Automatic identification of implantable cardioverter-defibrillator lead problems using intracardiac electrograms

被引:8
作者
Gunderson, BD [1 ]
Patel, AS [1 ]
Bounds, CA [1 ]
机构
[1] Medtronic Inc, Minneapolis, MN USA
来源
COMPUTERS IN CARDIOLOGY 2002, VOL 29 | 2002年 / 29卷
关键词
D O I
10.1109/CIC.2002.1166722
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implantable Cardioverter-Defibrillator (ICD) lead problems can cause inappropriate painful shocks or inappropriately withheld lifesaving shocks. ICDs contain storage for detected spontaneous episodes that can be analyzed to characterize lead performance. The goal of this project was to develop an automatic lead problem identification algorithm using stored episode data. The algorithm combines sensed RR interval patterns and electrogram (EGM) characteristics to identify non-cardiac (NC) oversensing (OS) problems (e.g. lead failures) and cardiac (C) OS problems (e.g. T-wave OS). Stored episodes from 59 patients with OS and 147 patients with no OS were used for evaluation. The sensitivities to identify the lead problems were 97.7% for NC-OS and 86.7% for C-OS with a specificity of 98.0%. Analysis of stored episodes with EGM maybe used to identify ICD lead problems with very high sensitivity and specificity.
引用
收藏
页码:121 / 124
页数:4
相关论文
共 5 条
  • [1] Sensing lead failure in implantable defibrillators: A comparison of two commonly used leads
    Degeratu, FT
    Khalighi, K
    Peters, RW
    Shorofsky, SR
    Gold, MR
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (01) : 21 - 24
  • [2] Sensing lead-related complications in patients with transvenous implantable cardioverter-defibrillators
    Lawton, JS
    Ellenbogen, KA
    Wood, MA
    Stambler, BS
    Herre, JM
    Nath, S
    Bernstein, RC
    DiMarco, JP
    Haines, DE
    Szentpetery, S
    Baker, LD
    Damiano, RJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 78 (06) : 647 - 651
  • [3] Late complications in patients with pectoral defibrillator implants with transvenous defibrillator lead systems:: High incidence of insulation breakdown
    Mehta, D
    Nayak, HM
    Singson, M
    Chao, S
    Pe, E
    Camuñas, JL
    Gomes, JA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10): : 1893 - 1900
  • [4] Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators
    Reiter, MJ
    Mann, DE
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) : 542 - 558
  • [5] SENSING PACING LEAD COMPLICATIONS WITH A NEWER GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - WORLDWIDE EXPERIENCE FROM THE GUARDIAN-ATP-4210 CLINICAL-TRIAL
    STAMBLER, BS
    WOOD, MA
    DAMIANO, RJ
    GREENWAY, PS
    SMUTKA, ML
    ELLENBOGEN, KA
    BATSFORD, W
    BAUERNFEIND, R
    BECKMAN, K
    BROWNSTEIN, S
    CASTLE, L
    DEERING, T
    ELLENBOGEN, K
    ESTES, NAM
    FRANKLIN, J
    FREIHLING, T
    HAISTY, K
    KEHOE, R
    LOVE, C
    LUCERI, R
    NACCARELLI, G
    PACIFICO, A
    PARSONNET, V
    PETROPOULOS, T
    SAKSENA, S
    SHARMA, A
    SINGER, I
    VENDITTI, F
    WALTER, P
    ZIPES, D
    DAVIS, MJ
    HELLESTRAND, K
    KUCHAR, D
    POWELL, A
    STAFFORD, W
    UTHER, J
    VOHRA, J
    DORIAN, P
    GULAMHUSEIN, S
    KERR, CR
    ROSENGARTEN, M
    ROY, D
    TANG, ASL
    WILLIAMS, W
    YEE, R
    MOLLER, M
    THOMSEN, PEB
    ALIOT, E
    LECLERQ, JF
    LEMAREC, H
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (01) : 123 - 132