Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension

被引:8
|
作者
Coleman, Andrew [1 ]
Steel, Stephen
DeGreef, Annemarie [2 ]
Shennan, Andrew [2 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Dept Med Phys, Rayne Inst, London SE1 7EH, England
[2] Kings Coll London, St Thomas Hosp, Maternal & Fetal Res Unit, London, England
关键词
automated device; blood pressure monitoring; oscillometric measurement; self-measurement; validation studies; DEVICE;
D O I
10.1097/MBP.0b013e32833713cf
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can be confident of their accuracy. Methods The device was validated using the European Society of Hypertension International Protocol. Data collection and analysis were carried out under the laboratory standard ISO17025. Adherence to the standard and protocol was audited on-site by the United Kingdom Accreditation Service. Nine sequential same-arm measurements were taken alternating between trained observers and the test device. Minor variations from the protocol are reported. Results The device passed both phases of the European Society of Hypertension International Protocol. The mean (standard deviation) of the difference between the observers and the device measurements was -1.2 (6.4) mmHg for systolic and 2.1 (6.8) mmHg for diastolic pressures, respectively. Conclusion The BP11 BP monitor can be recommended for home and professional use in an adult population. Blood Press Monit 15: 163-166 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 50 条
  • [31] Validation of the iHealth BP5 wireless upper arm blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Shang, Fujun
    Zhu, Yizheng
    Zhu, Zhenlai
    Liu, Lei
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2013, 18 (05) : 278 - 281
  • [32] Validation of the Beneware model ABP-021 ambulatory blood pressure monitor according to the revised 2010 European Society of hypertension international protocol
    Pereira, Telmo
    Guimaraes, Joao
    BLOOD PRESSURE MONITORING, 2018, 23 (04) : 210 - 213
  • [33] Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002
    Lueders, Stephan
    Krueger, Ralf
    Zemmrich, Claudia
    Forstner, Klaus
    Sturm, Claus-Dieter
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 248 - 252
  • [34] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [35] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 230 - 233
  • [36] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [37] Validation of the Pangao PG-800B11 blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols
    Bing, Sen
    Zhang, Cunhai
    Wang, Liping
    Li, Linyi
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2014, 19 (06) : 366 - 369
  • [38] Validation of the Kinetik Blood Pressure Monitor—Series 1 for use in adults at home and in clinical settings, according to the 2002 European Society of Hypertension International Protocol on the validation of blood pressure devices
    C. L. Schwartz
    K. Edwards
    W. Gamble
    A. Kirkham
    P. Lacy
    P. Lewis
    S. T. J. McDonagh
    C. Peers
    J. P. Sheppard
    P. Swales
    J. Howarth
    B. Williams
    Journal of Human Hypertension, 2021, 35 : 1046 - 1050
  • [39] Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Kollias, Anastasios
    Anagnostopoulos, Ioannis
    Gravvani, Areti
    Stambolliu, Emelina
    Bountzona, Ioanna
    Menti, Ariadni
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2020, 25 (01) : 50 - 52
  • [40] Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices
    Jones, SC
    Bilous, M
    Winship, S
    Finn, P
    Goodwin, J
    BLOOD PRESSURE MONITORING, 2004, 9 (04) : 219 - 223