Minimum sample frequency for multichannel intraluminal impedance measurement of the oesophagus

被引:26
作者
Bredenoord, AJ
Weusten, BLAM
Timmer, R
Smout, AJPM
机构
[1] St Antonius Hosp, Dept Gastroenterol, NL-3430 EM Nieuwegein, Netherlands
[2] Univ Med Ctr, Gastrointestinal Res Unit, Utrecht, Netherlands
关键词
electrical impedance; gastro-oesophageal reflux; oesophageal motility; oesophageal transit testing;
D O I
10.1111/j.1365-2982.2004.00575.x
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
In all systems for impedance monitoring signals are stored in digital format after analog-to-digital conversion at a predefined rate, the sample frequency. We aimed to find the minimum sample frequency required to evaluate oesophageal transit and gastro-oesophageal reflux studies using impedance monitoring. In 10 healthy subjects and five patients with gastro-oesophageal reflux disease impedance signals were acquired during a 90-min postprandial period, using a sample frequency of 1000 Hz. Additionally, 10 liquid swallows were recorded in the healthy subjects. A dedicated computer program was used to derive, from the original 1000-Hz files, series of new data files with sample frequencies of 500, 200, 100, 50, 20, 10, 8, 5 and 4 Hz. In all of these files, liquid and gas reflux events were identified. In the analysis of the swallows, bolus head advance time (BHAT) and total bolus transit time were measured. Reflux events were detectable down to a sample frequency of 50 Hz. In transit analysis errors for BHAT at frequencies below 8 Hz errors exceeded 2.5%. Therefore, in impedance monitoring the minimum sample frequency is 50 Hz for the detection of reflux and 8 Hz for the evaluation of oesophageal transit.
引用
收藏
页码:713 / 719
页数:7
相关论文
共 16 条
  • [1] [Anonymous], 1991, J GASTROINTEST MOTIL, V3, P151, DOI DOI 10.1111/J.1365-2982.1991.TB00061.X
  • [2] Aerophagia, gastric, and supragastric belching: a study using intraluminal electrical impedance monitoring
    Bredenoord, AJ
    Weusten, BLAM
    Sifrim, D
    Timmer, R
    Smout, AJPM
    [J]. GUT, 2004, 53 (11) : 1561 - 1565
  • [3] MEASURING ESOPHAGEAL MOTILITY WITH A NEW INTRALUMINAL IMPEDANCE DEVICE - FIRST CLINICAL-RESULTS IN REFLUX PATIENTS
    FASS, J
    SILNY, J
    BRAUN, J
    HEINDRICHS, U
    DREUW, B
    SCHUMPELICK, V
    RAU, G
    [J]. SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1994, 29 (08) : 693 - 702
  • [4] Detection of small bolus volumes using multiple intraluminal impedance in preterm infants
    Peter, CS
    Wiechers, C
    Bohnhorst, B
    Silny, J
    Poets, CF
    [J]. JOURNAL OF PEDIATRIC GASTROENTEROLOGY AND NUTRITION, 2003, 36 (03) : 381 - 384
  • [5] Patterns of gas and liquid reflux during transient lower oesophageal sphincter relaxation: a study using intraluminal electrical impedance
    Sifrim, D
    Silny, J
    Holloway, RH
    Janssens, JJ
    [J]. GUT, 1999, 44 (01) : 47 - 54
  • [6] Sifrim D, 2001, AM J GASTROENTEROL, V96, P647
  • [7] SILNY J, 1993, J GASTROINTEST MOTIL, V5, P107
  • [8] Relevance of ineffective oesophageal motility during oesophageal acid clearance
    Simrén, M
    Silny, J
    Holloway, R
    Tack, J
    Janssens, J
    Sifrim, D
    [J]. GUT, 2003, 52 (06) : 784 - 790
  • [9] Gastroesophageal reflux in infants: Evaluation of a new intraluminal impedance technique
    Skopnik, H
    Silny, J
    Heiber, O
    Schulz, J
    Rau, G
    Heimann, G
    [J]. JOURNAL OF PEDIATRIC GASTROENTEROLOGY AND NUTRITION, 1996, 23 (05) : 591 - 598
  • [10] Talley NJ, 1999, GUT, V45, P37