Synthetic single crystal diamond dosimeters for conformal radiation therapy application

被引:10
作者
Almaviva, S. [1 ]
Ciancaglioni, I. [1 ]
Consorti, R. [2 ]
De Notaristefani, F. [3 ]
Manfredotti, C. [4 ,5 ]
Marinelli, Marco [1 ]
Milani, E. [1 ]
Petrucci, A. [2 ]
Prestopino, G. [1 ]
Verona, C. [1 ]
Verona-Rinati, G. [1 ]
机构
[1] Univ Roma Tor Vergata, Ist Nazl Fis Nucl, Dipartimento Ingn Meccan, I-00133 Rome, Italy
[2] Osped San Filippo Neri, UO Fis Sanit, I-00135 Rome, Italy
[3] Univ Roma 3, Ist Nazl Fis Nucl, Dipartimento Ingn Elettron, I-00146 Rome, Italy
[4] Univ Turin, Expt Phys Dept, I-10125 Turin, Italy
[5] Univ Turin, Ctr Excellence NIS, I-10125 Turin, Italy
关键词
CVD Diamond Detector; Single Crystal Diamond; Solid State Dosimetry; IMRT Dosimetry; RADIOTHERAPY; PHOTON;
D O I
10.1016/j.diamond.2009.10.007
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A synthetic single crystal diamond based dosimeter in a p-type/intrinsic/nietal structure, operating in photovoltaic regime, is proposed for application in highly conformed radiotherapy dosimetry. The device was characterized by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV, obtained by a CLINAC DHX Varian accelerator. All measurements were performed in a water phantom and commercial ionization chambers were used for calibration and comparison. Results showed a very good agreement of the diamond device response, as compared with the reference dosimeters, fast response times and high spatial resolution. One of such diamond dosimeters was then tested using a real Intensity Modulated Radiation Therapy (IMRT) prostate cancer treatment plan and its performance was compared with the ones from ionization chambers and a 2D diode array. The obtained results clearly assess the suitability of synthetic single crystal diamond for dose measurements in highly conformed radiotherapy and particularly in IMRT applications. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:217 / 220
页数:4
相关论文
共 16 条
[1]   Synthetic single crystal diamond diodes for radiotherapy dosimetry [J].
Almaviva, S. ;
Marinelli, Marco ;
Milani, E. ;
Tucciarone, A. ;
Verona-Rinati, G. ;
Consorti, R. ;
Petrucci, A. ;
De Notaristefani, F. ;
Ciancaglioni, I. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 594 (02) :273-277
[2]   Synthetic single crystal diamond dosimeters for Intensity Modulated Radiation Therapy applications [J].
Almaviva, S. ;
Ciancaglioni, I. ;
Consorti, R. ;
De Notaristefani, F. ;
Manfredotti, C. ;
Marinelli, Marco ;
Milani, E. ;
Petrucci, A. ;
Prestopino, G. ;
Verona, C. ;
Verona-Rinati, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (01) :191-194
[3]   Intensity-modulated radiotherapy: Current status and issues of interest [J].
Boyer, AL ;
Butler, EB ;
DiPetrillo, TA ;
Engler, MJ ;
Fraass, B ;
Grant, W ;
Ling, CC ;
Low, DA ;
Mackie, TR ;
Mohan, R ;
Purdy, JA ;
Roach, M ;
Rosenman, JG ;
Verhey, LJ ;
Wong, JW ;
Cumberlin, RL ;
Stone, H ;
Palta, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (04) :880-914
[4]   Diamond dosimetry: Outcomes of the CANDIDO and CONRAD INFN projects [J].
Bucciolini, M ;
Borchi, E ;
Bruzzi, M ;
Casati, M ;
Cirrone, P ;
Cuttone, G ;
De Angelis, C ;
Lovik, I ;
Onori, S ;
Raffaele, L ;
Sciortino, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 552 (1-2) :189-196
[5]   An investigation of the operating characteristics of two PTW diamond detectors in photon and electron beams [J].
De Angelis, C ;
Onori, S ;
Pacilio, M ;
Cirrone, GAP ;
Cuttone, G ;
Raffaele, L ;
Bucciolini, M ;
Mazzocchi, S .
MEDICAL PHYSICS, 2002, 29 (02) :248-254
[6]   Performances of homoepitaxial single crystal diamond in diagnostic x-ray dosimetry [J].
Garino, Y ;
Lo Giudice, A ;
Manfredotti, C ;
Marinelli, M ;
Milani, E ;
Tucciarone, A ;
Verona-Rinati, G .
APPLIED PHYSICS LETTERS, 2006, 88 (15)
[7]   Recent improvements on the use of CVD diamond ionisation chambers for radiotherapy applications [J].
Guerrero, M. J. ;
Tromson, D. ;
Descamps, C. ;
Bergonzo, P. .
DIAMOND AND RELATED MATERIALS, 2006, 15 (4-8) :811-814
[8]   Requirements for synthetic diamond devices for radiotherapy dosimetry applications [J].
Guerrero, MJ ;
Tromson, D ;
Rebisz, M ;
Mer, C ;
Bazin, B ;
Bergonzo, P .
DIAMOND AND RELATED MATERIALS, 2004, 13 (11-12) :2046-2051
[9]   DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM [J].
HOBAN, PW ;
HEYDARIAN, M ;
BECKHAM, WA ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) :1219-1229
[10]   A study of radiation dosimeters based on synthetic HPHT diamond [J].
Marczewska, B. ;
Kupriyanov, I. ;
Pal'yanov, Yu. ;
Nowak, T. ;
Olko, P. ;
Rebisz, M. ;
Waligorski, M. P. R. .
DIAMOND AND RELATED MATERIALS, 2007, 16 (02) :191-195