Intraoperative fluoroscopy, portable X-ray, and CT: patient and operating room personnel radiation exposure in spinal surgery

被引:76
作者
Nelson, Elisha M. [1 ]
Monazzam, Shafagh M. [2 ]
Kim, Kee D. [3 ]
Seibert, J. Anthony [1 ]
Klineberg, Eric O. [2 ]
机构
[1] Univ Calif Davis, Dept Radiol, Sacramento, CA 95817 USA
[2] Univ Calif Davis, Dept Orthopaed, Sacramento, CA 95817 USA
[3] Univ Calif Davis, Dept Neurosurg, Sacramento, CA 95817 USA
关键词
Radiation; Intraoperative imaging; O-arm; C-arm; Portable X-ray; Fluoroscopy; PEDICLE SCREW INSERTION; IONIZING-RADIATION; ORTHOPEDIC SURGEONS; COMPUTED-TOMOGRAPHY; IMAGE GUIDANCE; PLACEMENT; ACCURACY; CANCER; RISK; COMPLICATIONS;
D O I
10.1016/j.spinee.2014.06.003
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND CONTEXT: Intraoperative imaging is essential in spinal surgery to both determine the correct level and place implants safely. Surgeons have a variety of options: C-arm fluoroscopy (C-arm), portable X-ray (XR) radiography, and portable cone-beam computed tomography (O-arm). Although these modalities have their respective advantages and disadvantages, direct comparison of radiation exposure to either the patient or the operating room (OR) staff has not been made. PURPOSE: To determine the amount of radiation exposure to patients and OR staff during spine surgery with C-arm, XR, and O-arm. STUDY DESIGN: An experimental model to assess radiation exposure to OR staff and phantom patient during spine surgery. METHODS: A plastic phantom was created to emulate patient volume and absorption scattering characteristics of a typical sized adult abdominal volume. Radiation exposure was measured with ion chamber dosimeters to determine entrance phantom and scatter exposures at common positions occupied by OR staff for C-arm, XR, and O-arm in typical image acquisition during spinal surgery. RESULTS: Single lateral (LAT)/posterior-anterior entrance patient radiation exposure for C-arm was on average 116/102 mR, single-exposure XR for LAT/anterior-posterior (AP) was 3,435/2,160 mR, and single-exposure O-arm for LAT/AP was 4,360/5,220 mR. O-arm surface exposure LAT/AP was equivalent to 38/41 C-arm and 1.5/2.4 XR exposures. The surgeon and surgeon assistant had higher levels of scatter radiation for C-arm, followed by O-arm and XR. For the LAT C-arm acquisition, a 7.7-fold increase in radiation exposure was measured on the X-ray tube side compared with the detector side. The anesthesiologist scatter radiation level for a single acquisition was highest for O-arm, followed by XR and C-arm. The radiologic technologist scatter radiation level was highest for XR, followed by O-arm and fluoroscopy. Overall radiation exposure to OR staff was less than 4.4 mR for a single acquisition in all modalities. CONCLUSIONS: Assessment of radiation risk to the patient and OR staff should be part of the decision for utilization of any specific imaging modality during spinal surgery. This study provides the surgeon with information to better weigh the risks and benefits of each imaging modality. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:2985 / 2991
页数:7
相关论文
共 41 条
  • [1] A prospective evaluation of the role for intraoperative x-ray in lumbar discectomy. Predictors of incorrect level exposure
    Ammerman, Joshua M.
    Ammerman, Matthew D.
    Dambrosia, James
    Ammerman, Bruce J.
    [J]. SURGICAL NEUROLOGY, 2006, 66 (05): : 470 - 474
  • [2] Use of CT-Based Intraoperative Spinal Navigation: Management of Radiation Exposure to Operator, Staff, and Patients
    Bandela, John R.
    Jacob, R. Patrick
    Arreola, Manuel
    Griglock, Thomas M.
    Bova, Frank
    Yang, Mu
    [J]. WORLD NEUROSURGERY, 2013, 79 (02) : 390 - 394
  • [3] MICROSURGERY VERSUS STANDARD REMOVAL OF THE HERNIATED LUMBAR-DISK - A 3-YEAR COMPARISON IN 150 CASES
    BARRIOS, C
    AHMED, M
    ARROTEGUI, J
    BJORNSSON, A
    GILLSTROM, P
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1990, 61 (05): : 399 - 403
  • [4] How accurately do novice surgeons place thoracic pedicle screws with the free hand technique?
    Bergeson, Ryan K.
    Schwend, Richard M.
    DeLucia, Tracey
    Silva, Selina R.
    Smith, Jason E.
    Avilucea, Frank R.
    [J]. SPINE, 2008, 33 (15) : E501 - E507
  • [5] Surgeon and patient radiation exposure in minimally invasive transforaminal lumbar interbody fusion
    Bindal, Rajesh K.
    Glaze, Sharon
    Ognoskie, Meghann
    Tunner, Van
    Malone, Robert
    Ghosh, Subrata
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2008, 9 (06) : 570 - 573
  • [6] Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know
    Brenner, DJ
    Doll, R
    Goodhead, DT
    Hall, EJ
    Land, CE
    Little, JB
    Lubin, JH
    Preston, DL
    Preston, RJ
    Puskin, JS
    Ron, E
    Sachs, RK
    Samet, JM
    Setlow, RB
    Zaider, M
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) : 13761 - 13766
  • [7] Complications of pediatric thoracolumbar and lumbar pedicle screws
    Brown, CA
    Lenke, LG
    Bridwell, KH
    Geideman, WM
    Hasan, SA
    Blanke, K
    [J]. SPINE, 1998, 23 (14) : 1566 - 1571
  • [8] Prevalence of Cancer in Female Orthopedic Surgeons in the United States
    Chou, Loretta B.
    Cox, Christopher A.
    Tung, Joanna J.
    Harris, Alex H. S.
    Brooks-Terrell, Daria
    Sieh, Weiva
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (01) : 240 - 244
  • [9] DO SURGICAL RESIDENTS, EMERGENCY-MEDICINE PHYSICIANS, AND NURSES EXPERIENCE SIGNIFICANT RADIATION EXPOSURE DURING THE RESUSCITATION OF TRAUMA PATIENTS
    CIRAULO, DL
    MARINI, CP
    LLOYD, GT
    FISHER, J
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1994, 36 (05) : 703 - 705
  • [10] COMPLICATIONS ASSOCIATED WITH THE TECHNIQUE OF PEDICLE SCREW FIXATION - A SELECTED SURVEY OF ABS MEMBERS
    ESSES, SI
    SACHS, BL
    DREYZIN, V
    [J]. SPINE, 1993, 18 (15) : 2231 - 2239