Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension

被引:8
|
作者
Coleman, Andrew [1 ]
Steel, Stephen
DeGreef, Annemarie [2 ]
Shennan, Andrew [2 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Dept Med Phys, Rayne Inst, London SE1 7EH, England
[2] Kings Coll London, St Thomas Hosp, Maternal & Fetal Res Unit, London, England
关键词
automated device; blood pressure monitoring; oscillometric measurement; self-measurement; validation studies; DEVICE;
D O I
10.1097/MBP.0b013e32833713cf
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can be confident of their accuracy. Methods The device was validated using the European Society of Hypertension International Protocol. Data collection and analysis were carried out under the laboratory standard ISO17025. Adherence to the standard and protocol was audited on-site by the United Kingdom Accreditation Service. Nine sequential same-arm measurements were taken alternating between trained observers and the test device. Minor variations from the protocol are reported. Results The device passed both phases of the European Society of Hypertension International Protocol. The mean (standard deviation) of the difference between the observers and the device measurements was -1.2 (6.4) mmHg for systolic and 2.1 (6.8) mmHg for diastolic pressures, respectively. Conclusion The BP11 BP monitor can be recommended for home and professional use in an adult population. Blood Press Monit 15: 163-166 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 50 条
  • [1] Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (06) : 329 - 331
  • [2] Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension protocol
    Longo, D
    Toffanin, G
    Garbelotto, R
    Zaetta, V
    Businaro, L
    Palatini, P
    BLOOD PRESSURE MONITORING, 2003, 8 (02) : 91 - 95
  • [3] Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (01) : 42 - 44
  • [4] Validation of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International protocol for validation of blood pressure measuring devices in adults
    Mallion, JJM
    Pierre, H
    Neuder, Y
    Ormezzano, O
    Baguet, JP
    BLOOD PRESSURE MONITORING, 2005, 10 (02) : 97 - 101
  • [5] Self-measurement of blood pressure: validation of the Braun BP VitalScan Plus 1650 monitor according to the European Society of Hypertension International Protocol
    Nolly, Hector
    Osso, Pablo
    Nolly, Alejandro
    Silva, Martin
    Nolly, Mariela
    Romero, Marcelo
    BLOOD PRESSURE MONITORING, 2008, 13 (02) : 101 - 106
  • [6] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beime, Beate
    Deutsch, Cornelia
    Krueger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (05) : 573 - 580
  • [7] Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the international protocol of the european society of hypertension
    Pini, C.
    Pastori, M.
    Baccheschi, J.
    Gerosa, P.
    Omboni, S.
    Parati, G.
    JOURNAL OF HYPERTENSION, 2007, 25 : S39 - S39
  • [8] Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the international protocol of the european society of hypertension
    Pini, C.
    Pastori, M.
    Natalizi, A.
    Gerosa, P.
    Omboni, S.
    Parati, G.
    JOURNAL OF HYPERTENSION, 2007, 25 : S39 - S39
  • [9] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 573 - 580
  • [10] Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Natalizi, Anna
    Gerosa, Pietro Francesco
    Frigerioa, Mauro
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2008, 13 (03) : 177 - 182