High-Precision Electrical Impedance Tomography Data Acquisition System for Brain Imaging

被引:100
作者
Shi, Xuetao [1 ]
Li, Weichen [1 ]
You, Fusheng [1 ]
Huo, Xuyang [2 ]
Xu, Canhua [1 ]
Ji, Zhenyu [1 ]
Liu, Ruigang [1 ]
Liu, Benyuan [1 ]
Li, Yandong [1 ]
Fu, Feng [1 ]
Dong, Xiuzhen [1 ]
机构
[1] Fourth Mil Med Univ, Sch Biomed Engn, Dept Biomed Engn, Xian 710032, Shaanxi, Peoples R China
[2] Jilin Med Coll, Fac Biomed Engn, Jilin 132013, Jilin, Peoples R China
基金
中国国家自然科学基金;
关键词
Electrical impedance; brain imaging; data acquisition; imaging monitoring; EIT SYSTEM; CALIBRATION; ELECTRODES; DESIGN; EEG;
D O I
10.1109/JSEN.2018.2836336
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In electrical impedance tomography (EIT), it is difficult to obtain the intracranial impedance due to the highly resistive skull enclosing the brain. Therefore, a high-precision data acquisition system is required for brain EIT. In this paper, we used a high-precision digital synthesis method and a digital demodulation technique with high noise immunity to eliminate random errors. Moreover, we focused on two problems encountered during EIT data acquisition: 1) the shunt effect on the excitation current due to the distributed capacitance between electrodes and ground and 2) high common-mode voltages in the boundary measurements. We designed a new electrode interface to reduce the influence of the distributed capacitance and a programmable current source to accurately compensate for the excited current. We also proposed a new voltmeter circuit with improved CMRR. Overall, this EIT data acquisition system can produce a programmable current with SNR greater than 89 dB. It can also measure the voltage difference precisely with CMRR higher than 75 dB with a 1-k Omega impedance imbalance. The results on a calibration model show that this system has a high SNR of 83 dB and a low reciprocity error of 0.125%. In addition, EIT imaging results were acquired using a brain physical phantom. The system can detect small disturbances of 0.35% in volume (1.99% of the cross-sectional area) and 17% in resistivity. Experiments on healthy volunteers also suggest that small intracranial impedance variations due to temporary occlusion and reperfusion of the unilateral carotid artery may be monitored by the system.
引用
收藏
页码:5974 / 5984
页数:11
相关论文
共 34 条
[1]   A Versatile and Reproducible Multi-Frequency Electrical Impedance Tomography System [J].
Avery, James ;
Dowrick, Thomas ;
Faulkner, Mayo ;
Goren, Nir ;
Holder, David .
SENSORS, 2017, 17 (02)
[2]   IMAGING SPATIAL DISTRIBUTIONS OF RESISTIVITY USING APPLIED POTENTIAL TOMOGRAPHY [J].
BARBER, CC ;
BROWN, BH ;
FREESTON, IL .
ELECTRONICS LETTERS, 1983, 19 (22) :933-935
[3]   Principles of electrical impedance tomography and its clinical application [J].
Bodenstein, Marc ;
David, Matthias ;
Markstaller, Klaus .
CRITICAL CARE MEDICINE, 2009, 37 (02) :713-724
[4]   Imaging with electricity: Report of the European Concerted Action on Impedance Tomography [J].
Boone, K ;
Barber, D ;
Brown, B .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1997, 21 (06) :201-232
[5]   A WIDE-BAND AC-COUPLED CURRENT SOURCE FOR ELECTRICAL-IMPEDANCE TOMOGRAPHY [J].
BRAGOS, R ;
ROSELL, J ;
RIU, P .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A91-A99
[6]  
Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
[7]  
Dong Xiu-Zhen, 2008, Chinese Journal of Biomedical Engineering, V27, P644
[8]   A comparison of two EIT systems suitable for imaging impedance changes in epilepsy [J].
Fabrizi, L. ;
McEwan, A. ;
Oh, T. ;
Woo, E. J. ;
Holder, D. S. .
PHYSIOLOGICAL MEASUREMENT, 2009, 30 (06) :S103-S120
[9]   An electrode addressing protocol for imaging brain function with electrical impedance tomography using a 16-channel semi-parallel system [J].
Fabrizi, L. ;
McEwan, A. ;
Oh, T. ;
Woo, E. J. ;
Holder, D. S. .
PHYSIOLOGICAL MEASUREMENT, 2009, 30 (06) :S85-S101
[10]  
Ferree TC, 2000, IEEE T BIO-MED ENG, V47, P1584, DOI 10.1109/10.887939