Electropolymerized hydrophobic polyazulene as solid-contacts in potassium-selective electrodes

被引:44
作者
He, Ning [1 ]
Gyurcsanyi, Robert E. [2 ]
Lindfors, Tom [1 ]
机构
[1] Abo Akad Univ, Fac Sci & Engn, Johan Gadolin Proc Chem Ctr, Analyt Chem Lab, Biskopsgatan 8, FIN-20500 Turku, Finland
[2] Budapest Univ Technol & Econ, Dept Inorgan & Analyt Chem, MTA BME Lendlet Chem Nanosensors Res Grp, Szt Gellert Ter 4, H-1111 Budapest, Hungary
基金
芬兰科学院;
关键词
PLASTICIZED POLY(VINYL CHLORIDE); ATTENUATED TOTAL-REFLECTION; POTENTIOMETRIC ION SENSORS; WATER-UPTAKE; POTENTIAL STABILITY; MEMBRANE ELECTRODES; CONDUCTING POLYMERS; SILICONE-RUBBER; FTIR-ATR; TRANSDUCER;
D O I
10.1039/c5an02664d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electropolymerized hydrophobic polyazulene (PAz) based solid-state potassium ion-selective electrodes (SC-ISEs) have been characterized in terms of their suitability for clinical application. Polarization of the PAz solid contact before applying the plasticised poly(vinyl chloride) based K+-selective membrane was implemented as a convenient approach to address the general problem of the irreproducible standard potential (E-0) of SC-ISEs. Using this method, the E-0 reproducibility among different electrodes was, in the worst case, +/- 7.9 mV (n = 4). The effectiveness of the redox buffer-free approach presented here in stabilizing E0 is strengthened by the absence of light, oxygen and carbon dioxide sensitivity of the PAz SC-ISEs. No evidence was found for the formation of an aqueous layer for the PAz-based SC-ISEs. Thus the hydrophobic carbon structure of PAz having a water contact angle of 98 +/- 11 degrees, which is slightly higher than that for graphene, can apparently efficiently counteract the aqueous layer formation. In terms of the specific application, the PAz solid contact ISEs were found to show a remarkably good potential stability at their first contact with an aqueous sample. We also confirmed that the PAz-based SC-ISEs can be used for the accurate determination of the K+ concentration in serum solutions. Overall, the PAz solid contact shows significant advantages as compared to the state-of-the-art of electrically conducting polymer based SC-ISEs.
引用
收藏
页码:2990 / 2997
页数:8
相关论文
共 57 条
[21]   Reduced Graphene Oxide Films as Solid Transducers in Potentiometric All-Solid-State Ion-Selective Electrodes [J].
Hernandez, Rafael ;
Riu, Jordi ;
Bobacka, Johan ;
Valles, Cristina ;
Jimenez, Pablo ;
Benito, Ana M. ;
Maser, Wolfgang K. ;
Xavier Rius, F. .
JOURNAL OF PHYSICAL CHEMISTRY C, 2012, 116 (42) :22570-22578
[22]   All-Solid-State Reference Electrodes Based on Colloid-Imprinted Mesoporous Carbon and Their Application in Disposable Paper-based Potentiometric Sensing Devices [J].
Hu, Jinbo ;
Ho, Kieu T. ;
Zou, Xu U. ;
Smyrl, William H. ;
Stein, Andreas ;
Buehlmann, Philippe .
ANALYTICAL CHEMISTRY, 2015, 87 (05) :2981-2987
[23]   Critical assessment of graphene as ion-to-electron transducer for all-solid-state potentiometric sensors [J].
Jaworska, Ewa ;
Lewandowski, Wiktor ;
Mieczkowski, Jozef ;
Maksymiuk, Krzysztof ;
Michalska, Agata .
TALANTA, 2012, 97 :414-419
[24]   Ion-selective electrodes with three-dimensionally ordered macroporous carbon as the solid contact [J].
Lai, Chun-Ze ;
Fierke, Melissa A. ;
Stein, Andreas ;
Buehlmann, Philippe .
ANALYTICAL CHEMISTRY, 2007, 79 (12) :4621-4626
[25]   Paper-Based Potentiometric Ion Sensing [J].
Lan, Wen-Jie ;
Zou, Xu U. ;
Hamedi, Mahiar M. ;
Hu, Jinbo ;
Parolo, Claudio ;
Maxwell, E. Jane ;
Buehlmann, Philippe ;
Whitesides, George M. .
ANALYTICAL CHEMISTRY, 2014, 86 (19) :9548-9553
[26]   APPLICATION OF ION-SELECTIVE ELECTRODES IN CLINICAL ANALYSIS [J].
LEWENSTAM, A ;
MAJZURAWSKA, M ;
HULANICKI, A .
ELECTROANALYSIS, 1991, 3 (08) :727-734
[27]   All-solid-state potassium-selective electrode using graphene as the solid contact [J].
Li, Fenghua ;
Ye, Junjin ;
Zhou, Min ;
Gan, Shiyu ;
Zhang, Qixian ;
Han, Dongxue ;
Niu, Li .
ANALYST, 2012, 137 (03) :618-623
[28]   Dual-sorption model of water uptake in poly(vinyl chloride)-based ion-selective membranes: Experimental water concentration and transport parameters [J].
Li, Z ;
Li, XZ ;
Petrovic, S ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 1996, 68 (10) :1717-1725
[29]   Stability of the inner polyaniline solid contact layer in all-solid-state K+-selective electrodes based on plasticized poly(vinyl chloride) [J].
Lindfors, T ;
Ivaska, A .
ANALYTICAL CHEMISTRY, 2004, 76 (15) :4387-4394
[30]   Light sensitivity and potential stability of electrically conducting polymers commonly used in solid contact ion-selective electrodes [J].
Lindfors, Tom .
JOURNAL OF SOLID STATE ELECTROCHEMISTRY, 2009, 13 (01) :77-89