The response of a large cylindrical 76 mm x 76 mm (height x diameter) cerium tribromide (CeBr3) detector was measured for prompt gamma rays. The total intrinsic activity of the CeBr3 detector, which was measured over 0.33-333 MeV range, was found to be 0.022 +/- 0.001 counts/s/cm(3). The partial intrinsic activity ( due to Ac-227 contamination), was measured over a energy range of 1.22-2.20 MeV energy, was found to be 0.007 +/- 0.001 counts/s/cm(3). Compared to intrinsic activities of LaBr3:Ce and LaCl3:Ce detectors of equivalent volume, the CeBr3 detector has 7-8 times less total intrinsic activity. The detector response for low energy prompt gamma rays was measured over 0.3-0.6 MeV gamma energy range using a portable neutron generator-based Prompt Gamma Neutron Activation Analysis (PGNAA) setup. The experimental yield of boron, cadmium and mercury prompt gamma-rays was measured from water samples contaminated with 0.75-2.5 wt mercury, 031-2.50 wt% boron, and 0.0625-0.500 wt% cadmium, respectively. An excellent agreement has been observed between the calculated and experimental yields of the gamma rays. Also minimum detection limit (MDC) of the CeBr3 detector was measured for boron, cadmium and mercury samples. The CeBr3 detector has 23% smaller value of MDCB and 18% larger value of MDCCd than those of a LaBr3:Ce detector of equivalent size. This study has shown that CeBr3 detector has an excellent response for the low energy prompt gamma-rays with almost an order of magnitude low intrinsic activity as compared to LaCl3:Ce and LaBr3:Ce detectors of equivalent volume. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:50 / 56
页数:7
相关论文
共 14 条
[1]   Advanced compact accelerator neutron generator technology for active neutron interrogation field work [J].
Chichester, D. L. ;
Simpson, J. D. ;
Lemchak, M. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2007, 271 (03) :629-637
[2]  
Choi H.D., 2006, DATABASE PROMPT GAMM
[3]   Light output and energy resolution of Ce3+-doped scintillators [J].
Dorenbos, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 486 (1-2) :208-213
[4]   CeBr3 scintillator development for possible use in space missions [J].
Drozdowski, W. ;
Dorenbos, P. ;
Bos, A. J. J. ;
Bizarri, G. ;
Owens, A. ;
Quarati, F. G. A. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :1391-1396
[5]  
Guss P. P., 2010, P SPIE, V7805
[6]   SIZE EFFECT ON NUCLEAR GAMMA-RAY ENERGY SPECTRA ACQUIRED BY DIFFERENT-SIZED CeBr3, LaBr3:Ce, AND NaI:Tl GAMMA-RAY DETECTORS [J].
Guss, Paul ;
Reed, Michael ;
Yuan, Ding ;
Beller, Denis ;
Cutler, Matthew ;
Contreras, Christopher ;
Mukhopadhyay, Sanjoy ;
Wilde, Scott .
NUCLEAR TECHNOLOGY, 2014, 185 (03) :309-321
[7]   Pulse height tests of a large diameter fast LaBr3:Ce scintillation detector [J].
Naqvi, A. A. ;
Khiari, F. Z. ;
Maslehuddin, M. ;
Gondal, M. A. ;
Al-Amoudi, O. S. B. ;
Ukashat, M. S. ;
Ilyas, A. M. ;
Liadi, F. A. ;
Isab, A. A. ;
Rehman, Khateeb-ur ;
Raashid, M. ;
Dastageer, M. A. .
APPLIED RADIATION AND ISOTOPES, 2015, 104 :224-231
[8]   Prompt gamma tests of LaBr3:Ce and BGO detectors for detection of hydrogen, carbon and oxygen in bulk samples [J].
Naqvi, A. A. ;
Al-Matouq, Fares A. ;
Khiari, F. Z. ;
Isab, A. A. ;
Rehman, Khateeb-ur ;
Raashid, M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 684 :82-87
[9]   Response tests of a LaCl3:Ce scintillation detector with low energy prompt gamma rays from boron and cadmium [J].
Naqvi, A. A. ;
Al-Anezi, M. S. ;
Kalakada, Zameer ;
Al Matouq, Faris A. ;
Maslehuddin, M. ;
Gondal, M. A. ;
Isab, A. A. ;
Khateeb-ur-Rehman ;
Dastageer, M. .
APPLIED RADIATION AND ISOTOPES, 2012, 70 (05) :882-887
[10]   Detection efficiency of low levels of boron and cadmium with a LaBr3:Ce scintillation detector [J].
Naqvi, A. A. ;
Al-Anezi, M. S. ;
Kalakada, Zameer ;
Isab, A. A. ;
Raashid, M. ;
Al Matouq, Faris Ahmed ;
Khateeb-ur-Rehman ;
Khiari, F. Z. ;
Garwan, M. A. ;
Al-Amoudi, O. S. B. ;
Maslehuddin, M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 665 :74-79