Resistance to Extrinsic Compression and Maintenance of Intraluminal Flow in Coil-Reinforced Stents (Silhouette® Scaffold Device): An In Vitro Study

被引:11
作者
Miyaoka, Ricardo [1 ]
Hendlin, Kari [1 ]
Monga, Manoj [1 ]
机构
[1] Univ Minnesota, Dept Urol Surg, Minneapolis, MN 55455 USA
关键词
IPSILATERAL URETERAL STENTS; OBSTRUCTION; EXPERIENCE; MANAGEMENT; CARCINOMA; FAILURE;
D O I
10.1089/end.2009.0513
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Stenting is the best approach to manage ureteral extrinsic compression and depends on intraluminal patency, which is mainly determined by cross-sectional stability. We evaluated the resistance to extrinsic compression of the Silhouette (R) Scaffold Device and its ability to simultaneously maintain intraluminal flow. Materials and Methods: Four 8 F Silhouette stents of different lengths were evaluated for radial compression and intraluminal flow by an MTS Micro Bionix Testing System using Testworks II software. Compression was exerted in 0.2 mm increment cycles until a maximum load cell of 5 N was reached. The Young's modulus, E, was calculated from each trial using engineering stress. Flow rates were compared with the 10.2 F Cook Amplatz. Analysis of variance was used to detect differences between Scaffold stents, and Student's t-test for differences between baseline flow rates. A level of significance of p < 0.05 was used. Results: Cross-sectional area and E did not differ throughout different length Scaffold stents (average 28,856 +/- 2592; 95% confidence interval 27,447, 30,265; p - 0.426). The inner lumen of the 10.2 F Cook Amplatz was larger by 0.125 mm. Baseline flow rate of the Cook Amplatz was significantly higher than that of the Silhouette Scaffold (1.089 vs. 0.660 mL/second, p = 0.001), but the drop in the flow rate was four times faster for the Amplatz. Conclusion: The resistance of 8 F Silhouette Scaffold Device is extremely higher than that of previously tested stents, and it is uniform along the shaft and does not vary with length. The Cook Amplatz 10.2 F stent has higher flow at baseline and 2 mm compression, but its flow rate drops four times faster than the Silhouette Scaffold.
引用
收藏
页码:595 / 598
页数:4
相关论文
共 16 条
[1]   Metal mesh stents for ureteral obstruction caused by hormone-resistant carcinoma of prostate [J].
Ahmed, M ;
Bishop, MC ;
Bates, CP ;
Manhire, AR .
JOURNAL OF ENDOUROLOGY, 1999, 13 (03) :221-224
[2]  
AVARANTINOS E, 2007, J ENDOUROL, V21, P1297
[3]   Mechanism of ureteral stent flow:: A comparative in vivo study [J].
Brewer, AV ;
Elbahnasy, AM ;
Bercowsky, E ;
Maxwell, KL ;
Shalhav, AL ;
Kahn, SA ;
McDougall, EM ;
Clayman, RV .
JOURNAL OF ENDOUROLOGY, 1999, 13 (04) :269-271
[4]   Analysis of Ureteral Stent Compression Force and its Role in Malignant Obstruction [J].
Christman, Matthew S. ;
L'Esperance, James O. ;
Choe, Chong H. ;
Stroup, Sean P. ;
Auge, Brian K. .
JOURNAL OF UROLOGY, 2009, 181 (01) :392-396
[5]   Experimental study of wallstent self-expandable metal stent in ureteral implantation [J].
Desgrandchamps, F ;
Tuchschmid, Y ;
CochandPriollet, B ;
Therin, M ;
Teillac, P ;
LeDuc, A .
JOURNAL OF ENDOUROLOGY, 1995, 9 (06) :477-481
[6]   HIGH FAILURE RATE OF INDWELLING URETERAL STENTS IN PATIENTS WITH EXTRINSIC OBSTRUCTION - EXPERIENCE AT 2 INSTITUTIONS [J].
DOCIMO, SG ;
DEWOLF, WC .
JOURNAL OF UROLOGY, 1989, 142 (02) :277-279
[7]   Management of ureteral obstruction secondary to pelvic malignancies [J].
Feng, MI ;
Bellman, GC ;
Shapiro, CE .
JOURNAL OF ENDOUROLOGY, 1999, 13 (07) :521-524
[8]   EXPERIENCE WITH INDWELLING URETERAL STENT CATHETERS [J].
GIBBONS, RP ;
CORREA, RJ ;
CUMMINGS, KB ;
MASON, JT .
JOURNAL OF UROLOGY, 1976, 115 (01) :22-26
[9]   In vitro evaluation of ureteral stent compression [J].
Hendlin, K ;
Vedula, K ;
Horn, C ;
Monga, M .
UROLOGY, 2006, 67 (04) :679-682
[10]   THE DOUBLE-J URETERAL STENT - INVIVO AND INVITRO FLOW STUDIES [J].
HUBNER, WA ;
PLAS, EG ;
STOLLER, ML .
JOURNAL OF UROLOGY, 1992, 148 (02) :278-280