Accuracy of Methemoglobin Detection by Pulse CO-Oximetry During Hypoxia

被引:43
作者
Feiner, John R. [1 ]
Bickler, Philip E. [1 ]
Mannheimer, Paul D. [2 ]
机构
[1] Univ Calif San Francisco, Dept Anesthesia & Perioperat Care, San Francisco, CA 94143 USA
[2] Resp & Monitoring Solut, Boulder, CO USA
关键词
OXYGEN-SATURATION; READINGS;
D O I
10.1213/ANE.0b013e3181c91bb6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Methemoglobin in the blood cannot be detected by conventional pulse oximetry, although it can bias the oximeter's estimate (Spo(2)) of the true arterial functional oxygen saturation (Sao(2)). A recently introduced "Pulse CO-Oximeter" (Masimo Rainbow SET (R) Radical-7 Pulse CO-Oximeter, Masimo Corp., Irvine, CA) is intended to additionally monitor noninvasively the fractional carboxyhemoglobin and methemoglobin content in blood. The purpose of our study was to determine whether hypoxia affects the new device's estimated methemoglobin reading accuracy, and whether the presence of methemoglobin impairs the ability of the Radical-7 and a conventional pulse oximeter (Nonin 9700, Nonin Medical Inc., Plymouth, MN) to detect decreases in Sao(2). METHODS: Eight and 6 healthy adults were included in 2 study groups, respectively, each fitted with multiple sensors and a radial arterial catheter for blood sampling. In the first group, IV administration of approximately 300 mg sodium nitrite increased subjects' methemoglobin level to a 7% to 8% target and hypoxia was induced to different levels of Sao(2) (70%-100%) by varying fractional inspired oxygen. In the second group, 15% methemoglobin at room air and 80% Sao(2) were targeted. Pulse CO-oximeter readings were compared with arterial blood values measured using a Radiometer multiwavelength hemoximeter. Pulse CO-oximeter methemoglobin reading performance was analyzed by observing the incidence of meaningful reading errors at the various hypoxia levels. This was used to determine the impact on predictive values for detecting methemoglobinemia. Spo(2) reading bias, precision, and root mean square error were evaluated during conditions of elevated methemoglobin. RESULTS: Observations spanned 66.2% to 99% Sao(2) and 0.6% to 14.4% methemoglobin over the 2 groups (170 blood draws). Masimo methemoglobin reading bias and precision over the full Sao(2) span was 7.7%+/- 13.0%. Best accuracy was found in the 95% to 100% Sao(2) range (1.9%+/- 2.5%), progressing to its worst in the 70% to 80% range (24.8%+/- 15.6%). Occurrence of methemoglobin readings in error >5% increased over each 5-point decrease in Sao(2) (P < 0.05). Masimo Spo(2) readings were biased -6.3% +/- 3.0% in the 95% to 100% Sao(2) range with 4% to 8.3% methemoglobin. Both the Radical-7 and Nonin 9700 pulse oximeters accurately detected decreases in Sao(2) <90% with 4% to 15% methemoglobin, despite displaying low Spo(2) readings when Sao(2) was >95%. CONCLUSIONS: The Radical-7's methemoglobin readings become progressively more inaccurate as Sao(2) decreases <95%, at times overestimating true values by 10% to 40%. Elevated methemoglobin causes the Spo(2) readings to underestimate Sao(2) similar to conventional 2-wavelength pulse oximeters at high saturation. Spo(2) readings from both types of instruments continue to trend downward during the development of hypoxemia (Sao(2) <90%) with methemoglobin levels up to 15%. (Anesth Analg 2010;111:143-8)
引用
收藏
页码:143 / 148
页数:6
相关论文
共 18 条
[1]   Acquired methemoglobinemia - A retrospective series of 138 cases at 2 teaching hospitals [J].
Ash-Bernal, R ;
Wise, R ;
Wright, SM .
MEDICINE, 2004, 83 (05) :265-273
[2]   EFFECTS OF METHEMOGLOBINEMIA ON PULSE OXIMETRY AND MIXED VENOUS OXIMETRY [J].
BARKER, SJ ;
TREMPER, KK ;
HYATT, J .
ANESTHESIOLOGY, 1989, 70 (01) :112-117
[3]   Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry - A human volunteer study [J].
Barker, Steven J. ;
Curry, Jeremy ;
Redford, Daniel ;
Morgan, Scott .
ANESTHESIOLOGY, 2006, 105 (05) :892-897
[4]  
BEBOUT DE, 2008, RESP THER, V3, P39
[5]   Confirmation of the pulse oximetry gap in carbon monoxide poisoning [J].
Bozeman, WP ;
Myers, RAM ;
Barish, RA .
ANNALS OF EMERGENCY MEDICINE, 1997, 30 (05) :608-611
[6]   The effect of methemoglobin on the equilibrium between oxygen and hemoglobin [J].
Darling, RC ;
Roughton, FJW .
AMERICAN JOURNAL OF PHYSIOLOGY, 1942, 137 (01) :0056-0068
[7]   Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and Gender [J].
Feiner, John R. ;
Severinghaus, John W. ;
Bickler, Philip E. .
ANESTHESIA AND ANALGESIA, 2007, 105 :S18-S23
[8]   The Light-Tissue Interaction of Pulse Oximetry [J].
Mannheimer, Paul D. .
ANESTHESIA AND ANALGESIA, 2007, 105 :S10-S17
[9]  
*MAS CORP, 2008, RAD 7 SIGN EXTR PULS
[10]  
Nijland R, 1995, Acta Anaesthesiol Scand Suppl, V107, P49