Nonlinearity measurements of high-power laser detectors at NIST

被引:11
作者
Li, XY [1 ]
Scott, T [1 ]
Yang, S [1 ]
Cromer, C [1 ]
Dowell, M [1 ]
机构
[1] Natl Inst Stand & Technol, Boulder, CO 80305 USA
关键词
attenuation method; calibration; calorimeter; chopper wheel; high-power laser; nonlinearity measurement; thermal detector;
D O I
10.6028/jres.109.030
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We briefly explain the fundamentals of detector nonlinearity applicable to both electrical and optical nonlinearity measurements. We specifically discuss the attenuation method for optical nonlinearity measurement that the NIST system is based upon, and we review the possible sources of nonlinearity inherent to thermal detectors used with high-power lasers. We also describe, in detail, the NIST nonlinearity measurement system, in which detector responsivity can be measured at wavelengths of 1.06 mum and 10.6 mum, over a power range from 1 W to 1000 W. We present the data processing method used and show measurement results depicting both positive and negative nonlinear behavior. The expanded uncertainty of a typical NIST high-power laser detector calibration including nonlinearity characterization is about 1.3%.
引用
收藏
页码:429 / 434
页数:6
相关论文
共 9 条
[1]  
Lara H, 2003, LASER FOCUS WORLD, V39, P137
[2]   Power measurement standards for high-power lasers:: comparison between the NIST and the PTB [J].
Li, X ;
Scott, TR ;
Cromer, CL ;
Keenan, D ;
Brandt, F ;
Möstl, K .
METROLOGIA, 2000, 37 (05) :445-447
[3]  
MOONY B, 1996, IND LASER REV NOV, P25
[4]  
Scott T. R., 1988, Proceedings of the SPIE - The International Society for Optical Engineering, V888, P48, DOI 10.1117/12.944232
[5]  
Taylor B. N., 1994, 1297 NIST
[6]  
Vayshenker I, 2000, NIST SPECIAL PUBLICA, P250
[7]  
WEST DE, 1977, 685 NBS
[8]   REFERENCE CALORIMETER FOR LASER ENERGY MEASUREMENTS [J].
WEST, ED ;
SCHMIDT, LB ;
RASMUSSEN, AL ;
CASE, WE .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1972, A 76 (01) :13-+
[9]  
Yang S, 1995, NIST TECHNICAL NOTE, V1376