High pulmonary vascular resistance in addition to low right ventricular stroke work index effectively predicts biventricular assist device requirement

被引:24
作者
Imamura, Teruhiko [1 ]
Kinugawa, Koichiro [1 ]
Kinoshita, Osamu [2 ]
Nawata, Kan [2 ]
Ono, Minoru [2 ]
机构
[1] Univ Tokyo, Grad Sch Med, Dept Therapeut Strategy Heart Failure, Bunkyo Ku, 7-3-1 Hongo, Tokyo 1138655, Japan
[2] Univ Tokyo, Grad Sch Med, Dept Cardiac Surg, Tokyo 1138655, Japan
关键词
VAD; Heart failure; INTERMACS; HEART-FAILURE; RISK SCORE; HYPERTENSION;
D O I
10.1007/s10047-015-0867-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Although the right ventricular stroke work index (RVSWI) is a good index for RV function, a low RVSWI is not necessarily an indicator for the need for a right ventricular assist device at the time of left VAD implantation. We here aimed to determine a more precise indicator for the need for a biventricular assist device (BiVAD). In total, 116 patients (mean age, 38 +/- A 14 years), who underwent hemodynamic assessments preoperatively including 12 BiVAD patients, and had been followed at our institute from 2003 to 2015, were included. Multivariate logistic regression analysis indicated that RVSWI and pulmonary vascular resistance (PVR) were independent predictors of BiVAD requirement (P < 0.05 for both). In addition, all patients were classified into 4 groups: (1) normal (RVSWI > 5 g/m, PVR < 3.7 WU), (2) pulmonary hypertension (RVSWI > 5, PVR > 3.7), (3) RV failure (RVSWI < 5, PVR < 3.7), and (4) both pulmonary hypertension and RV failure (RVSWI < 5, PVR > 3.7), and examined. Most of the patients in Group 4 (75 %), with acutely depressed hemodynamics and inflammatory responses in the myocardium, required BiVAD. Overall, patients with BiVAD had a worse survival rate as compared with those with LVAD alone. In conclusion, high PVR in addition to low RVSWI effectively predicts BiVAD requirement.
引用
收藏
页码:44 / 53
页数:10
相关论文
共 19 条
[1]   Survival after biventricular assist device implantation: An analysis of the Interagency Registry for Mechanically Assisted Circulatory Support database [J].
Cleveland, Joseph C., Jr. ;
Naftel, David C. ;
Reece, T. Brett ;
Murray, Margaret ;
Antaki, James ;
Pagani, Francis D. ;
Kirklin, James K. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2011, 30 (08) :862-869
[2]   Early planned institution of biventricular mechanical circulatory support results in improved outcomes compared with delayed conversion of a left ventricular assist device to a biventricular assist device [J].
Fitzpatrick, J. Raymond, III ;
Frederick, John R. ;
Hiesinger, William ;
Hsu, Vivian M. ;
McCormick, Ryan C. ;
Kozin, Elliott D. ;
Laporte, Carine M. ;
O'Hara, Mary Lou ;
Howell, Elan ;
Dougherty, Deborah ;
Cohen, Jeffrey E. ;
Southerland, Kevin W. ;
Howard, Jessica L. ;
Paulson, Carter ;
Acker, Michael A. ;
Morris, Rohinton J. ;
Woo, Y. Joseph .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (04) :971-977
[3]   Risk Score Derived from Pre-operative Data Analysis Predicts the Need for Biventricular Mechanical Circulatory Support [J].
Fitzpatrick, J. Raymond, III ;
Frederick, John R. ;
Hsu, Vivian M. ;
Kozin, Elliott D. ;
O'Hara, Mary Lou ;
Howell, Elan ;
Dougherty, Deborah ;
McCormick, Ryan C. ;
Laporte, Carine A. ;
Cohen, Jeffrey E. ;
Southerland, Kevin W. ;
Howard, Jessica L. ;
Jessup, Mariell L. ;
Morris, Rohinton J. ;
Acker, Michael A. ;
Woo, Y. Joseph .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (12) :1286-1292
[4]   Preoperative risk factors for right ventricular failure after implantable left ventricular assist device insertion [J].
Fukamachi, K ;
McCarthy, PM ;
Smedira, NG ;
Vargo, RL ;
Starling, RC ;
Young, JB .
ANNALS OF THORACIC SURGERY, 1999, 68 (06) :2181-2184
[5]   Biventricular failure with low pulmonary vascular resistance was managed by left ventricular assist device alone without right-sided mechanical support [J].
Imamura, Teruhiko ;
Kinugawa, Koichiro ;
Nitta, Daisuke ;
Hatano, Masaru ;
Kinoshita, Osamu ;
Nawata, Kan ;
Ono, Minoru .
JOURNAL OF ARTIFICIAL ORGANS, 2015, 18 (03) :272-275
[6]   Late-Onset Right Ventricular Failure in Patients With Preoperative Small Left Ventricle After Implantation of Continuous Flow Left Ventricular Assist Device [J].
Imamura, Teruhiko ;
Kinugawa, Koichiro ;
Kato, Naoko ;
Muraoka, Hironori ;
Fujino, Takeo ;
Inaba, Toshiro ;
Maki, Hisataka ;
Kinoshita, Osamu ;
Hatano, Masaru ;
Kyo, Shunei ;
Ono, Minoru .
CIRCULATION JOURNAL, 2014, 78 (03) :625-633
[7]   Novel Risk Scoring System With Preoperative Objective Parameters Gives a Good Prediction of 1-Year Mortality in Patients With a Left Ventricular Assist Device [J].
Imamura, Teruhiko ;
Kinugawa, Koichiro ;
Shiga, Taro ;
Endo, Miyoko ;
Kato, Naoko ;
Inaba, Toshiro ;
Maki, Hisataka ;
Hatano, Masaru ;
Yao, Atsushi ;
Nishimura, Takashi ;
Hirata, Yasunobu ;
Kyo, Shunei ;
Ono, Minoru ;
Nagai, Ryozo .
CIRCULATION JOURNAL, 2012, 76 (08) :1895-1903
[8]   How to Treat Stage D Heart Failure? - When to Implant Left Ventricular Assist Devices in the Era of Continuous Flow Pumps? - [J].
Kinugawa, Koichiro .
CIRCULATION JOURNAL, 2011, 75 (09) :2038-2045
[9]   Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: Incidence, risk factors, and effect on outcomes [J].
Kormos, Robert L. ;
Teuteberg, Jeffrey J. ;
Pagani, Francis D. ;
Russell, Stuart D. ;
John, Ranjit ;
Miller, Leslie W. ;
Massey, Todd ;
Milano, Carmelo A. ;
Moazami, Nader ;
Sundareswaran, Kartik S. ;
Farrar, David J. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (05) :1316-1324
[10]   RIGHT-VENTRICULAR PULMONARY-ARTERIAL INTERACTIONS [J].
KUSSMAUL, WG ;
NOORDERGRAAF, A ;
LASKEY, WK .