Blind deconvolution in model-based iterative reconstruction for CT using a normalized sparsity measure

被引:11
作者
Hehn, Lorenz [1 ,2 ,3 ,4 ]
Tilley, Steven [5 ]
Pfeiffer, Franz [1 ,2 ,3 ,4 ]
Stayman, J. Webster [5 ]
机构
[1] Tech Univ Munich, Dept Phys, Chair Biomed Phys, D-85748 Garching, Germany
[2] Tech Univ Munich, Munich Sch BioEngn, D-85748 Garching, Germany
[3] Tech Univ Munich, Sch Med, Dept Diagnost & Intervent Radiol, D-81675 Munich, Germany
[4] Tech Univ Munich, Klinikum Rechts Isar, D-81675 Munich, Germany
[5] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
关键词
blind deconvolution; statistical image reconstruction; iterative reconstruction; computed tomography; IMAGE-RECONSTRUCTION; DETECTOR BLUR;
D O I
10.1088/1361-6560/ab489e
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Model-based iterative reconstruction techniques for CT that include a description of the noise statistics and a physical forward model of the image formation process have proven to increase image quality for many applications. Specifically, including models of the system blur into the physical forward model and thus implicitly performing a deconvolution of the projections during tomographic reconstruction, could demonstrate distinct improvements, especially in terms of resolution. However, the results strongly rely on an exact characterization of all components contributing to the system blur. Such characterizations can be laborious and even a slight mismatch can diminish image quality significantly. Therefore, we introduce a novel objective function, which enables us to jointly estimate system blur parameters during tomographic reconstruction. Conventional objective functions are biased in terms of blur and can yield lowest cost to blurred reconstructions with low noise levels. A key feature of our objective function is a new normalized sparsity measure for CT based on total-variation regularization, constructed to be less biased in terms of blur. We outline a solving strategy for jointly recovering low-dimensional blur parameters during tomographic reconstruction. We perform an extensive simulation study, evaluating the performance of the regularization and the dependency of the different parts of the objective function on the blur parameters. Scenarios with different regularization strengths and system blurs are investigated, demonstrating that we can recover the blur parameter used for the simulations. The proposed strategy is validated and the dependency of the objective function with the number of iterations is analyzed. Finally, our approach is experimentally validated on test-bench data of a human wrist phantom, where the estimated blur parameter coincides well with visual inspection. Our findings are not restricted to attenuation-based CT and may facilitate the recovery of more complex imaging model parameters.
引用
收藏
页数:13
相关论文
共 28 条
  • [1] Blind Deconvolution Using Convex Programming
    Ahmed, Ali
    Recht, Benjamin
    Romberg, Justin
    [J]. IEEE TRANSACTIONS ON INFORMATION THEORY, 2014, 60 (03) : 1711 - 1732
  • [2] [Anonymous], 1999, FORBILD PHANTOMS
  • [3] Behling R., 2015, Modern diagnostic x-ray sources: Technology, manufacturing, reliability, DOI DOI 10.1201/B18655
  • [4] Total variation blind deconvolution
    Chan, TF
    Wong, CK
    [J]. IEEE TRANSACTIONS ON IMAGE PROCESSING, 1998, 7 (03) : 370 - 375
  • [5] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [6] Incorporation of system resolution compensation (RC) in the ordered-subset transmission (OSTR) algorithm for transmission imaging in SPECT
    Feng, Bing
    Fessler, Jeffrey A.
    King, Michael A.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (07) : 941 - 949
  • [7] Fessler J. A., 2000, Handbook of Medical Imaging, Volume, V2, P1, DOI DOI 10.1117/3.831079
  • [8] Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation
    Gang, Grace J.
    Stayman, J. Webster
    Zbijewski, Wojciech
    Siewerdsen, Jeffrey H.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 260 - 278
  • [9] Simultaneous deblurring and iterative reconstruction of CBCT for image guided brain radiosurgery
    Hashemi, SayedMasoud
    Song, William Y.
    Sahgal, Arjun
    Lee, Young
    Huynh, Christopher
    Grouza, Vladimir
    Nordstrom, Hakan
    Eriksson, Markus
    Dorenlot, Antoine
    Regis, Jean Marie
    Mainprize, James G.
    Ruschin, Mark
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (07) : 2521 - 2541
  • [10] Huber P. J., 2011, International Encyclopedia of Statistical Science, P1248, DOI 10.1007/978-3-642-04898-2_594