Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension

被引:8
|
作者
Coleman, Andrew [1 ]
Steel, Stephen
DeGreef, Annemarie [2 ]
Shennan, Andrew [2 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Dept Med Phys, Rayne Inst, London SE1 7EH, England
[2] Kings Coll London, St Thomas Hosp, Maternal & Fetal Res Unit, London, England
关键词
automated device; blood pressure monitoring; oscillometric measurement; self-measurement; validation studies; DEVICE;
D O I
10.1097/MBP.0b013e32833713cf
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can be confident of their accuracy. Methods The device was validated using the European Society of Hypertension International Protocol. Data collection and analysis were carried out under the laboratory standard ISO17025. Adherence to the standard and protocol was audited on-site by the United Kingdom Accreditation Service. Nine sequential same-arm measurements were taken alternating between trained observers and the test device. Minor variations from the protocol are reported. Results The device passed both phases of the European Society of Hypertension International Protocol. The mean (standard deviation) of the difference between the observers and the device measurements was -1.2 (6.4) mmHg for systolic and 2.1 (6.8) mmHg for diastolic pressures, respectively. Conclusion The BP11 BP monitor can be recommended for home and professional use in an adult population. Blood Press Monit 15: 163-166 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 50 条
  • [41] Validation of the AVITA BPM15S wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Zeng, Wei-Fang
    Zhang, Lu
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 183 - 186
  • [42] Validation of Transtek blood pressure monitor TMB-1491 for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Tian, Huiyong
    Zeng, Sijian
    Zhong, Xiaoyan
    Gong, Wei
    Liu, Wenjun
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 280 - 282
  • [43] Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010
    Michael Abou-Dakn
    Sabine Wenzel
    Journal of Human Hypertension, 2018, 32 : 770 - 774
  • [44] Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, Michael
    Wenzel, Sabine
    JOURNAL OF HUMAN HYPERTENSION, 2018, 32 (11) : 770 - 774
  • [45] Validation of the Kinetik Blood Pressure Monitor-Series 1 for use in adults at home and in clinical settings, according to the 2002 European Society of Hypertension International Protocol on the validation of blood pressure devices
    Schwartz, C. L.
    Edwards, K.
    Gamble, W.
    Kirkham, A.
    Lacy, P.
    Lewis, P.
    McDonagh, S. T. J.
    Peers, C.
    Sheppard, J. P.
    Swales, P.
    Howarth, J.
    Williams, B.
    JOURNAL OF HUMAN HYPERTENSION, 2021, 35 (11) : 1046 - 1050
  • [46] Validation of the Tensoval Duo Control II blood pressure monitor for clinic use and self-measurement according to the British Hypertension Society protocol and the European Society of Hypertension International Protocol revision 2010
    de Greeff, Annemarie
    Shennan, Andrew H.
    BLOOD PRESSURE MONITORING, 2013, 18 (03) : 161 - 166
  • [47] Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension
    Belghazi, Jalil
    El Feghali, Ramzi N.
    Moussalem, Therese
    Rejdych, Maya
    Asmar, Roland G.
    VASCULAR HEALTH AND RISK MANAGEMENT, 2007, 3 (04) : 389 - 400
  • [48] Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension
    Topouchian, Jirar
    Agnoletti, Davide
    Blacher, Jacques
    Youssef, Ahmed
    Ibanez, Isabel
    Khabouth, Jose
    Khawaja, Salwa
    Beaino, Layale
    Asmar, Roland
    VASCULAR HEALTH AND RISK MANAGEMENT, 2011, 7 : 709 - 717
  • [49] Lack of validation of the Dixtal (DX 2020) upper arm blood pressure monitor, in oscillometric mode, for clinical use in an intensive care unit, according to the European Society of Hypertension-International Protocol revision 2010
    Gothardo, Ana C. L. O.
    Savioli, Amanda F.
    Santos, Dayanna S.
    Lamas, Jose L. T.
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 215 - 218
  • [50] Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Chen, Can
    Shang, Fujun
    Wang, Jiepin
    Chen, Jianghong
    Ji, Na
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 253 - 256