Quantification of blood clotting kinetics II: Thromboelastograph analysis and measurement of erythrocyte sedimentation rate using magnetoelastic sensors

被引:18
|
作者
Roy, Somnath C.
Ong, Keat G.
Zeng, Kefeng
Grimes, Craig A. [1 ]
机构
[1] Penn State Univ, Mat Res Inst, Dept Elect Engn, Mat Res Lab 217, University Pk, PA 16802 USA
[2] Michigan Technol Univ, Dept Biomed Engn, Houghton, MI 49931 USA
[3] KMG2 Sensors Corp, State Coll, PA 16803 USA
关键词
thromboelastograph (TEG); erythrocyte sedimentation rate (ESR); blood clotting kinetics; clot; magnetoelastic;
D O I
10.1166/sl.2007.218
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report on the development of a low cost, versatile diagnostic system capable of simultaneously performing a thromboelastograph (TEG) analysis and measuring the erythrocyte sedimentation rate (ESR) for the detection of hemophilia, von Willebrand disease, polymyalgia rheumatica, temporal arteritis, and anemia. This diagnostic system is based on use of magnetoelastic sensors that, when excited by a magnetic field, resonate thereby generating a secondary magnetic detectable by a remotely located magnetic coil. By measuring changes in the resonance frequency and resonance amplitude of the sensor both liquid viscosity, and mass loading of the sensor, can be accurately determined. This allows the sensor to monitor real-time blood clotting kinetics as well as clot strength, hence perform a thromboelastograph analysis, as well as determine ESR by measuring mass loading of the sensor as associated with particle settling.
引用
收藏
页码:432 / 440
页数:9
相关论文
共 27 条
  • [1] Quantification of blood clotting kinetics I: Determination of activated clotting times as a function of heparin concentration using magnetoelastic sensors
    Zeng, Kefeng
    Roy, Somnath C.
    Grimes, Craig A.
    SENSOR LETTERS, 2007, 5 (02) : 425 - 431
  • [2] Remote query analysis of blood clotting utilizing magnetoelastic sensors.
    Issa, M
    Grimes, CA
    Bachas, LG
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2000, 220 : U95 - U95
  • [4] A microfluidic erythrocyte sedimentation rate analyzer using rouleaux formation kinetics
    Ziya Isiksacan
    Mohammad Asghari
    Caglar Elbuken
    Microfluidics and Nanofluidics, 2017, 21
  • [5] A microfluidic erythrocyte sedimentation rate analyzer using rouleaux formation kinetics
    Isiksacan, Ziya
    Asghari, Mohammad
    Elbuken, Caglar
    MICROFLUIDICS AND NANOFLUIDICS, 2017, 21 (03)
  • [6] Acoustic measurement of erythrocyte sedimentation rate and other parameters of blood without reagents
    Tamarova, E
    Dolgov, V
    Klemin, V
    CLINICA CHIMICA ACTA, 2005, 355 : S297 - S297
  • [7] Erythrocyte Sedimentation Rate Measurement Using as a Rapid Alternative to the Westergren Method
    Hashemi, Reza
    Majidi, Alireza
    Motamed, Hassan
    Amini, Afshin
    Najari, Fares
    Tabatabaey, Ali
    EMERGENCY, 2015, 3 (02): : 50 - 53
  • [8] STORAGE OF BLOOD-SAMPLES FOR MEASUREMENT OF PLASMA VISCOSITY AND ERYTHROCYTE SEDIMENTATION-RATE
    CASWELL, M
    STUART, J
    CLINICAL HEMORHEOLOGY, 1992, 12 (02): : 309 - 315
  • [9] EDTA samples for the measurement of erythrocyte sedimentation rate using Sedimatic 100 analyser
    Horsti, J
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2001, 39 (10) : 1001 - 1003
  • [10] Erythrocyte sedimentation rate (ESR) and blood clotting time (BCT) in crossbred (Hariana x Friesian) preparturient cows and its calves
    Prabha, B
    Singh, C
    INDIAN JOURNAL OF ANIMAL SCIENCES, 2000, 70 (06): : 594 - 595