Left Ventricular Assist Device Flow Pattern Analysis Using a Novel Model Incorporating Left Ventricular Pulsatility

被引:14
作者
Grinstein, Jonathan [1 ]
Torii, Ryo [2 ]
Bourantas, Christos V. [3 ]
Garcia-Garcia, Hector M. [4 ]
机构
[1] Univ Chicago, Dept Med, Cardiol Sect, Chicago, IL 60637 USA
[2] UCL, Dept Mech Engn, London, England
[3] Barts Heart Ctr, Dept Cardiol, London, England
[4] MedStar Cardiovasc Res Network, Washington, DC USA
关键词
computational fluid dynamics; LVAD flow; Programable flow algorithms; shear stress; shear rate; COMPUTATIONAL FLUID-DYNAMICS; HEARTMATE II; BLOOD-FLOW; SUPPORT; THROMBOSIS; STRESS; SYSTEM;
D O I
10.1097/MAT.0000000000001341
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our current understanding of flow through the circuit of left ventricular assist device (LVAD), left ventricle and ascending aorta remains incompletely understood. Computational fluid dynamics, which allow for analysis of flow in the cardiovascular system, have been used for this purpose, although current simulation models have failed to fully incorporate the interplay between the pulsatile left ventricle and continuous-flow generated by the LVAD. Flow-through the LVAD is dependent on the interaction between device and patient-specific factors with suboptimal flow patterns evoking increased risk of LVAD-related complications. Computational fluid dynamics can be used to analyze how different pump and patient factors affect flow patterns in the left ventricle and the aorta. Computational fluid dynamics simulations were carried out on a patient with a HeartMate II. Simulations were also conducted for theoretical scenarios substituting HeartWare HVAD, HeartMate 3 (HM3) in continuous mode and HM3 with Artificial Pulse. An anatomical model of the patient was reconstructed from computed tomography (CT) images, and the LVAD outflow was used as the inflow boundary condition. The LVAD outflow was calculated separately using a lumped-parameter-model of the systemic circulation, which was calibrated to the patient based on the patient-specific ventricular volume change reconstructed from 4 dimensional computed tomography and pulmonary capillary wedge pressure tracings. The LVADs were implemented in the lumped-parameter-model via published pressure head versus flow (H-Q) curves. To quantify the flushing effect, virtual contrast agent was released in the ascending aorta and its flushing over the cycles was quantified. Shear stress acting on the aortic endothelium and shear rate in the bloodstream were also quantified as indicators of normal/abnormal blood flow, especially the latter being a biomarker of platelet activation and hemolysis. LVAD speeds for the HVAD and HM3 were selected to match flow rates for the patient's HMII (9,000 RPM for HMII, 5,500 RPM for HM3, and 2,200 RPM for HVAD), the cardiac outputs were 5.81 L/min, 5.83 L/min, and 5.92 L/min, respectively. The velocity of blood flow in the outflow cannula was higher in the HVAD than in the two HeartMate pumps with a cycle average (range) of 0.92 m/s (0.78-1.19 m/s), 0.91 m/s (0.86-1.00 m/s), and 1.74 m/s (1.40-2.24 m/s) for HMII, HM3, and HVAD, respectively. Artificial pulse increased the peak flow rate to 9.84 L/min for the HM3 but the overall cardiac output was 5.96 L/min, which was similar to the continuous mode. Artificial pulse markedly decreased blood stagnation in the ascending aorta; after six cardiac cycles, 48% of the blood was flushed out from the ascending aorta under the continuous operation mode while 60% was flushed under artificial pulse. Shear stress and shear rate in the aortic arch were higher with the HVAD compared to the HMII and HM3, respectively (shear stress: 1.76 vs. 1.33 vs. 1.33 Pa, shear rate: 136 vs. 91.5 vs. 89.4 s(-1)). Pump-specific factors such as LVAD type and programmed flow algorithms lead to unique flow patterns which influence blood stagnation, shear stress, and platelet activation. The pump-patient interaction can be studied using a novel computational fluid dynamics model to better understand and potentially mitigate the risk of downstream LVAD complications.
引用
收藏
页码:724 / 732
页数:9
相关论文
共 50 条
  • [41] Readmissions after continuous flow left ventricular assist device implantation
    Kimura, Mitsutoshi
    Nawata, Kan
    Kinoshita, Osamu
    Yamauchi, Haruo
    Hoshino, Yasuhiro
    Hatano, Masaru
    Amiya, Eisuke
    Kashiwa, Koichi
    Endo, Miyoko
    Kagami, Yukie
    Nemoto, Mariko
    Ono, Minoru
    JOURNAL OF ARTIFICIAL ORGANS, 2017, 20 (04) : 311 - 317
  • [42] CrossTalk proposal: Blood flow pulsatility in left ventricular assist device patients is essential to maintain normal brain physiology
    Stohr, Eric J.
    McDonnell, Barry J.
    Colombo, Paolo C.
    Willey, Joshua Z.
    JOURNAL OF PHYSIOLOGY-LONDON, 2019, 597 (02): : 353 - 356
  • [43] Continuous-flow left ventricular assist device implantation in the presence of a hostile ventricular apex
    Atluri, Pavan
    Dymond, Daniel J.
    Woo, Y. Joseph
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (04) : 981 - 982
  • [44] Continuous-Flow Left Ventricular Assist Device Implantation in Patients With a Small Left Ventricle
    Kawabori, Masashi
    Kurihara, Chitaru
    Sugiura, Tadahisa
    Cohn, William E.
    Civitello, Andrew B.
    Frazier, O. Howard
    Morgan, Jeffrey A.
    ANNALS OF THORACIC SURGERY, 2018, 105 (03) : 799 - 806
  • [45] Computational fluid dynamics analysis of a Maglev centrifugal left ventricular assist device
    Burgreen, GW
    Loree, HM
    Bourque, K
    Dague, C
    Poirier, VL
    Farrar, D
    Hampton, E
    Wu, ZJ
    Gempp, TM
    Schöb, R
    ARTIFICIAL ORGANS, 2004, 28 (10) : 874 - 880
  • [46] Sound analysis of a left ventricular assist device: A technical evaluation of iOS devices
    Sundbom, Per
    Roth, Michael
    Granfeldt, Hans
    Karlsson, Daniel M.
    Ahn, Henrik
    Gustafsson, Fredrik
    Hubbert, Laila
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2018, 41 (05) : 254 - 260
  • [47] In silico analysis of pressure distribution and flow profiles across an experimental left ventricular assist device accessory
    Osypka, Anna
    Meissner, Florian
    Ozturk, Deniz
    Windisch, Roxane
    Vestner, Heiko
    Galbas, Michelle Costa
    Czerny, Martin
    Bothe, Wolfgang
    INTERDISCIPLINARY CARDIOVASCULAR AND THORACIC SURGERY, 2025, 40 (03):
  • [48] Diagnosis and Treatment Algorithm for Blood Flow Obstructions in Patients With Left Ventricular Assist Device
    Scandroglio, Anna Mara
    Kaufmann, Friedrich
    Pieri, Marina
    Kretzschmar, Alexandra
    Mueller, Marcus
    Pergantis, Panagiotis
    Dreysse, Stephan
    Falk, Volkmar
    Krabatsch, Thomas
    Potapov, Evgenij V.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (23) : 2758 - 2768
  • [49] Asymptomatic Sustained Ventricular Fibrillation in a Patient With Left Ventricular Assist Device
    Busch, Mathias C.
    Haap, Michael
    Kristen, Arnt
    Haas, Christian S.
    ANNALS OF EMERGENCY MEDICINE, 2011, 57 (01) : 25 - 28
  • [50] Anthracycline Treatment and Ventricular Remodeling in Left Ventricular Assist Device Patients
    Segura, Ana Maria
    Radovancevic, Rajko
    Demirozu, Zumrat T.
    Frazier, O. H.
    Buja, L. Maximilian
    TEXAS HEART INSTITUTE JOURNAL, 2014, 42 (02): : 124 - 130