His bundle pacing capture threshold stability during long-term follow-up and correlation with lead slack

被引:34
|
作者
Beer, Dominik [1 ]
Subzposh, Faiz A. [1 ]
Colburn, Shaun [1 ]
Naperkowski, Angela [1 ]
Vijayaraman, Pugazhendhi [1 ]
机构
[1] Cardiac Electrophysiol Geisinger Heart Inst MC 36, 1000 E Mt Blvd, Wilkes Barre, PA 18711 USA
来源
EUROPACE | 2021年 / 23卷 / 05期
关键词
His bundle pacing; Conduction system pacing; Lead slack; Pacemaker; Capture threshold; Lead revision; PERMANENT;
D O I
10.1093/europace/euaa350
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims His bundle pacing (HBP) is the most physiologic form of pacing. Long-term HBP capture threshold stability and its relation to lead characteristics at the time of implantation have not been adequately described. The aim of this study was to characterize HB capture threshold in follow-up and to identify potential lead characteristics predictive of lead capture instability. Methods and results Consecutive patients with successful HBP for bradycardia indications were identified from the Geisinger HBP registry. His bundle capture thresholds, baseline comorbidities, and radiographic lead slack characteristics were analysed. An increase in HB capture threshold >= 1V above implant values at any time during follow-up was tracked. Forty-four of the 294 studied (15%) experienced HB capture threshold increase by >= 1 V. Threshold increase was seen early (41% by 8 weeks, 66% by 1 year). Eighteen (6%) patients required lead revision in follow-up. Abnormal slack shape was associated with a trend toward capture threshold increase [hazard ratio (HR) 2.07; 95% confidence interval (CI) 0.9-4.6; P=0.08]. Non-perpendicular angle of lead insertion on radiography was associated with the capture threshold increase (HR 2.81, 95% CI 1.4-5.8; P <0.01). Conclusion His bundle capture threshold remains stable in the majority (85%) of patients. Implant characteristics may predict the threshold rise. Further evaluation of the aetiology of threshold increase and design changes in lead and delivery systems may lead to chronically stable capture thresholds. [GRAPHICS] .
引用
收藏
页码:757 / 766
页数:10
相关论文
共 50 条
  • [21] RESULTS OF LONG-TERM FOLLOW-UP OF PERMANENT ATRIAL-PACING IN PATIENTS WITH SICK SINUS SYNDROME
    KOLK, R
    LAANE, P
    ROOSE, I
    SAMARUTEL, J
    KARDIOLOGIYA, 1994, 34 (7-8) : 5 - 7
  • [22] Long-Term Follow-up Impact of Dual-Chamber Pacing on Patients with Hypertrophic Obstructive Cardiomyopathy
    Hu Yue-Cheng
    Li Zuo-Cheng
    Li Xi-Ming
    Yuan, David Zhe
    Jin Dong-Xia
    Zhang Ying-Yi
    Yang Hui-Ming
    Cong Hong-Liang
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (01): : 86 - 93
  • [23] Upgrade to his bundle pacing in pacing-dependent patients referred for pulse generator change: Feasibility and intermediate term follow up
    Ye, Yang
    Zhang, Zuwen
    Sheng, Xia
    Wang, Bei
    Chen, Shiquan
    Pan, Yiwen
    Lan, Jingliang
    Sun, Yaxun
    Luan, Yi
    Jiang, Chenyang
    Fu, Guosheng
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2018, 260 : 88 - 92
  • [24] Long-term follow-up of DDDR closed-loop cardiac pacing for the prevention of recurrent vasovagal syncope
    Bortnik, Miriam
    Occhetta, Eraldo
    Dell'Era, Gabriele
    Secco, Gioel G.
    Degiovanni, Anna
    Plebani, Laura
    Marino, Paolo
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2012, 13 (04) : 242 - 245
  • [25] Feasibility of Permanent His Bundle Pacing in the Elderly vs the Very Elderly. A Single-Center Mid-Term Follow-Up Study
    Pestrea, Catalin
    Cicala, Ecaterina
    Gherghina, Alexandra
    Ortan, Florin
    Pop, Dana
    CLINICAL INTERVENTIONS IN AGING, 2023, 18 : 941 - 949
  • [26] Outcomes of His-bundle pacing upgrade after long-term right ventricular pacing and/or pacing-induced cardiomyopathy: Insights into disease progression
    Vijayaraman, Pugazhendhi
    Herweg, Bengt
    Dandamudi, Gopi
    Mittal, Suneet
    Bhatt, Advay G.
    Marcantoni, Lina
    Naperkowski, Angela
    Sharma, Parikshit S.
    Zanon, Francesco
    HEART RHYTHM, 2019, 16 (10) : 1554 - 1561
  • [27] Pacemaker replacement in nonagenarians: Procedural safety and long-term follow-up
    Loirat, Aurelie
    Feneon, Damien
    Behaghel, Albin
    Behar, Nathalie
    Le Helloco, Alain
    Mabo, Philippe
    Daubert, Jean-Claude
    Leclercq, Christophe
    Martins, Raphael P.
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2015, 108 (6-7) : 367 - 374
  • [28] Programming and follow-up of patients with His bundle pacing; [Programmierung und Nachverfolgung bei Patienten mit His-Bündel-Stimulation]
    Bakelants E.
    Zweerink A.
    Burri H.
    Herzschrittmachertherapie + Elektrophysiologie, 2020, 31 (2) : 177 - 182
  • [29] Single chamber atrial pacing:: A realistic option in sinus node disease:: A long-term follow-up study of 213 patients
    Hoeijer, Carl J.
    Hoeglund, Peter
    Schueller, Hans
    Brandt, Johan
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (06): : 740 - 747
  • [30] Clinical safety and efficacy of single-chamber atrial pacing-in sick sinus syndrome:: Long-term follow-up
    Moriñigo, JL
    Arribas, A
    Ledesma, C
    Sánchez, PL
    Martín, F
    Luengo, CM
    REVISTA ESPANOLA DE CARDIOLOGIA, 2002, 55 (12): : 1267 - 1272