Motion artifact reducing reconstruction of 4D CT image data for the analysis of respiratory dynamics

被引:30
作者
Werner, R.
Ehrhardt, J.
Frenzel, T.
Saering, D.
Lu, W.
Low, D.
Handels, H.
机构
[1] Univ Hamburg, Med Ctr Hamburg Eppendorf, Dept Med Informat, D-20246 Hamburg, Germany
[2] AK St Georg Hosp, Hermann Holthusen Inst Radiotherapy, Hamburg, Germany
[3] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO USA
关键词
computed X-ray tomography; 4D CT; computer-assisted image analysis; respiratory motion; lung cancer;
D O I
10.1160/ME9040
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Objectives: Respiratory motion represents a major problem in radiotherapy of thoracic and abdominal tumors. Methods for compensation require comprehensive knowledge of underlying dynamics. Therefore, 4D (= 3D + t) CT data can be helpful. But modern CT scanners cannot scan a large region of interest simultaneously. So patients hove to be scanned in segments. Commonly used approaches for reconstructing the data segments into 4D CT images cause motion artifacts. In order to reduce the artifacts, a new method for 4D CT reconstruction is presented. The resulting data sets are used to analyze respiratory motion. Methods. Spatiotemporal CT image sequences of lung cancer patients were acquired using a multi-slice CT in cine mode during free breathing. 4D CT reconstruction was done by optical flow based temporal interpolation. The resulting 4D image data were compared with data generated by the commonly used nearest neighbor reconstruction. Subsequent motion analysis is mainly concerned with tumor mobility. Results: The presented optical flow-based method enables the reconstruction of 3D CT images at arbitrarily chosen points of the patient's breathing cycle. A considerable reduction of motion artifacts has been proven in eight patient data sets. Motion analysis showed that tumor mobility differs strongly between the patients. Conclusions. Due to the proved reduction of motion artifacts, the optical flow-based 4D CT reconstruction offers the possibility of high-quality motion analysis. Because the method is based on an interpolation scheme, it additionally has the potential to enable the reconstruction of 4D CT data from a lesser number of scans.
引用
收藏
页码:254 / 260
页数:7
相关论文
共 50 条
[41]   A multiple points method for 4D CT image sorting [J].
Gianoli, Chiara ;
Riboldi, Marco ;
Spadea, Maria Francesca ;
Travaini, Laura Lavinia ;
Ferrari, Mahila ;
Mei, Riccardo ;
Orecchia, Roberto ;
Baroni, Guido .
MEDICAL PHYSICS, 2011, 38 (02) :656-667
[42]   A continuous 4D motion model from multiple respiratory cycles for use in lung radiotherapy [J].
McClelland, Jamie R. ;
Blackall, Jane M. ;
Tarte, Segolene ;
Chandler, Adam C. ;
Hughes, Simon ;
Ahmad, Shahreen ;
Landau, David B. ;
Hawkes, David J. .
MEDICAL PHYSICS, 2006, 33 (09) :3348-3358
[43]   Tumor and normal tissue motion in the thorax during respiration: Analysis of volumetric and positional variations using 4D CT [J].
Weiss, Elisabeth ;
Wijesooriya, Krishni ;
Dill, S. Vaughn ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (01) :296-307
[44]   Direction-Dependent Regularization for Improved Estimation of Liver and Lung Motion in 4D Image Data [J].
Schmidt-Richberg, Alexander ;
Ehrhardt, Jan ;
Werner, Rene ;
Handels, Heinz .
MEDICAL IMAGING 2010: IMAGE PROCESSING, 2010, 7623
[45]   4D medical image computing and visualization of lung tumor mobility in spatio-temporal CT image data [J].
Handels, Heinz ;
Werner, Rene ;
Schmidt, Rainer ;
Frenzel, Thorsten ;
Lu, Wei ;
Low, Daniel ;
Ehrhardt, Jan .
INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2007, 76 (SUPPL. 3) :S433-S439
[46]   Tensor grid based image registration with application to ventilation estimation on 4D CT lung data [J].
Ruppertshofen, Heike ;
Kabus, Sven ;
Fischer, Bernd .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2010, 5 (06) :583-593
[47]   Tensor grid based image registration with application to ventilation estimation on 4D CT lung data [J].
Heike Ruppertshofen ;
Sven Kabus ;
Bernd Fischer .
International Journal of Computer Assisted Radiology and Surgery, 2010, 5 :583-593
[48]   Inter-Subject Variability of Respiratory Motion from 4D MRI [J].
Abd Rahni, Ashrani Aizzuddin ;
Lewis, Emma ;
Wells, Kevin .
2015 INTERNATIONAL CONFERENCE ON BIOSIGNAL ANALYSIS, PROCESSING AND SYSTEMS (ICBAPS), 2015,
[49]   Resolving Variable Respiratory Motion From Unsorted 4D Computed Tomography [J].
Huang, Yuliang ;
Eiben, Bjoern ;
Thielemans, Kris ;
McClelland, Jamie R. .
MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION - MICCAI 2024, PT I, 2024, 15001 :588-597
[50]   Motion artifacts occurring at the lung/diaphragm interface using 4D CT attenuation correction of 4D PET scans [J].
Killoran, Joseph H. ;
Gerbaudo, Victor H. ;
Mamede, Marcelo ;
Ionascu, Dan ;
Park, Sang-June ;
Berbeco, Ross .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (04) :261-274