Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors

被引:45
作者
Peng, Hao [1 ]
Levin, Craig S. [1 ]
机构
[1] Stanford Univ, Sch Med, Dept Radiol, Mol Imaging Program, Stanford, CA 94305 USA
关键词
POSITRON EMISSION MAMMOGRAPHY; CANCER; SENSITIVITY; SCANNER; INSTRUMENTATION; DENSITY;
D O I
10.1088/0031-9155/55/9/022
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We studied the performance of a dual-panel positron emission tomography ( PET) camera dedicated to breast cancer imaging using Monte Carlo simulation. The proposed system consists of two 4 cm thick 12 x 15 cm(2) area cadmium zinc telluride (CZT) panels with adjustable separation, which can be put in close proximity to the breast and/or axillary nodes. Unique characteristics distinguishing the proposed system from previous efforts in breast-dedicated PET instrumentation are the deployment of CZT detectors with superior spatial and energy resolution, using a cross-strip electrode readout scheme to enable 3D positioning of individual photon interaction coordinates in the CZT, which includes directly measured photon depth-of-interaction (DOI), and arranging the detector slabs edge-on with respect to incoming 511 keV photons for high photon sensitivity. The simulation results show that the proposed CZT dual-panel PET system is able to achieve superior performance in terms of photon sensitivity, noise equivalent count rate, spatial resolution and lesion visualization. The proposed system is expected to achieve similar to 32% photon sensitivity for a point source at the center and a 4 cm panel separation. For a simplified breast phantom adjacent to heart and torso compartments, the peak noise equivalent count (NEC) rate is predicted to be similar to 94.2 kcts s(-1) (breast volume: 720 cm(3) and activity concentration: 3.7 kBq cm(-3)) for a similar to 10% energy window around 511 keV and similar to 8 ns coincidence time window. The system achieves 1 mm intrinsic spatial resolution anywhere between the two panels with a 4 cm panel separation if the detectors have DOI resolution less than 2 mm. For a 3 mm DOI resolution, the system exhibits excellent sphere resolution uniformity (sigma(rms)/mean) <= 10%) across a 4 cm width FOV. Simulation results indicate that the system exhibits superior hot sphere visualization and is expected to visualize 2 mm diameter spheres with a 5:1 activity concentration ratio within roughly 7 min imaging time. Furthermore, we observe that the degree of spatial resolution degradation along the direction orthogonal to the two panels that is typical of a limited angle tomography configuration is mitigated by having high-resolution DOI capabilities that enable more accurate positioning of oblique response lines.
引用
收藏
页码:2761 / 2788
页数:28
相关论文
共 55 条
  • [1] PET/CT scanner instrumentation, challenges, and solutions
    Alessio, AM
    Kinahan, PE
    Cheng, PM
    Vesselle, H
    Karp, JS
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2004, 42 (06) : 1017 - +
  • [2] Avril N, 2001, J NUCL MED, V42, P9
  • [3] Breast cancer imaging studies with a variable field of view PET cameral
    Baghaei, H
    Wong, WH
    Uribe, J
    Li, HD
    Zhang, N
    Wang, Y
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) : 1080 - 1084
  • [4] SENSITIVITY AND SPECIFICITY OF 1ST SCREEN MAMMOGRAPHY IN THE CANADIAN NATIONAL BREAST SCREENING STUDY - A PRELIMINARY-REPORT FROM 5 CENTERS
    BAINES, CJ
    MILLER, AB
    WALL, C
    MCFARLANE, DV
    SIMOR, IS
    JONG, R
    SHAPIRO, BJ
    AUDET, L
    PETITCLERC, M
    OUIMETOLIVA, D
    LADOUCEUR, J
    HEBERT, G
    MINUK, T
    HARDY, G
    STANDING, HK
    [J]. RADIOLOGY, 1986, 160 (02) : 295 - 298
  • [5] CHARGE-TRANSPORT IN ARRAYS OF SEMICONDUCTOR GAMMA-RAY DETECTORS
    BARRETT, HH
    ESKIN, JD
    BARBER, HB
    [J]. PHYSICAL REVIEW LETTERS, 1995, 75 (01) : 156 - 159
  • [6] BERGER MJ, 1999, 8 NIST
  • [7] Computed tomography for imaging the breast
    Boone, John M.
    Kwan, Alex L. C.
    Yang, Kai
    Burkett, George W.
    Lindfors, Karen K.
    Nelson, Thomas R.
    [J]. JOURNAL OF MAMMARY GLAND BIOLOGY AND NEOPLASIA, 2006, 11 (02) : 103 - 111
  • [8] Initial Characterization of a Dedicated Breast PET/CT Scanner During Human Imaging
    Bowen, Spencer L.
    Wu, Yibao
    Chaudhari, Abhijit J.
    Fu, Lin
    Packard, Nathan J.
    Burkett, George W.
    Yang, Kai
    Lindfors, Karen K.
    Shelton, David K.
    Hagge, Rosalie
    Borowsky, Alexander D.
    Martinez, Steve R.
    Qi, Jinyi
    Boone, John M.
    Cherry, Simon R.
    Badawi, Ramsey D.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (09) : 1401 - 1408
  • [9] Brasse D, 2005, J NUCL MED, V46, P859
  • [10] Cone-beam volume CT breast imaging: Feasibility study
    Chen, B
    Ning, R
    [J]. MEDICAL PHYSICS, 2002, 29 (05) : 755 - 770