Automated 3D registration of Magnetic Resonance Angiography, 3D Power Doppler, and 3D B-mode ultrasound images of carotid bifurcation

被引:5
作者
Slomka, PJ [1 ]
Mandel, J [1 ]
Fenster, A [1 ]
Downey, D [1 ]
机构
[1] London Hlth Sci Ctr, London, ON N6A 4G5, Canada
来源
MEDICAL IMAGING 2000: IMAGE PROCESSING, PTS 1 AND 2 | 2000年 / 3979卷
关键词
image registration; ultrasound; magnetic resonance angiography;
D O I
10.1117/12.387695
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To allow a more objective interpretation of 3D carotid bifurcation images, we have implemented and evaluated on patient data, automated volume registration of 3D magnetic resonance angiography (MRA), 3D power Doppler (PD) ultrasound, and 3D B-mode ultrasound. Our algorithm maximizes the mutual information between the thresholded intensities of the MRA and PD images. The B-mode images, acquired simultaneously in the same orientation as PD, are registered to the MRA using the transformation obtained from the MRA-PD registration. To test the algorithm we misaligned clinical ultrasound images and simulated mismatches between the datasets due to different appearances of diseased vessels by removing 3D sections of voxels from each of the paired scans. All registrations were assessed visually using integrated 3D volume, surface, and 2D slice display. 97% of images misaligned within a range of 40 degrees and 40 pixels were correctly registered. The deviation from the mean registration parameters due to the simulated defects was 1.6+/-2.5 degrees, 1.5+/-1.6 pixels in X, Y and 0.7+/-0.7 pixels in Z direction. The algorithm can be used to register carotid images with misalignment range of 40 pixels in X, Y directions, 10 pixels in Z direction and 40 degrees rotations, even in the case of different image appearances due to vessel stenoses.
引用
收藏
页码:332 / 341
页数:4
相关论文
共 28 条
[1]   CEREBRAL MR-ANGIOGRAPHY WITH MULTIPLE OVERLAPPING THIN SLAB ACQUISITION .1. QUANTITATIVE-ANALYSIS OF VESSEL VISIBILITY [J].
BLATTER, DD ;
PARKER, DL ;
ROBISON, RO .
RADIOLOGY, 1991, 179 (03) :805-811
[2]   A real vessel phantom for imaging experimentation [J].
Dabrowski, W ;
DunmoreBuyze, J ;
Rankin, RN ;
Holdsworth, DW ;
Fenster, A .
MEDICAL PHYSICS, 1997, 24 (05) :687-693
[3]  
Dey D, 1999, J NUCL MED, V40, P448
[4]   VASCULAR IMAGING WITH A 3-DIMENSIONAL POWER DOPPLER SYSTEM [J].
DOWNEY, DB ;
FENSTER, A .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (03) :665-668
[5]   FURTHER COMMENTS ON THE MEASUREMENT OF CAROTID STENOSIS FROM ANGIOGRAMS [J].
ELIASZIW, M ;
SMITH, RF ;
SINGH, N ;
HOLDSWORTH, DW ;
FOX, AJ ;
BARNETT, HJM .
STROKE, 1994, 25 (12) :2445-2449
[6]  
Fahrig R, 1997, AM J NEURORADIOL, V18, P1507
[7]   Three-dimensional ultrasound imaging of the vasculature [J].
Fenster, A ;
Lee, D ;
Sherebrin, S ;
Rankin, R ;
Downey, D .
ULTRASONICS, 1998, 36 (1-5) :629-633
[8]   Model-based quantitation of 3-D magnetic resonance angiographic images [J].
Frangi, AF ;
Niessen, WJ ;
Hoogeveen, RM ;
van Walsum, T ;
Viergever, MA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (10) :946-956
[9]   A GEOMETRICALLY ACCURATE VASCULAR PHANTOM FOR COMPARATIVE-STUDIES OF X-RAY, ULTRASOUND, AND MAGNETIC-RESONANCE VASCULAR IMAGING - CONSTRUCTION AND GEOMETRICAL VERIFICATION [J].
FRAYNE, R ;
GOWMAN, LM ;
RICKEY, DW ;
HOLDSWORTH, DW ;
PICOT, PA ;
DRANGOVA, M ;
CHU, KC ;
CALDWELL, CB ;
FENSTER, A ;
RUTT, BK .
MEDICAL PHYSICS, 1993, 20 (02) :415-425
[10]  
Gilardi M C, 1996, Q J Nucl Med, V40, P142