Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beams

被引:33
|
作者
Li, XA
Ma, CM
Salhani, D
机构
[1] Ottawa Reg Canc Ctr, Dept Med Phys, Ottawa, ON K1Y 4K7, Canada
[2] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1997年 / 42卷 / 12期
关键词
D O I
10.1088/0031-9155/42/12/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work, nine commonly used dosimetry detectors have been investigated to determine suitable relative dosimeters for kilovoltage x-ray beams. By comparison with the Monte Carlo calculated data, it was determined that for the detectors studied the PTW N23342, Markus and NACP parallel-plate chambers are more suitable for the measurement of percentage depth dose (PDD) data for this beam quality range with an uncertainty of about 3%. A diode detector may be used to measure the PDD for the 100 kVp beam, but it is not suitable for higher energies (300 kVp). The Capintec parallel-plate chamber may be adequate for medium-energy photons, but it has a slightly higher uncertainty for low-energy x-rays (100 kVp). For the measurement of beam profiles, diode and film yield incorrect profile tails, which can be corrected using the RK ionization chamber.
引用
收藏
页码:2561 / 2568