Interference of dextran in biuret-type assays of serum proteins

被引:4
|
作者
Delanghe, JR
Hamers, N
Taes, YE
Libeer, JC
机构
[1] State Univ Ghent Hosp, Clin Chem Lab, B-9000 Ghent, Belgium
[2] IPH, Inst Publ Hlth, Brussels, Belgium
关键词
biuret; dextran interference; plasma expanders; total protein assay;
D O I
10.1515/CCLM.2005.011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Dextran interference in biuret-type assays of total serum proteins was investigated in a Belgian National External Quality Assurance Survey with 256 participants. In vitro supplementation of therapeutic (10 % Gentran 70(R)) dextran concentrations showed a broadly varying (from 0 to 20 %) negative interference. The analytical interference was found to depend on both the sodium hydroxide and tartrate concentrations in the reagent formulation. The dry chemistry biuret method was not affected by the dextran interference. In a number of cases, the effects observed may be of clinical importance. Both clinicians and laboratory staff should be aware of the persistence of this analytical problem.