Development of a multiparametric analyser based on ISFET sensors applied to process control in the wine industry

被引:22
作者
Artigas, J
Jiménez, C [1 ]
Domínguez, C
Mínguez, S
Gonzalo, A
Alonso, J
机构
[1] CSIC, IMB CNM, Inst Microelect Barcelona, Bellaterra 08193, Barcelona, Spain
[2] Inst Catala Vinya & Vi, INCAVI, Vilafranca del Penedes, Barcelona, Spain
关键词
wine monitoring; ion-exchange resin; tartaric stabilisation; ISFET sensors; continuous flow system;
D O I
10.1016/S0925-4005(02)00464-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The continuous monitoring of processes in the food industry, and in particular in wine production, holds great interest as it can help to increase the quality of the goods produced and also the efficiency of the production process. One process that is feasible to control is the stabilisation of wine by means of ion-exchange resins. The use of this technique is of great promise since it allows continuous wine treatment and also on-line monitoring systems can be applied to determine the saturation of the resin in addition to the fraction of wine requiring stabilisation. The development of a pH, calcium and potassium analyser for on-line monitoring of wine in a continuous flow system is reported here. The potentiometric sensors used are ion-selective field-effect transistors (ISFETs). The stability of these sensors in hydro-alcoholic media was studied obtaining a sensitivity of 56 mV per decade for the pH ISFET and sensitivities of -26 mV per decade and -50 mV per decade for calcium and potassium sensors, respectively. The long-term stability of pH-ISFETs was 7 and 4 months for cation sensors. Results obtained in the continuous flow system with wine samples demonstrated that the proposed system is capable to detect the saturation point of the ion-exchange resin in wine stabilisation process and also to determine the ratio of wine required to stabilise it. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:199 / 204
页数:6
相关论文
共 14 条
  • [1] FLOW-THROUGH PH-ISFET AS DETECTOR IN AUTOMATED DETERMINATIONS
    ALEGRET, S
    BARTROLI, J
    JIMENEZ, C
    DELVALLE, M
    DOMINGUEZ, C
    CABRUJA, E
    MERLOS, A
    [J]. ELECTROANALYSIS, 1991, 3 (4-5) : 349 - 354
  • [2] PH-ISFET WITH NMOS TECHNOLOGY
    ALEGRET, S
    BARTROLI, J
    JIMENEZ, C
    DELVALLE, M
    DOMINGUEZ, C
    CABRUJA, E
    MERLOS, A
    [J]. ELECTROANALYSIS, 1991, 3 (4-5) : 355 - 360
  • [3] Development of a photopolymerisable membrane for calcium ion sensors -: Application to soil drainage waters
    Artigas, J
    Beltran, A
    Jiménez, C
    Bartrolí, J
    Alonso, J
    [J]. ANALYTICA CHIMICA ACTA, 2001, 426 (01) : 3 - 10
  • [4] Beltran A, 2002, ELECTROANAL, V14, P213, DOI 10.1002/1521-4109(200202)14:3<213::AID-ELAN213>3.0.CO
  • [5] 2-U
  • [6] PHOTOCURABLE POLYMER MATRICES FOR POTASSIUM-SENSITIVE ION-SELECTIVE ELECTRODE MEMBRANES
    BRATOV, A
    ABRAMOVA, N
    MUNOZ, J
    DOMINGUEZ, C
    ALEGRET, S
    BARTROLI, J
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (19) : 3589 - 3595
  • [7] BRATOV A, 1995, SENSOR ACTUAT B-CHEM, V24, P823
  • [8] BRUGIRARD A, 1991, ASPECTS PRACTIQUES T
  • [9] PHOTOCURED POLYMERS IN ION-SELECTIVE ELECTRODE MEMBRANES
    CATTRALL, RW
    ILES, PJ
    HAMILTON, IC
    [J]. ANALYTICA CHIMICA ACTA, 1985, 169 (MAR) : 403 - 406
  • [10] Flanzy C., 2000, Enologia: Fundamentos Cientificos y Tecnologicos