A quality assurance tool for high-dose-rate brachytherapy

被引:27
作者
Rickey, Daniel W. [1 ,2 ,3 ]
Sasaki, David [1 ]
Bews, Jeff [1 ,2 ,3 ]
机构
[1] CancerCare Manitoba, Med Phys, Winnipeg, MB R3E 0V9, Canada
[2] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[3] Univ Manitoba, Dept Radiol, Winnipeg, MB R3A 1R9, Canada
关键词
high-dose-rate brachytherapy; quality assurance; radiochromic film; diode detector; RADIOCHROMIC FILM; HDR BRACHYTHERAPY; STEPPING SOURCE; CANCER; TIME;
D O I
10.1118/1.3425786
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work was to develop a quality assurance (QA) tool for high-dose-rate (HDR) brachytherapy that would quickly and easily verify both source positioning (dwell positions) and durations (dwell times). Methods: The authors constructed a QA tool that combined radiochromic film to verify position with four photodiode detectors to verify dwell times. To characterize the temporal accuracy of the tool, a function generator powered four red light-emitting diodes that were optically coupled to the four photodiode detectors. The QA tool was used to verify the dwell positions and times of a commercial brachytherapy afterloader. Measurements of dwell time were independently verified by a one-dimensional optical camera that acquired 1000 lines/s. Results: The temporal accuracy of the QA tool was found to be about 1 ms. For visual assessment, the source position could be located within about 0.5 mm. Evaluating the accuracy and precision of an HDR brachytherapy afterloader, the authors found that the bias in dwell time can exceed 60 ms and the dwell time associated with the first dwell position had an unexpectedly large standard deviation of 30 ms. They found that the source locations were much easier to locate on the film if a plastic catheter was used instead of a metal treatment tube. Scanning the films enabled the dwell positions to be determined within about 0.2 mm. Conclusions: For pretreatment QA, the authors found that this tool allowed verification of dwell positions and dwell times in about 6 min. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3425786]
引用
收藏
页码:2525 / 2532
页数:8
相关论文
共 24 条
[1]   Quality assurance of HDR 192Ir sources using a Fricke dosimeter [J].
Austerlitz, C. ;
Mota, H. ;
Almeida, C. E. ;
Allison, R. ;
Sibata, C. .
MEDICAL PHYSICS, 2007, 34 (04) :1348-1353
[2]   Review of MammoSite brachytherapy: Advantages, disadvantages and clinical outcomes [J].
Bensaleh, Saleh ;
Bezak, Eva ;
Borg, Martin .
ACTA ONCOLOGICA, 2009, 48 (04) :487-494
[3]   Size and positioning reproducibility of an 192Ir brachytherapy stepping source [J].
Berndt, A ;
Rickey, DW ;
Rathee, S ;
Bews, J .
MEDICAL PHYSICS, 2000, 27 (01) :129-131
[4]   QUALITY ASSURANCE TOOL FOR HIGH-DOSE-RATE BRACHYTHERAPY [J].
DEWERD, LA ;
JURSINIC, P ;
KITCHEN, R ;
THOMADSEN, BR .
MEDICAL PHYSICS, 1995, 22 (04) :435-440
[5]   Comparison of dosimetric standards of USA and France for HDR brachytherapy [J].
Douysset, G ;
Gouriou, J ;
Delaunay, F ;
DeWerd, L ;
Stump, K ;
Micka, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (09) :1961-1978
[6]   Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera [J].
Duan, J ;
Macey, DJ ;
Pareek, PN ;
Brezovich, IA .
MEDICAL PHYSICS, 2001, 28 (02) :167-173
[7]   High dose-rate brachytherapy source position quality assurance using radiochromic film [J].
Evans, M. D. C. ;
Devic, S. ;
Podgorsak, E. B. .
MEDICAL DOSIMETRY, 2007, 32 (01) :13-15
[8]   High-dose-rate brachytherapy in the treatment of recurrent and residual head and neck cancer [J].
Glatzel, M ;
Büntzel, J ;
Schröder, D ;
Küttner, K ;
Fröhlich, D .
LARYNGOSCOPE, 2002, 112 (08) :1366-1371
[9]   High dose rate brachytherapy for prostate cancer [J].
Hoskin, P. .
CANCER RADIOTHERAPIE, 2008, 12 (6-7) :512-514
[10]   HDR MICROSELECTRON QUALITY-ASSURANCE STUDIES USING A WELL-TYPE ION-CHAMBER [J].
JONES, CH .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (01) :95-101