A perforated diamond anvil cell for high-energy x-ray diffraction of liquids and amorphous solids at high pressure

被引:30
作者
Soignard, Emmanuel [1 ]
Benmore, Chris J. [2 ,3 ]
Yarger, Jeffery L. [2 ,4 ]
机构
[1] Arizona State Univ, LeRoy Eyring Ctr Solid State Sci, Tempe, AZ 85287 USA
[2] Arizona State Univ, Dept Phys, Tempe, AZ 85287 USA
[3] Argonne Natl Lab, Adv Photon Source, Xray Sci Div, Argonne, IL 60439 USA
[4] Arizona State Univ, Dept Chem & Biochem, Tempe, AZ 85287 USA
关键词
diamond; X-ray diffraction; X-ray scattering; SUPERCRITICAL CONDITIONS; MULTICHANNEL COLLIMATOR; TRANSITION; DETECTOR; ELEMENTS; GLASSES; XAFS;
D O I
10.1063/1.3356977
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Diamond anvil cells (DACs) are widely used for the study of materials at high pressure. The typical diamonds used are between 1 and 3 mm thick, while the sample contained within the opposing diamonds is often just a few microns in thickness. Hence, any absorbance or scattering from diamond can cause a significant background or interference when probing a sample in a DAC. By perforating the diamond to within 50-100 mu m of the sample, the amount of diamond and the resulting background or interference can be dramatically reduced. The DAC presented in this article is designed to study amorphous materials at high pressure using high-energy x-ray scattering (>60 keV) using laser-perforated diamonds. A small diameter perforation maintains structural integrity and has allowed us to reach pressures >50 GPa, while dramatically decreasing the intensity of the x-ray diffraction background (primarily Compton scattering) when compared to studies using solid diamonds. This cell design allows us for the first time measurement of x-ray scattering from light (low Z) amorphous materials. Here, we present data for two examples using the described DAC with one and two perforated diamond geometries for the high-pressure structural studies of SiO(2) glass and B(2)O(3) glass.
引用
收藏
页数:9
相关论文
共 37 条
  • [1] Modified hydrothermal diamond anvil cells for XAFS analyses of elements with low energy absorption edges in aqueous solutions at sub- and supercritical conditions
    Bassett, WA
    Anderson, AJ
    Mayanovic, RA
    Chou, IM
    [J]. ZEITSCHRIFT FUR KRISTALLOGRAPHIE, 2000, 215 (12): : 711 - 717
  • [2] Hydrothermal diamond anvil cell for XAFS studies of first-row transition elements in aqueous solution up to supercritical conditions
    Bassett, WA
    Anderson, AJ
    Mayanovic, RA
    Chou, IM
    [J]. CHEMICAL GEOLOGY, 2000, 167 (1-2) : 3 - 10
  • [3] Diamond anvil cell, 50th birthday
    Bassett, William A.
    [J]. HIGH PRESSURE RESEARCH, 2009, 29 (02) : CP5 - 186
  • [4] Structural and topological changes in silica glass at pressure
    Benmore, C. J.
    Soignard, E.
    Amin, S. A.
    Guthrie, M.
    Shastri, S. D.
    Lee, P. L.
    Yarger, J. L.
    [J]. PHYSICAL REVIEW B, 2010, 81 (05)
  • [5] X-ray diffraction and adsorption at extreme pressures
    Brister, K
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (04) : 1629 - 1647
  • [6] CHAPMAN KW, J APPL CRYS IN PRESS
  • [7] Applications of an amorphous silicon-based area detector for high-resolution, high-sensitivity and fast time-resolved pair distribution function measurements
    Chupas, Peter J.
    Chapman, Karena W.
    Lee, Peter L.
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2007, 40 : 463 - 470
  • [8] Breakdown of intermediate-range order in liquid GeSe2 at high pressure
    Crichton, WA
    Mezouar, M
    Grande, T
    Stolen, S
    Grzechnik, A
    [J]. NATURE, 2001, 414 (6864) : 622 - 625
  • [9] Applications of perforated diamond anvils for very high-pressure research
    Dadashev, A
    Pasternak, MP
    Rozenberg, GK
    Taylor, RD
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (06) : 2633 - 2637
  • [10] Quantitative structure factor and density measurements of high-pressure fluids in diamond anvil cells by x-ray diffraction: Argon and water
    Eggert, JH
    Weck, G
    Loubeyre, P
    Mezouar, M
    [J]. PHYSICAL REVIEW B, 2002, 65 (17): : 1 - 12