Simple and robust ICP-MS method for simultaneous determination of serum Co and Cr in routine clinical practice

被引:15
作者
Choi, Hyun-Jung [1 ,2 ,3 ]
Lim, Seung-Jae [4 ]
Park, Youn-Soo [4 ]
Lee, Soo-Youn [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Lab Med & Genet, Seoul 135710, South Korea
[2] Chonnam Natl Univ, Sch Med, Dept Lab Med, Gwanju, South Korea
[3] Chonnam Natl Univ Hosp, Gwanju, South Korea
[4] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Orthoped Surg, Seoul 135710, South Korea
关键词
ICP-MS; Cobalt; Chromium; Arthroplasty; PLASMA-MASS SPECTROMETRY; HIP-ARTHROPLASTY PATIENTS; ON-METAL PROSTHESES; WHOLE-BLOOD; ION LEVELS; CHROMIUM; COBALT; REPLACEMENT; DEBRIS;
D O I
10.1016/j.cca.2014.09.019
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Chromium (Cr) and cobalt (Co) metal ions released from orthopedic prostheses, such as metal-on-metal (MoM) bearings, may be related to wear of the bearing surface. These ions may serve as indicators of the in vivo performance of MoM bearing surfaces. We developed a simple and sensitive assay suitable for the simultaneous measurement of serum Cr and Co using inductively coupled plasma-mass spectrometry (ICP-MS) in clinical routine laboratory practice to evaluate analytical performance and clinical utility. Methods: We evaluated the linearity, precision, and lower limit of quantification (LOQ) of an ICP-MS method to determine serum Cr and Co concentration in accordance with U.S. Food and Drug Administration guidance for bioanalytical method validation. This method was used to determine the serum Cr and Co levels of 236 clinical samples from 74 patients with MoM hip prosthesis and 51 healthy controls. Results: The assay showed good linearity (0-20 mu g/l, r(2) > 0.999) and satisfactory accuracy for all tested concentrations of Cr and Co (%bias, -1.5% to 2.5%, -3.3% to 1.6%, respectively). The CVs for both metal ions were <10% for LOQ (0.02 mu g/l Cr, 0.01 mu g/l Co) and within 5% for other concentrations (intra- and inter-assay CV, 1.2-2.6 and 1.9-4.4% of Cr; 1.4-2.7 and 1.9-4.7% of Co). The serum Cr and Co concentrations (mean +/- SD) were 0.60 +/- 0.12 mu g/l and 0.29 +/- 0.15 mu g/l in 51 healthy subjects. In 74 patients (185 serum samples) after HRA (median duration implanted, 48 months; range, 1-138 months), the concentrations of Cr and Co were 2.75 mu g/l (range, 0.61-116.80 mu g/l) and 1.49 mu g/l (range, 0.12-127.80 mu g/l), respectively. Conclusions: ICP-MS can simultaneously quantify serum Cr and Co levels with suitable analytical performance including a wide analytical range and very low LOQ. This assay will be very useful for evaluating the nutritional status of Cr and Co in healthy humans as well as monitoring the performance of orthopedic prostheses after hip resurfacing arthroplasty. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:91 / 96
页数:6
相关论文
共 25 条
[1]  
Alimonti Alessandro, 2005, Ann Ist Super Sanita, V41, P181
[2]  
[Anonymous], 2001, Guidance for industry - bioanalytical method validation
[3]   Metal ion levels in the blood of patients after hip resurfacing: A comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses [J].
Antoniou, John ;
Zukor, David J. ;
Mwale, Fackson ;
Minarik, William ;
Petit, Alain ;
Huk, Olga L. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :142-148
[4]   Evaluation of the Method for Analyzing Chromium, Cobalt and Titanium Ion Levels in the Blood Following Hip Replacement with a Metal-on-Metal Prosthesis [J].
Barry, Janie ;
Lavigne, Martin ;
Vendittoli, Pascal-Andre .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2013, 37 (02) :90-96
[5]   Evaluation of methods for trace-element determination with emphasis on their usability in the clinical routine laboratory [J].
Bolann, B. J. ;
Rahil-Khazen, R. ;
Henriksen, H. ;
Isrenn, R. ;
Ulvik, R. J. .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2007, 67 (04) :353-366
[6]  
Bozic KJ, 2004, CLIN ORTHOP RELAT R, P18
[7]  
Case CP, 2001, CLIN CHEM, V47, P275
[8]   Highly sensitive measurement of whole blood chromium by inductively coupled plasma mass spectrometry [J].
Cieslak, Wendy ;
Pap, Kathleen ;
Bunch, Dustin R. ;
Reineks, Edmunds ;
Jackson, Raymond ;
Le, Roxanne Stein ;
Wang, Sihe .
CLINICAL BIOCHEMISTRY, 2013, 46 (03) :266-270
[9]   Determination of trace elements in serum by dynamic reaction cell inductively coupled plasma mass spectrometry - Developing of a method with a desolvating system nebulizer [J].
D'Ilio, S. ;
Violante, N. ;
Caimi, S. ;
Di Gregorio, M. ;
Petrucci, F. ;
Senofonte, O. .
ANALYTICA CHIMICA ACTA, 2006, 573 :432-438
[10]   Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream [J].
De Pasquale, Dalila ;
Stea, Susanna ;
Squarzoni, Stefano ;
Bordini, Barbara ;
Amabile, Marilina ;
Catalani, Simona ;
Apostoli, Pietro ;
Toni, Aldo .
INTERNATIONAL ORTHOPAEDICS, 2014, 38 (03) :469-475