Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension

被引:8
作者
Coleman, Andrew [1 ]
Steel, Stephen
DeGreef, Annemarie [2 ]
Shennan, Andrew [2 ]
机构
[1] Guys & St Thomas NHS Fdn Trust, Dept Med Phys, Rayne Inst, London SE1 7EH, England
[2] Kings Coll London, St Thomas Hosp, Maternal & Fetal Res Unit, London, England
关键词
automated device; blood pressure monitoring; oscillometric measurement; self-measurement; validation studies; DEVICE;
D O I
10.1097/MBP.0b013e32833713cf
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The BP11 is currently one of the lowest-cost blood pressure (BP) monitors available in the UK and has sold approximately 1.5 million units since its launch in 2006. As access to home-use BP monitoring increases, it is important that such devices are validated so that consumers and clinicians can be confident of their accuracy. Methods The device was validated using the European Society of Hypertension International Protocol. Data collection and analysis were carried out under the laboratory standard ISO17025. Adherence to the standard and protocol was audited on-site by the United Kingdom Accreditation Service. Nine sequential same-arm measurements were taken alternating between trained observers and the test device. Minor variations from the protocol are reported. Results The device passed both phases of the European Society of Hypertension International Protocol. The mean (standard deviation) of the difference between the observers and the device measurements was -1.2 (6.4) mmHg for systolic and 2.1 (6.8) mmHg for diastolic pressures, respectively. Conclusion The BP11 BP monitor can be recommended for home and professional use in an adult population. Blood Press Monit 15: 163-166 (C) 2010 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 7 条
[1]  
*ASS ADV MED INSTR, 2002, ANSI AAMI
[2]   Validation of A&D UA-85X device for blood pressure measurement [J].
Bonso, Elisa ;
Ragazzo, Fabio ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2008, 13 (06) :339-341
[3]  
*BS, EN106011996 BS
[4]  
*BS, 170252000 BS EN ISOI
[5]   Validation of the visocor HM40 wrist blood pressure measuring device according to the International Protocol [J].
Dorigatti, Francesca ;
Bonso, Elisa ;
Saladini, Francesca ;
Palatini, Paolo .
BLOOD PRESSURE MONITORING, 2009, 14 (02) :83-86
[6]   The Omron Elite 7300W home blood pressure monitor passes the European Society of Hypertension International Validation Protocol for women and men [J].
Grim, Clarence E. ;
Grim, Carlene M. .
BLOOD PRESSURE MONITORING, 2009, 14 (02) :87-90
[7]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17