Automated cardiac output measurements by ultrasound are inaccurate at high cardiac outputs

被引:2
作者
Basdogan, F
Visser, W
Struijk, PC
Jansen, JRC
Vletter, WB
Wladimiroff, JW
Lotgering, FK
机构
[1] Acad Hosp Rotterdam, Dept Obstet & Gynecol, NL-3000 CB Rotterdam, Netherlands
[2] Acad Hosp Rotterdam, Dept Pulm Dis, NL-3000 CB Rotterdam, Netherlands
[3] Acad Hosp Rotterdam, Dept Cardiol, NL-3000 CB Rotterdam, Netherlands
关键词
cardiac output; echocardiography; pregnancy;
D O I
10.1046/j.1469-0705.2000.00128.x
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective The sonographic technique of automated cardiac output measurement (ACM) is a promising new method to measure cardiac output and could be of use in a high-risk obstetric unit in the treatment of pre-eclamptic patients. The aim was to determine the accuracy of the ACM method. Design Comparative study of the sonographic technique of ACM versus cardiac output measured by thermodilution (TD). Methods The study included 39 intensive care patients, 21 men, 13 non-pregnant women and five severely pre-eclamptic pregnant patients, with a wide range of cardiac outputs, in whom TD catheters had been inserted for clinical reasons. Two separate experienced observers, blinded to the results obtained with the other method, performed four successive measurements in each patient with either the ACA-I or TD technique. The averaged cardiac output value per patient and method tons used for comparison. Results Cardiac output was successfully measured with ACM and TD in 85 and 100% of patients, respectively. Mean cardiac output measured by ACM (6.77 +/- 1.90 L/min) was significantly lower than that measured by TD (9.12 +/- 3.06 L/min). Although cardiac output values obtained with ACM were significantly correlated with those measured by TD, the ACM values were consistently lower than TD values in the higher cardiac output range; the relationship was represented by ACM = 0.35 TD + 3.55 L/min (r =0.57, P < 0.001). The (ACM - TD) difference increased significantly with cardiac output, through a difference in strobe volume, not in heart rate. Conclusion The ACM is not an accurate tool to measure cardiac output in patients with a high cardiac output, including treated pre-eclamptic women.
引用
收藏
页码:508 / 512
页数:5
相关论文
共 10 条
[1]  
Assmann P E, 1988, J Am Soc Echocardiogr, V1, P393
[2]   CENTRAL HEMODYNAMIC ASSESSMENT OF NORMAL TERM PREGNANCY [J].
CLARK, SL ;
COTTON, DB ;
LEE, W ;
BISHOP, C ;
HILL, T ;
SOUTHWICK, J ;
PIVARNIK, J ;
SPILLMAN, T ;
DEVORE, GR ;
PHELAN, J ;
HANKINS, GDV ;
BENEDETTI, TJ ;
TOLLEY, D .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1989, 161 (06) :1439-1442
[3]   AORTIC MOTION - A POTENTIAL PITFALL IN CT IMAGING OF DISSECTION IN THE ASCENDING AORTA [J].
DUVERNOY, O ;
COULDEN, R ;
YTTERBERG, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1995, 19 (04) :569-572
[4]   THE THERMODILUTION METHOD FOR THE CLINICAL-ASSESSMENT OF CARDIAC-OUTPUT [J].
JANSEN, JRC .
INTENSIVE CARE MEDICINE, 1995, 21 (08) :691-697
[5]  
Nesser J., 1996, Journal of the American College of Cardiology, V27, p269A
[6]  
Sun JP, 1997, CIRCULATION, V95, P932
[7]   Automatic cardiac output measurement (ACOM): Clinical applications of a new noninvasive tool [J].
Trindade, PT ;
Brown, P ;
Puryear, JV ;
Popylisen, S ;
Schnittger, I .
INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 1998, 14 (03) :147-154
[8]  
TSUJINO H, 1996, MED REV, V58, P13
[9]   Maternal and perinatal outcome of temporizing management in 254 consecutive patients with severe pre-eclampsia remote from term [J].
Visser, W ;
Wallenburg, HCS .
EUROPEAN JOURNAL OF OBSTETRICS GYNECOLOGY AND REPRODUCTIVE BIOLOGY, 1995, 63 (02) :147-154
[10]   CENTRAL HEMODYNAMIC OBSERVATIONS IN UNTREATED PREECLAMPTIC PATIENTS [J].
VISSER, W ;
WALLENBURG, HCS .
HYPERTENSION, 1991, 17 (06) :1072-1077