Real-time impedimetric droplet measurement (iDM)

被引:21
作者
Saateh, Abtin [1 ]
Kalantarifard, Ali [1 ]
Celik, Oguz Tolga [1 ]
Asghari, Mohammad [1 ]
Serhatlioglu, Murat [1 ]
Elbuken, Caglar [1 ]
机构
[1] Bilkent Univ, Natl Nanotechnol Res Ctr UNAM, Inst Mat Sci & Nanotechnol, TR-06800 Ankara, Turkey
关键词
VELOCITY FLUCTUATIONS; FLOW; MICROFLUIDICS; SIZE; CHIP; GENERATION; SYSTEM;
D O I
10.1039/c9lc00641a
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Droplet-based microfluidic systems require a precise control of droplet physical properties; hence, measuring the morphological properties of droplets is critical to obtain high sensitivity analysis. The ability to perform such measurements in real-time is another demand which has not been addressed yet. In this study, we used coplanar electrodes configured in the differential measurement mode for impedimetric measurement of size and velocity. To obtain the size of the droplets, detailed 3D finite element simulations of the system were performed. The interaction of the non-uniform electric field and the droplet was investigated. Electrode geometry optimization steps were described and design guideline rules were laid out. User-friendly software was developed for real-time observation of droplet length and velocity together with in situ statistical analysis results. A comparison between impedimetric and optical measurement tools is given. Finally, to illustrate the benefit of having real-time analysis, iDM was used to synthesize particles with a predefined monodispersity limit and to study the response times of syringe pump and pressure pump driven droplet generation devices. This analysis allows one to evaluate the 'warm-up' time for a droplet generator system, after which droplets reach the desired steady-state size required by the application of interest.
引用
收藏
页码:3815 / 3824
页数:10
相关论文
共 45 条
  • [1] MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES
    ALTMAN, DG
    BLAND, JM
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) : 307 - 317
  • [2] Droplet morphometry and velocimetry (DMV): a video processing software for time-resolved, label-free tracking of droplet parameters
    Basu, Amar S.
    [J]. LAB ON A CHIP, 2013, 13 (10) : 1892 - 1901
  • [3] Observed velocity fluctuations in monodisperse droplet generators
    Beer, Neil Reginald
    Rose, Klint Aaron
    Kennedy, Ian M.
    [J]. LAB ON A CHIP, 2009, 9 (06) : 838 - 840
  • [4] LiNbO3 integrated system for opto-microfluidic sensing
    Bettella, G.
    Zamboni, R.
    Pozza, G.
    Zaltron, A.
    Montevecchi, C.
    Pierno, M.
    Mistura, G.
    Sada, C.
    Gauthier-Manuel, L.
    Chauvet, M.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2019, 282 : 391 - 398
  • [5] The level-set method applied to droplet dynamics in the presence of an electric field
    Bjorklund, Erik
    [J]. COMPUTERS & FLUIDS, 2009, 38 (02) : 358 - 369
  • [6] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [7] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [8] Budd J. R., 2013, MEASUREMENT PROCEDUR
  • [9] Microfluidic high-throughput encapsulation and hydrodynamic self-sorting of single cells
    Chabert, Max
    Viovy, Jean-Louis
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2008, 105 (09) : 3191 - 3196
  • [10] Automated droplet measurement (ADM): an enhanced video processing software for rapid droplet measurements
    Chong, Zhuang Zhi
    Tor, Shu Beng
    Ganan-Calvo, Alfonso M.
    Chong, Zhuang Jie
    Loh, Ngiap Hiang
    Nam-Trung Nguyen
    Tan, Say Hwa
    [J]. MICROFLUIDICS AND NANOFLUIDICS, 2016, 20 (04)