Studies on a microfluidic chip based on continuous flow PCR amplification system

被引:0
作者
Liu, JH [1 ]
Yin, XF [1 ]
Xu, GM [1 ]
Fang, ZL [1 ]
Chen, HZ [1 ]
机构
[1] Zhejiang Univ, Dept Chem, Inst Microanal Syst, Hangzhou 310028, Peoples R China
来源
CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE | 2003年 / 24卷 / 02期
关键词
PCR amplification; microfluidic chip; DNA;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A PCR amplification system comprising a syringe pump sample processor, a microfluidic PCR chip and a 3-temperature-zone heating block was developed in this work. The system is capable of performing continuous PCR amplification with a low carryover between neighboring samples. The substrate thickness and thermal insulation of the temperature control unit were optimized. A 24-cycle PCR amplification of a 500-base pair fragment from lambda-DNA was successfully performed with a total reaction time of 4. 9 min.
引用
收藏
页码:232 / 235
页数:4
相关论文
共 12 条
[1]   Integrated system for rapid PCR-based DNA analysis in microfluidic devices [J].
Khandurina, J ;
McKnight, TE ;
Jacobson, SC ;
Waters, LC ;
Foote, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2000, 72 (13) :2995-3000
[2]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048
[3]  
LEGALLY ET, 2000, SENSOR ACTUAT B-CHEM, V63, P138
[4]  
MA LR, 2000, BIOCHIP, P88
[5]   A miniature analytical instrument for nucleic acids based on micromachined silicon reaction chambers [J].
Northrup, MA ;
Benett, B ;
Hadley, D ;
Landre, P ;
Lehew, S ;
Richards, J ;
Stratton, P .
ANALYTICAL CHEMISTRY, 1998, 70 (05) :918-922
[6]   Infrared-mediated thermocycling for ultrafast polymerase chain reaction amplification of DNA [J].
Oda, RP ;
Strausbauch, MA ;
Huhmer, AFR ;
Borson, N ;
Jurrens, SR ;
Craighead, J ;
Wettstein, PJ ;
Eckloff, B ;
Kline, B ;
Landers, JP .
ANALYTICAL CHEMISTRY, 1998, 70 (20) :4361-4368
[7]   Miniaturized flow-through PCR with different template types in a silicon chip thermocycler [J].
Schneegass, I ;
Bräutigam, R ;
Köhler, JM .
LAB ON A CHIP, 2001, 1 (01) :42-49
[8]   Optimization of the performance of the polymerase chain reaction in silicon-based microstructures [J].
Taylor, TB ;
WinnDeen, ES ;
Picozza, E ;
Woudenberg, TM ;
Albin, M .
NUCLEIC ACIDS RESEARCH, 1997, 25 (15) :3164-3168
[9]   Multiple sample PCR amplification and electrophoretic analysis on a microchip [J].
Waters, LC ;
Jacobson, SC ;
Kroutchinina, N ;
Khandurina, J ;
Foote, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 1998, 70 (24) :5172-5176
[10]   Microchip device for cell lysis, multiplex PCR amplification, and electrophoretic sizing [J].
Waters, LC ;
Jacobson, SC ;
Kroutchinina, N ;
Khandurina, J ;
Foote, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 1998, 70 (01) :158-162