Complications associated with the use of the esophageal-tracheal combitube

被引:58
|
作者
Vézina, D
Lessard, MR
Bussières, J
Topping, C
Trépanier, CA
机构
[1] Univ Quebec, Ctr Hosp, Dept Anesthesie, St Foy, PQ G1V 2M3, Canada
[2] Univ Laval, Dept Anesthesie, Hop Laval, Quebec City, PQ G1K 7P4, Canada
[3] Regie Reg Sante & Serv Sociaux Quebec, Dept Anesthesie, Quebec City, PQ, Canada
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1998年 / 45卷 / 01期
关键词
Difficult Airway; Subcutaneous Emphysema; Tracheal Lumen; Emergency Medical Technician; Proximal Balloon;
D O I
10.1007/BF03011999
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Purpose: To report four cases of subcutaneous emphysema, pneumomediastinum and pneumoperitoneum associated with the use of-the Esophageal-Tracheal Combitube((R)) (ETC) during prehospital management of cardiac arrest. Clinical features: Between September 1994 and April 1996, 1139 patients were resuscitated with the ETC and the semiautomated external defibrillator as part of the CPR protocol for prehospital management of cardiac arrest by basic emergency medical technicians, Eight of these patients presented with subcutaneous emphysema. Four of them, declared dead after arrival in the emergency room (ER), had autopsy studies, In two, autopsy revealed large (6 and 6.5 cm respectively) longitudinal transparietal lacerations of the anterior wall of the oesophagus. Multiple superficial lacerations of the oesophagus were also present in another patient, while no lesion of the airway or the oesophagus was found in the last patient. Conclusion: These cases suggest that subcutaneous emphysema, pneumomediastinum and pneumoperitoneum might be complications associated with the use of the FTC. Al least in two cases, oesophageal laceration appears to be the mechanism by which these complications occurred.
引用
收藏
页码:76 / 80
页数:5
相关论文
共 50 条
  • [1] Complications associated with the use of the Esophageal-Tracheal Combitube
    Daniel Vézina
    Martin R. Lessard
    Jean Bussières
    Claude Topping
    Claude A. Trépanier
    Canadian Journal of Anaesthesia, 1998, 45 : 76 - 80
  • [2] The Esophageal-Tracheal Combitube® and esophageal injuries
    Sonia J. Vaida
    Canadian Journal of Anesthesia, 2007, 54 : 491 - 492
  • [3] The esophageal-tracheal combitube® and esophageal injuries
    Vaida, Sonia J.
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2007, 54 (06): : 491 - 492
  • [4] Esophageal and tracheal distortion by the Esophageal-Tracheal Combitube: a cadaver study
    Daniel Vézina
    Claude A. Trépanier
    Martin R. Lessard
    Jean Bussières
    Canadian Journal of Anaesthesia, 1999, 46 : 393 - 397
  • [5] Use of the esophageal tracheal combitube™ by basic emergency medical technicians
    Lefrançois, DP
    Dufour, DG
    RESUSCITATION, 2002, 52 (01) : 77 - 83
  • [6] Complications with the Combitube
    Peter Krafft
    Michael Frass
    Allan P. Reed
    Canadian Journal of Anaesthesia, 1998, 45 : 823 - 824
  • [7] Retrospective study of complications associated with the combitube®
    Marie-Claude Vézina
    Pierre C. Nicole
    Claude A. Trépanier
    Martin R. Lessard
    Canadian Journal of Anesthesia, 2005, 52 (Suppl 1): : A125 - A125
  • [8] Bronchoscopy via a redesigned Combitube(TM) in the esophageal position - A clinical evaluation
    Krafft, P
    Roggla, M
    Fridrich, P
    Locker, GJ
    Frass, M
    Benumof, JL
    ANESTHESIOLOGY, 1997, 86 (05) : 1041 - 1045
  • [9] Tracheal intubation related complications in the prehospital setting
    Caruana, Emmanuel
    Duchateau, Francois-Xavier
    Cornaglia, Carole
    Devaud, Marie-Laure
    Pirracchio, Romain
    EMERGENCY MEDICINE JOURNAL, 2015, 32 (11) : 882 - 887
  • [10] Management of tracheal perforations: Potential complications and pitfalls
    Prange, T.
    EQUINE VETERINARY EDUCATION, 2015, 27 (11) : 566 - 568