His bundle pacing capture threshold stability during long-term follow-up and correlation with lead slack

被引:36
作者
Beer, Dominik [1 ]
Subzposh, Faiz A. [1 ]
Colburn, Shaun [1 ]
Naperkowski, Angela [1 ]
Vijayaraman, Pugazhendhi [1 ]
机构
[1] Cardiac Electrophysiol Geisinger Heart Inst MC 36, 1000 E Mt Blvd, Wilkes Barre, PA 18711 USA
来源
EUROPACE | 2021年 / 23卷 / 05期
关键词
His bundle pacing; Conduction system pacing; Lead slack; Pacemaker; Capture threshold; Lead revision; PERMANENT;
D O I
10.1093/europace/euaa350
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims His bundle pacing (HBP) is the most physiologic form of pacing. Long-term HBP capture threshold stability and its relation to lead characteristics at the time of implantation have not been adequately described. The aim of this study was to characterize HB capture threshold in follow-up and to identify potential lead characteristics predictive of lead capture instability. Methods and results Consecutive patients with successful HBP for bradycardia indications were identified from the Geisinger HBP registry. His bundle capture thresholds, baseline comorbidities, and radiographic lead slack characteristics were analysed. An increase in HB capture threshold >= 1V above implant values at any time during follow-up was tracked. Forty-four of the 294 studied (15%) experienced HB capture threshold increase by >= 1 V. Threshold increase was seen early (41% by 8 weeks, 66% by 1 year). Eighteen (6%) patients required lead revision in follow-up. Abnormal slack shape was associated with a trend toward capture threshold increase [hazard ratio (HR) 2.07; 95% confidence interval (CI) 0.9-4.6; P=0.08]. Non-perpendicular angle of lead insertion on radiography was associated with the capture threshold increase (HR 2.81, 95% CI 1.4-5.8; P <0.01). Conclusion His bundle capture threshold remains stable in the majority (85%) of patients. Implant characteristics may predict the threshold rise. Further evaluation of the aetiology of threshold increase and design changes in lead and delivery systems may lead to chronically stable capture thresholds. [GRAPHICS] .
引用
收藏
页码:757 / 766
页数:10
相关论文
共 20 条
[1]  
Abdelrahman M, 2018, J AM COLL CARDIOL, V71, P2319, DOI [10.1016/j.jacc.2018.02.048, 10.1016/S0735-1097(18)33122-X]
[2]   His-bundle pacing: impact of social media [J].
Beer, Dominik ;
Dandamudi, Gopi ;
Mandrola, John M. ;
Friedman, Paul A. ;
Vijayaraman, Pugazhendhi .
EUROPACE, 2019, 21 (10) :1445-1450
[3]   Clinical Outcomes of Selective Versus Nonselective His Bundle Pacing [J].
Beer, Dominik ;
Sharma, Parikshit S. ;
Subzposh, Faiz A. ;
Naperkowski, Angela ;
Pietrasik, Grzegorz M. ;
Durr, Brendan ;
Qureshi, Maria ;
Panikkath, Ragesh ;
Abdelrahman, Mohamed ;
Williams, Brent A. ;
Hanifin, Jillian L. ;
Zimberg, Ryan ;
Austin, Kelly ;
Macuch, Brooke ;
Trohman, Richard G. ;
Vanenkevort, Erin A. ;
Dandamudi, Gopi ;
Vijayaraman, Pugazhendhi .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (07) :766-774
[4]   Pacemaker and ICD leads: Strategies for long-term management [J].
Borek, P. Peter ;
Wilkoff, Bruce L. .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2008, 23 (01) :59-72
[5]   2019 ESC Guidelines for the management of patients with supraventricular tachycardia The Task Force for the management of patients with supraventricular tachycardia of the European Society of Cardiology (ESC): Developed in collaboration with the Association for European Paediatric and Congenital Cardiology (AEPC) [J].
Brugada, Josep ;
Katritsis, Demosthenes G. ;
Arbelo, Elena ;
Arribas, Fernando ;
Bax, Jeroen J. ;
Blomstrom-Lundqvist, Carina ;
Calkins, Hugh ;
Corrado, Domenico ;
Deftereos, Spyridon G. ;
Diller, Gerhard-Paul ;
Gomez-Doblas, Juan J. ;
Gorenek, Bulent ;
Grace, Andrew ;
Ho, Siew Yen ;
Kaski, Juan-Carlos ;
Kuck, Karl-Heinz ;
Lambiase, Pier David ;
Sacher, Frederic ;
Sarquella-Brugada, Georgia ;
Suwalski, Piotr ;
Zaza, Antonio .
EUROPEAN HEART JOURNAL, 2020, 41 (05) :655-720
[6]   How to perform permanent His bundle pacing in routine clinical practice [J].
Dandamudi, Gopi ;
Vijayaraman, Pugazhendhi .
HEART RHYTHM, 2016, 13 (06) :1362-1366
[7]   Permanent, direct his-bundle pacing - A novel approach to cardiac pacing in patients with normal His-purkinje activation [J].
Deshmukh, P ;
Casavant, DA ;
Romanyshyn, M ;
Anderson, K .
CIRCULATION, 2000, 101 (08) :869-877
[8]   Cardiac Resynchronization Therapy in Patients With Nonischemic Cardiomyopathy Using Left Bundle Branch Pacing [J].
Huang, Weijian ;
Wu, Shengjie ;
Vijayaraman, Pugazhendhi ;
Su, Lan ;
Chen, Xueying ;
Cai, Bingni ;
Zou, Jiangang ;
Lan, Rongfang ;
Fu, Guosheng ;
Mao, Guangyun ;
Ellenbogen, Kenneth A. ;
Whinnett, Zachary, I ;
Tung, Roderick .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2020, 6 (07) :849-858
[9]  
Huang Weijian, 2018, Card Electrophysiol Clin, V10, P519, DOI 10.1016/j.ccep.2018.05.016
[10]  
Kusumoto FM, 2019, CIRCULATION, V140, pE333, DOI 10.1161/CIR.0000000000000627