Use of Thermoluminescence Dosimetry for QA in High-Dose-Rate Skin Surface Brachytherapy with Custom-Flap Applicator
    Manna, Francesco
    Pugliese, Mariagabriella
    Buonanno, Francesca
    Gherardi, Federica
    Iannacone, Eva
    La Verde, Giuseppe
    Muto, Paolo
    Arrichiello, Cecilia
    SENSORS, 2023, 23 (07)
  • [23] Applicability of Glass Dosimeters for In-vivo Dosimetry in Brachytherapy
    Moon, Sun Young
    Son, Jaeman
    Yoon, Myonggeun
    Jeang, EunHee
    Lim, Young Kyung
    Chung, Weon Kyu
    Kim, Dong Wook
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 72 (11) : 1320 - 1325
  • [24] High Dose Rate Brachytherapy for Prostate Cancer - Physical Aspects
    Miszczak, Dorota
    Dabrowski, Tomasz
    Nahajowski, Damian
    Lesiak, Jan
    Kisielewiez, Kamil
    Kudzia, Roksana
    Kukielka, Andrzej
    Walasek, Tomasz
    Byrski, Edward
    Waligorski, Michal
    SOME ASPECTS OF MEDICAL PHYSICS - IN VIVO AND IN VITRO STUDIES, 2010, 1 : 182 - 185
  • [25] High dose-rate brachytherapy in the treatment of prostate cancer
    Mendez, Lucas C.
    Morton, Gerard C.
    TRANSLATIONAL ANDROLOGY AND UROLOGY, 2018, 7 (03) : 357 - 370
  • [26] Two-dimensional in vivo rectal dosimetry during high-dose-rate brachytherapy for cervical cancer: a phantom study
    Jeang, Eun Hee
    Goh, Youngmoon
    Cho, Kwan Ho
    Min, Soonki
    Choi, Sang Hyoun
    Jeong, Hojin
    Jo, Kwanghyun
    Lee, Nuri
    Song, Sanghyeon
    Lee, Se Byeong
    Shin, Dongho
    Kim, Yeon-Joo
    Kim, Joo-Young
    Kim, Dae Yong
    Hwang, Ui-Jung
    Lim, Young Kyung
    ACTA ONCOLOGICA, 2018, 57 (10) : 1359 - 1366
  • [27] High dose rate brachytherapy as monotherapy for localised prostate cancer
    Strouthos, Iosif
    Tselis, Nikolaos
    Chatzikonstantinou, Georgios
    Butt, Saeed
    Baltas, Dimos
    Bon, Dimitra
    Milickovic, Natasa
    Zamboglou, Nikolaos
    RADIOTHERAPY AND ONCOLOGY, 2018, 126 (02) : 270 - 277
  • [28] An in vivo investigative protocol for HDR prostate brachytherapy using urethral and rectal thermoluminescence dosimetry
    Toye, Warren
    Das, Ram
    Kron, Tomas
    Franich, Rick
    Johnston, Peter
    Duchesne, Gillian
    RADIOTHERAPY AND ONCOLOGY, 2009, 91 (02) : 243 - 248
  • [29] Towards real time in-vivo rectal dosimetry during trans-rectal ultrasound based high dose rate prostate brachytherapy using MOSkin dosimeters
    Poder, Joel
    Howie, Andrew
    Brown, Ryan
    Bucci, Joseph
    Rosenfeld, Anatoly
    Enari, Komiti
    Schreiber, Kristine
    Carrara, Mauro
    Bece, Andrej
    Malouf, David
    Cutajar, Dean
    RADIOTHERAPY AND ONCOLOGY, 2020, 151 : 273 - 279
  • [30] Dwell time verification in brachytherapy based on time resolved in vivo dosimetry
    Johansen, J. G.
    Kertzscher, G.
    Jorgensen, E. B.
    Rylander, S.
    Bentzen, L.
    Hokland, S. B.
    Sondergaard, C. S.
    With, A. K. M.
    Buus, S.
    Tanderup, K.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 60 : 156 - 161