MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children

被引:52
作者
Mukundan, Srinivasan, Jr. [1 ]
Wang, Page Inman [1 ]
Frush, Donald P. [1 ]
Yoshizumi, Terry [1 ]
Marcus, Jeffrey [1 ]
Kloeblen, Emily [1 ]
Moore, Meredith [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27705 USA
关键词
bismuth shielding; CT; dosimetry; eye; head and neck imaging; pediatric imaging; radiation dose; safety;
D O I
10.2214/AJR.06.1146
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of our study was to measure radiation dose to the orbit during pediatric cranial CT with and without bismuth shielding using a novel dosimetry system. Cranial CT was performed on a pediatric anthropomorphic phantom, with and without bismuth eye shields. A solid-state metal oxide semiconductor field effect transistor ( MOSFET) dosimeter was used to obtain real-time dose measurements. CONCLUSION. Bismuth shielding reduced radiation dose to the eye by up to 42%; shield artifact fell outside the diagnostic area of interest.
引用
收藏
页码:1648 / 1650
页数:3
相关论文
共 9 条
[1]   Is it time to retire the CTDI for CT quality assurance and dose optimization? [J].
Brenner, DJ .
MEDICAL PHYSICS, 2005, 32 (10) :3225-3226
[2]  
Colombo Paola, 2004, Radiol Med, V108, P560
[3]  
Hopper KD, 2001, AM J NEURORADIOL, V22, P1194
[4]   CT dose reduction and dose management tools: Overview of available options [J].
McCollough, Cynthia H. ;
Bruesewitz, Michael R. ;
Kofler, James M., Jr. .
RADIOGRAPHICS, 2006, 26 (02) :503-U14
[5]   Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users' needs? [J].
McLaughlin, DJ ;
Mooney, RB .
CLINICAL RADIOLOGY, 2004, 59 (05) :446-450
[6]  
MERRIAM GR, 1957, AMER J ROENTGENOL RA, V77, P759
[7]   Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: A Monte Carlo study [J].
Perisinakis, K ;
Raissaki, M ;
Tzedakis, A ;
Theocharopoulos, N ;
Damilakis, J ;
Gourtsoyiannis, N .
MEDICAL PHYSICS, 2005, 32 (04) :1024-1030
[8]  
Task Group on Accident Prevention and Safety in Radiation Therapy, 2000, Ann ICRP, V30, P7
[9]   GANTRY ANGULATION IN BRAIN CT - DOSAGE IMPLICATIONS, EFFECT ON POSTERIOR-FOSSA ARTIFACTS, AND CURRENT INTERNATIONAL PRACTICE [J].
YEOMAN, LJ ;
HOWARTH, L ;
BRITTEN, A ;
COTTERILL, A ;
ADAM, EJ .
RADIOLOGY, 1992, 184 (01) :113-116