Estimating changes in cardiac output using an implanted hemodynamic monitor in heart failure patients

被引:5
|
作者
Stahlberg, Marcus [1 ]
Damgaard, Morten [2 ]
Ersgard, David
Norsk, Peter [2 ]
Karamanoglu, Mustafa [3 ]
Linde, Cecilia
Braunschweig, Frieder
机构
[1] Karolinska Inst, Dept Cardiol, Res Unit, Karolinska Univ Hosp, S-17176 Stockholm, Sweden
[2] Univ Copenhagen, Dept Biomed Sci, Fac Hlth Sci, DK-1168 Copenhagen, Denmark
[3] Medtronic Inc, NT&D Res, Minneapolis, MN USA
关键词
Cardiac output; hemodynamic monitoring; heart failure; pressure waveform; RIGHT-VENTRICULAR PRESSURE; FEASIBILITY; MULTICENTER; DIAGNOSIS; SYSTEM;
D O I
10.3109/14017431003631987
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The aim of this study was to evaluate an algorithm that estimates changes in cardiac output (CO) from right ventricular (RV) pressure waveforms derived from an implantable hemodynamic monitor (IHM) in heart failure patients. Design. Twelve heart failure patients (NYHA II-III, EF 32%) with an implantable hemodynamic monitor (Chronicle (R)) were included in this study. Changes in cardiac output were provoked by body position change at rest (left lateral supine, horizontal supine, sitting, and standing) and a steady state bicycle exercise at 20 watts. Estimated CO derived from the IHM (CO(IHM)) was compared to CO measured with inert gas rebreathing (CO(RB)), echocardiography (COE(CHO)) and impedance cardiography (CO(ICG)). CO(RB) was considered the reference method. Results. The median intra-patient correlation coefficient comparing CO(RB) and CO(IHM) was 0.83 (range: 0.63-0.98). Comparing CO(RB) with CO(ECHO) and CO(ICG) resulted in mean intra-patient correlation coefficients of 0.73 (-0.29-0.94) and 0.63 (-0.29-0.96). In a statistical model where slope and intercept was considered random between patients the coefficient of determination (R(2)) comparing CO(RB) and CO(IHM) was 0.91. Mean bias was -0.39 L/min (11%). Limits of agreement were +/- 1.56 L/min and relative error was 21%. Conclusions. A simple algorithm based on RV pressure wave form characteristics derived from an IHM can be used to estimate changes in CO in heart failure patients. These findings encourage further research aiming to improve and validate the algorithm.
引用
收藏
页码:201 / 208
页数:8
相关论文
共 50 条
  • [21] Validity of Hemodynamic Monitoring Using Inert Gas Rebreathing Method in Patients With Chronic Heart Failure and Those Implanted With a Left Ventricular Assist Device
    Okwose, Nduka C.
    Bouzas-Cruz, Noelia
    Fernandez, Oscar Gonzalez
    Koshy, Aaron
    Green, Thomas
    Woods, Andrew
    Robinson-Smith, Nicola
    Tovey, Sian
    Mcdiarmid, Adam
    Parry, Gareth
    Schueler, Stephan
    Macgowan, Guy A.
    Jakovljevic, Djordje G.
    JOURNAL OF CARDIAC FAILURE, 2021, 27 (04) : 414 - 418
  • [22] Improved Accuracy of Automated Estimation of Cardiac Output Using Circulation Time in Patients with Heart Failure
    Dajani, Hilmi R.
    Hosokawa, Kazuya
    Ando, Shin-ichi
    JOURNAL OF CARDIAC FAILURE, 2016, 22 (11) : 925 - 927
  • [23] Relationship of changes in cardiac output to changes in heart rate in medical ICU patients
    Sasse, SA
    Chen, PA
    Mahutte, CK
    INTENSIVE CARE MEDICINE, 1996, 22 (05) : 409 - 414
  • [24] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Wonshill Koh
    Kristin A. Schneider
    Huaiyu Zang
    Sarosh P. Batlivala
    Matthew P. Monteleone
    Alexis L. Benscoter
    Meghan M. Chlebowski
    Ilias D. Iliopoulos
    David S. Cooper
    Pediatric Cardiology, 2022, 43 : 1205 - 1213
  • [25] Non-invasive measurement of cardiac output in heart failure patients using a new foreign gas rebreathing technique
    Gabrielsen, A
    Videbæk, R
    Schou, M
    Damgaard, M
    Kastrup, J
    Norsk, P
    CLINICAL SCIENCE, 2002, 102 (02) : 247 - 252
  • [26] Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology
    Koh, Wonshill
    Schneider, Kristin A.
    Zang, Huaiyu
    Batlivala, Sarosh P.
    Monteleone, Matthew P.
    Benscoter, Alexis L.
    Chlebowski, Meghan M.
    Iliopoulos, Ilias D.
    Cooper, David S.
    PEDIATRIC CARDIOLOGY, 2022, 43 (06) : 1205 - 1213
  • [27] Myocardial remodeling in patients with chronic heart failure and implanted cardiac contractility modulators
    Safiullina, Alfiya A.
    Uskach, Tatiana M.
    Dobrovolskaya, Svetlana, V
    Saidova, Marina A.
    Zhirov, Igor, V
    Tereshchenko, Sergey N.
    TERAPEVTICHESKII ARKHIV, 2021, 93 (12) : 1443 - 1450
  • [28] Pulse contour cardiac output monitoring in acute heart failure patients
    Wernly, Bernhard
    Lichtenauer, Michael
    Franz, Marcus
    Fritzenwanger, Michael
    Kabisch, Bjoern
    Figulla, Hans-Reiner
    Jung, Christian
    WIENER KLINISCHE WOCHENSCHRIFT, 2016, 128 (23-24) : 864 - 869
  • [29] Bioimpedance cardiac output measurements in patients with presumed congestive heart failure
    Weiss, SJ
    Kulik, JP
    Calloway, E
    ACADEMIC EMERGENCY MEDICINE, 1997, 4 (06) : 568 - 573
  • [30] The Use of the Internet to Remotely Monitor Patients with Heart Failure
    Maric, Biljana
    Kaan, Annemarie
    Araki, Yuriko
    Ignaszewski, Andrew
    Lear, Scott A.
    TELEMEDICINE JOURNAL AND E-HEALTH, 2010, 16 (01): : 26 - 33