Technical Note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry

被引:46
作者
Desroches, Joannie [1 ]
Bouchard, Hugo [1 ,2 ]
Lacroix, Frederic [1 ,2 ]
机构
[1] CHUM, Dept Radiooncol, Montreal, PQ H2L 4M1, Canada
[2] Univ Montreal, Dept Phys, Montreal, PQ H3T 1J4, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
biomedical optical imaging; calibration; dosimetry; error statistics; radiation therapy; RADIOCHROMIC FILM; ENERGY-DEPENDENCE; UNCERTAINTY ANALYSIS; DOSE VERIFICATION; THERAPY; SENSITIVITY; COMMITTEE; AAPM; MVP;
D O I
10.1118/1.3355895
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study is to determine the effect on the measured optical density of scanning on either side of a Gafchromic EBT and EBT2 film using an Epson (Epson Canada Ltd., Toronto, Ontario) 10000XL flat bed scanner. Methods: Calibration curves were constructed using EBT2 film scanned in landscape orientation in both reflection and transmission mode on an Epson 10000XL scanner. Calibration curves were also constructed using EBT film. Potential errors due to an optical density difference from scanning the film on either side ("face up" or "face down") were simulated. Results: Scanning the film face up or face down on the scanner bed while keeping the film angular orientation constant affects the measured optical density when scanning in reflection mode. In contrast, no statistically significant effect was seen when scanning in transmission mode. This effect can significantly affect relative and absolute dose measurements. As an application example, the authors demonstrate potential errors of 17.8% by inverting the film scanning side on the gamma index for 3%-3 mm criteria on a head and neck intensity modulated radiotherapy plan, and errors in absolute dose measurements ranging from 10% to 35% between 2 and 5 Gy. Conclusions: Process consistency is the key to obtaining accurate and precise results in Gafchromic film dosimetry. When scanning in reflection mode, care must be taken to place the film consistently on the same side on the scanner bed.
引用
收藏
页码:1565 / 1570
页数:6
相关论文
共 25 条
[1]   On the characterization and uncertainty analysis of radiochromic film dosimetry [J].
Bouchard, Hugo ;
Lacroix, Frederic ;
Beaudoin, Gilles ;
Carrier, Jean-Francois ;
Kawrakow, Iwan .
MEDICAL PHYSICS, 2009, 36 (06) :1931-1946
[2]   Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) :60-62
[3]   Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack [J].
Chiu-Tsao, Sou-Tung ;
Chan, Maria F. .
MEDICAL PHYSICS, 2009, 36 (06) :2074-2083
[4]   Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies [J].
Chiu-Tsao, ST ;
Ho, Y ;
Shankar, R ;
Wang, L ;
Harrison, LB .
MEDICAL PHYSICS, 2005, 32 (11) :3350-3354
[5]   Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Watts, Ronald J. ;
Ahnesjo, Anders ;
Gibbons, John ;
Li, X. Allen ;
Lowenstein, Jessica ;
Mitra, Raj K. ;
Simon, William E. ;
Zhu, Timothy C. .
MEDICAL PHYSICS, 2008, 35 (09) :4186-4215
[6]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[7]   Dosimetric properties of improved GafChromic films for seven different digitizers [J].
Devic, S ;
Seuntjens, J ;
Hegyi, G ;
Podgorsak, EB ;
Soares, CG ;
Kirov, AS ;
Ali, I ;
Williamson, JF ;
Elizondo, A .
MEDICAL PHYSICS, 2004, 31 (09) :2392-2401
[8]   Absorption spectroscopy of EBT model GAFCHROMIC™ film [J].
Devic, Slobodan ;
Tomic, Nada ;
Pang, Zhiyu ;
Seuntjens, Jan ;
Podgorsak, Ervin B. ;
Soares, Christopher G. .
MEDICAL PHYSICS, 2007, 34 (01) :112-118
[9]  
*EBT2, INT SPEC PROD
[10]   An assessment of GafChromic film for measuring 50 kV and 100 kV percentage depth dose curves [J].
Fletcher, Claire Lesley ;
Mills, John A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11) :N209-N218