Use of flow cytometry and SNARF to calibrate and measure intracellular pH in NSO cells

被引:25
作者
Bond, J [1 ]
Varley, J [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Chem Engn & Chem Technol, London SW7 2AZ, England
关键词
calibration; flow cytometry; intracellular pH; NSO; 1,5-(and-6)-carboxy seminaphtorhodafluor-1-acetoxymethyl ester;
D O I
10.1002/cyto.a.20066
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: Two calibration methods have been proposed for determining the relation between the fluorescence ratio of a pH-sensitive fluorescent indicator and intracellular pH (pH(i)). The first method uses nigericin to clamp pH(i) to external pH (pH(e)) and the second is the null point method. We compared these different calibration methods, solution conditions, and temperatures by using flow cytometry and the fluorescent dye 1,5- (and-6)-carboxy seminaphtorhodafluor-1-acetoxymethyl ester with an NS0 cell line. Methods: The nigericin method was performed in glucose solutions supplemented with KCl and 2-(N-morpholino)ethane sulphonic acid plus tris(hydroxymethyl)aminomethane (solution 1A), a mixture of K2HPO4/KH2PO4 in glucose-solution supplemented solutions (solution 2A), or bicarbonate buffered growth medium supplemented with K2HPO4/KH2PO4 (solution 2B); this allowed a range of pH, values to be used. The effect of temperature (22degreesC or 37degreesC) on the nigericin calibration curve was also investigated. The null point method was performed by using a series of solutions with a mixture of weak acid and base with a known pHi response. Results: Using solution 1A as the calibration solution resulted in acidic values of pH(i) for cells cultured in medium as compared with the values achieved with solution 2A. Using solution 213 did not affect the calibration curve. For the temperatures considered in this study, there was no affect on the calibration curve, but temperature did affect the pHi value of cells in phosphate buffered saline. The pseudo-null point method used with flow cytometry resulted in a calibration curve that was significantly different (P < 0.05) from that achieved using the nigericin method. Conclusions: Our data indicates that the choice of calibration solution can affect the reported pHi value; therefore, careful choice of solution is important. (C) 2005 Wiley-Liss, Inc.
引用
收藏
页码:43 / 50
页数:8
相关论文
共 22 条
[11]   ARGININE VASOPRESSIN ENHANCES PHI REGULATION IN THE PRESENCE OF HCO3- BY STIMULATING 3 ACID-BASE TRANSPORT-SYSTEMS [J].
GANZ, MB ;
BOYARSKY, G ;
STERZEL, RB ;
BORON, WF .
NATURE, 1989, 337 (6208) :648-651
[12]  
Karagiannis J, 2001, J CELL SCI, V114, P2929
[13]  
KURTZ I, 1993, METHOD CELL BIOL, V38, P183
[14]   Characterization of intracellular pH regulation in the guinea-pig ventricular myocyte [J].
Leem, CH ;
Lagadic-Gossmann, D ;
Vaughan-Jones, RD .
JOURNAL OF PHYSIOLOGY-LONDON, 1999, 517 (01) :159-180
[15]   Out-of-equilibrium pH transients in the guinea-pig ventricular myocyte [J].
Leem, CH ;
Vaughan-Jones, RD .
JOURNAL OF PHYSIOLOGY-LONDON, 1998, 509 (02) :471-485
[16]   REGULATION OF INTRACELLULAR PH IN EUKARYOTIC CELLS [J].
MADSHUS, IH .
BIOCHEMICAL JOURNAL, 1988, 250 (01) :1-8
[17]   NEW METHOD FOR CALCULATING PHI FROM ACCURATELY MEASURED CHANGES IN PHI INDUCED BY A WEAK ACID AND BASE [J].
SZATKOWSKI, MS ;
THOMAS, RC .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1986, 407 (01) :59-63
[18]   INTRACELLULAR PH MEASUREMENTS IN EHRLICH ASCITES TUMOR-CELLS UTILIZING SPECTROSCOPIC PROBES GENERATED INSITU [J].
THOMAS, JA ;
BUCHSBAUM, RN ;
ZIMNIAK, A ;
RACKER, E .
BIOCHEMISTRY, 1979, 18 (11) :2210-2218
[19]  
VAUGHANJONES RD, 1979, J PHYSL, V1, P111
[20]  
WELSH JP, 1996, FLOW CYTOMETRY APPL, P163