Reconfigurable MRI coil technology can substantially reduce RF heating of deep brain stimulation implants: First in-vitro study of RF heating reduction in bilateral DBS leads at 1.5 T

被引:42
作者
Golestanirad, Laleh [1 ,2 ]
Kazemivalipour, Ehsan [3 ]
Keil, Boris [4 ]
Downs, Sean [5 ]
Kirsch, John [5 ]
Elahi, Behzad [6 ]
Pilitsis, Julie [7 ]
Wald, Lawrence L. [5 ]
机构
[1] Northwestern Univ, Dept Biomed Engn, McCormick Sch Engn, Evanston, IL 60208 USA
[2] Northwestern Univ, Feinberg Sch Med, Dept Radiol, Chicago, IL 60611 USA
[3] Bilkent Univ, Dept Elect & Elect Engn, Ankara, Turkey
[4] Inst Med Phys & Radiat Protect, Dept Life Sci Engn, Giessen, Germany
[5] Massachusetts Gen Hosp, AA Martinos Ctr Biomed Imaging, Boston, MA 02114 USA
[6] Bryan Hlth, Dept Neurol, Lincoln, NE USA
[7] Albany Med Ctr, Dept Neurosurg, Albany, NY USA
关键词
CARDIAC-PACEMAKERS; SAFETY; WIRES; TESLA; SAR;
D O I
10.1371/journal.pone.0220043
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Patients with deep brain stimulation (DBS) implants can significantly benefit from magnetic resonance imaging (MRI), however access to MRI is restricted in these patients because of safety concerns due to RF heating of the leads. Recently we introduced a patient-adjustable reconfigurable transmit coil for low-SAR imaging of DBS at 1.5T. A previous simulation study demonstrated a substantial reduction in the local SAR around single DBS leads in 9 unilateral lead models. This work reports the first experimental results of temperature measurement at the tips of bilateral DBS leads with realistic trajectories extracted from postoperative CT images of 10 patients (20 leads in total). A total of 200 measurements were performed to record temperature rise at the tips of the leads during 2 minutes of scanning with the coil rotated to cover all accessible rotation angles. In all patients, we were able to find an optimum coil rotation angle and reduced the heating of both left and right leads to a level below the heating produced by the body coil. An average heat reduction of 65% was achieved for bilateral leads. When considering each lead alone, an average heat reduction of 80% was achieved. Our results suggest that reconfigurable coil technology introduces a promising approach for imaging of patients with DBS implants.
引用
收藏
页数:17
相关论文
共 45 条
[1]   Deep brain stimulation in Tourette's syndrome [J].
Ackermans, Linda ;
Temel, Yasin ;
Visser-Vandewalle, Veerle .
NEUROTHERAPEUTICS, 2008, 5 (02) :339-344
[2]   REDUCTION OF MAGNETIC RESONANCE IMAGING-RELATED HEATING IN DEEP BRAIN STIMULATION LEADS USING A LEAD MANAGEMENT DEVICE [J].
Baker, Kenneth B. ;
Tkach, Jean ;
Hall, John D. ;
Nyenhuis, John A. ;
Shellock, Frank G. ;
Rezai, Ali R. .
NEUROSURGERY, 2005, 57 (04) :392-396
[3]   In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: Role of implant geometry [J].
Calcagnini, Giovanni ;
Triventi, Michele ;
Censi, Federica ;
Mattei, Eugenio ;
Bartolini, Pietro ;
Kainz, Wolfgang ;
Bassen, Howard I. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (04) :879-886
[4]   Clinical Improvement in a Treatment-Resistant Patient With Schizophrenia Treated With Deep Brain Stimulation [J].
Corripio, Iluminada ;
Sarro, Salvador ;
McKenna, Peter J. ;
Molet, Joan ;
Alvarez, Enric ;
Pomarol-Clotet, Edith ;
Portella, Maria J. .
BIOLOGICAL PSYCHIATRY, 2016, 80 (08) :E69-E70
[5]   Bio-heat transfer model of deep brain stimulation-induced temperature changes [J].
Elwassif, Maged M. ;
Kong, Qingjun ;
Vazquez, Maribel ;
Bikson, Marom .
JOURNAL OF NEURAL ENGINEERING, 2006, 3 (04) :306-315
[6]  
Eryaman Y., 2012, MAGNETIC RESONANCE M
[7]   A simple geometric analysis method for measuring and mitigating RF induced currents on Deep Brain Stimulation leads by multichannel transmission/reception [J].
Eryaman, Yigitcan ;
Kobayashi, Naoharu ;
Moen, Sean ;
Aman, Joshua ;
Grant, Andrea ;
Vaughan, J. Thomas ;
Molnar, Gregory ;
Park, Michael C. ;
Vitek, Jerrold ;
Adriany, Gregor ;
Ugurbil, Kamil ;
Harel, Noam .
NEUROIMAGE, 2019, 184 :658-668
[8]   Parallel Transmit Pulse Design for Patients with Deep Brain Stimulation Implants [J].
Eryaman, Yigitcan ;
Guerin, Bastien ;
Akgun, Can ;
Herraiz, Joaquin L. ;
Martin, Adrian ;
Torrado-Carvajal, Angel ;
Malpica, Norberto ;
Hernandez-Tamames, Juan A. ;
Schiavi, Emanuele ;
Adalsteinsson, Elfar ;
Wald, Lawrence L. .
MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (05) :1896-1903
[9]   Reduction of Implant RF Heating Through Modification of Transmit Coil Electric Field [J].
Eryaman, Yigitcan ;
Akin, Burak ;
Atalar, Ergin .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (05) :1305-1313
[10]  
Finelli DA, 2002, AM J NEURORADIOL, V23, P1795