Investigating the Quantitative Fidelity of Prospectively Undersampled Chemical Shift Imaging in Muscular Dystrophy with Compressed Sensing and Parallel Imaging Reconstruction

被引:34
作者
Hollingsworth, Kieren G. [1 ]
Higgins, David M. [2 ]
McCallum, Michelle [3 ]
Ward, Louise [1 ]
Coombs, Anna [1 ]
Straub, Volker [3 ]
机构
[1] Newcastle Univ, Newcastle Magnet Resonance Ctr, Inst Cellular Med, Newcastle Upon Tyne NE4 5PL, Tyne & Wear, England
[2] Philips Healthcare, Guildford, Surrey, Uruguay
[3] Newcastle Univ, Inst Med Genet, Newcastle Upon Tyne NE4 5PL, Tyne & Wear, England
基金
英国医学研究理事会;
关键词
MRI; BMD; compressed sensing; biomarker; muscular dystrophy; WATER-FAT SEPARATION; MRI; QUANTIFICATION; MUSCLE; ALGORITHM; FRACTION; DIXON;
D O I
10.1002/mrm.25072
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeFat fraction measurement in muscular dystrophy has an important role to play in future therapy trials. Undersampled data acquisition reconstructed by combined compressed sensing and parallel imaging (CS-PI) can potentially reduce trial cost and improve compliance. These benefits are only gained from prospectively undersampled acquisitions. MethodsEight patients with Becker muscular dystrophy were recruited and prospectively undersampled data at ratios of 3.65x, 4.94x, and 6.42x were acquired in addition to fully sampled data: equivalent coherent undersamplings were acquired for reconstruction with parallel imaging alone (PI). Fat fraction maps and maps of total signal were created using a combined compressed sensing/parallel imaging (CS-PI) reconstruction. ResultsThe CS-PI reconstructions are of sufficient quality to allow muscle delineation at 3.65x and 4.94x undersampling but some muscles were obscured at 6.42x. When plotted against the fat fractions derived from fully sampled data, non-significant bias and 95% limits of agreement of 1.58%, 2.17% and 2.41% were found for the three CS-PI reconstructions, while a 3.36x PI reconstruction yields 2.78%, 1.8 times worse than the equivalent CS-PI reconstruction. ConclusionProspective undersampling and CS-PI reconstruction of muscle fat fraction mapping can be used to accelerate muscle fat fraction measurement in muscular dystrophy. Magn Reson Med 72:1610-1619, 2014. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:1610 / 1619
页数:10
相关论文
共 34 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
Chebrolu VV, 2020, J MAGN RESON IMAGING, V32, P493
[3]   STOCHASTIC SAMPLING IN COMPUTER-GRAPHICS [J].
COOK, RL .
ACM TRANSACTIONS ON GRAPHICS, 1986, 5 (01) :51-72
[4]   Compressed Sensing for Chemical Shift-Based Water-Fat Separation [J].
Doneva, Mariya ;
Boernert, Peter ;
Eggers, Holger ;
Mertins, Alfred ;
Pauly, John ;
Lustig, Michael .
MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (06) :1749-1759
[5]   Dual-Echo Dixon Imaging with Flexible Choice of Echo Times [J].
Eggers, Holger ;
Brendel, Bernhard ;
Duijndam, Adri ;
Herigault, Gwenael .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (01) :96-107
[6]   An EM algorithm for wavelet-based image restoration [J].
Figueiredo, MAT ;
Nowak, RD .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 2003, 12 (08) :906-916
[7]   Quantitative MRI can detect subclinical disease progression in muscular dystrophy [J].
Fischmann, Arne ;
Hafner, Patricia ;
Fasler, Susanne ;
Gloor, Monika ;
Bieri, Oliver ;
Studler, Ueli ;
Fischer, Dirk .
JOURNAL OF NEUROLOGY, 2012, 259 (08) :1648-1654
[8]   Muscle fat-fraction and mapping in Duchenne muscular dystrophy: evaluation of disease distribution and correlation with clinical assessments [J].
Gaeta, Michele ;
Messina, Sonia ;
Mileto, Achille ;
Vita, Gian Luca ;
Ascenti, Giorgio ;
Vinci, Sergio ;
Bottari, Antonio ;
Vita, Giuseppe ;
Settineri, Nicola ;
Bruschetta, Daniele ;
Racchiusa, Sergio ;
Minutoli, Fabio .
SKELETAL RADIOLOGY, 2012, 41 (08) :955-961
[9]   In vivo characterization of the liver fat 1H MR spectrum [J].
Hamilton, Gavin ;
Yokoo, Takeshi ;
Bydder, Mark ;
Cruite, Irene ;
Schroeder, Michael E. ;
Sirlin, Claude B. ;
Middleton, Michael S. .
NMR IN BIOMEDICINE, 2011, 24 (07) :784-790
[10]   Addressing phase errors in fat-water imaging using a mixed magnitude/complex fitting method [J].
Hernando, D. ;
Hines, C. D. G. ;
Yu, H. ;
Reeder, S. B. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (03) :638-644