Validation of ApneaLink™ Plus for the diagnosis of sleep apnea

被引:31
作者
Cho, Jae Hoon [1 ]
Kim, Hyun Jun [2 ]
机构
[1] Konkuk Univ, Coll Med, Dept Otorhinolaryngol Head & Neck Surg, Seoul, South Korea
[2] Ajou Univ, Sch Med, Dept Otolaryngol, Suwon 443721, Gyeonggi Do, South Korea
关键词
Obstructive sleep apnea; Portable sleep monitoring; Validation; PORTABLE RECORDING DEVICE; AIR-FLOW CHANNEL; PRACTICE PARAMETERS; MONITOR;
D O I
10.1007/s11325-017-1532-3
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose The aim of this study is to evaluate the validation of ApneaLink (TM) Plus (ALP) based on a large number of subjects in a prospective design. Method We recommended enrolling of 200 consecutive patients who had been referred because of habitual snoring or witnessed apnea during sleep. If consent was obtained, patients underwent standard polysomnography (PSG) and ALP evaluation simultaneously at the hospital (ALP(lab)), and repeated ALP at home (ALP(home)) within 1 month. The parameters of ALP were scored based on three different systems, Manual, Auto AASM, and Auto scoring systems. Result Among the 200 patients who were recommended for enrollment, 149 completed the study. Sensitivity was good for all three scoring systems, while specificity was poor for the Auto scoring system. The area under curve was highest for the manual scoring system and lowest for the auto scoring system, and increased as the apnea-hypopnea index (AHI) cutoff value increased from 5 to 30. Regarding agreement with PSG, the manual scoring system was most superior, followed by the Auto AASM, and Auto scoring systems. The degree of agreement between PSG and ALP was affected by sleep efficiency and the arousal index. Moderate agreement between PSG and ALP based on central apnea index was observed. Conclusion ALP is an accurate and reliable device for the diagnosis of OSA. Manual scoring is recommended; however, Auto AASM is also acceptable. ALP detects central sleep apnea to a certain degree. Both sleep efficiency and the arousal index affect the results of ALP.
引用
收藏
页码:799 / 807
页数:9
相关论文
共 18 条
  • [1] Validation of ApneaLink Ox™ for the diagnosis of obstructive sleep apnea
    Alberto Nigro, Carlos
    Dibur, Eduardo
    Malnis, Silvana
    Grandval, Sofia
    Nogueira, Facundo
    [J]. SLEEP AND BREATHING, 2013, 17 (01) : 259 - 266
  • [2] Practice Parameters for the Surgical Modifications of the Upper Airway for Obstructive Sleep Apnea in Adults
    Aurora, R. Nisha
    Casey, Kenneth R.
    Kristo, David
    Auerbach, Sanford
    Bista, Sabin R.
    Chowdhuri, Susmita
    Karippot, Anoop
    Lamm, Carin
    Ramar, Kannan
    Zak, Rochelle
    Morgenthaler, Timothy I.
    [J]. SLEEP, 2010, 33 (10) : 1408 - 1413
  • [3] Evaluation of the accuracy of manual and automatic scoring of a single airflow channel in patients with a high probability of obstructive sleep apnea
    BaHammam, Ahmed
    Sharif, Munir
    Gacuan, Divinagracia E.
    George, Smitha
    [J]. MEDICAL SCIENCE MONITOR, 2011, 17 (02): : MT13 - MT19
  • [4] Rules for Scoring Respiratory Events in Sleep: Update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events
    Berry, Richard B.
    Budhiraja, Rohit
    Gottlieb, Daniel J.
    Gozal, David
    Iber, Conrad
    Kapur, Vishesh K.
    Marcus, Carole L.
    Mehra, Reena
    Parthasarathy, Sairam
    Quan, Stuart F.
    Redline, Susan
    Strohl, Kingman P.
    Ward, Sally L. Davidson
    Tangredi, Michelle M.
    [J]. JOURNAL OF CLINICAL SLEEP MEDICINE, 2012, 8 (05): : 597 - 619
  • [5] Nightly variability in the indices of sleep-disordered breathing in men being evaluated for impotence with consecutive night polysomnograms
    Chediak, AD
    AcevedoCrespo, JC
    Seiden, DJ
    Kim, HH
    Kiel, MH
    [J]. SLEEP, 1996, 19 (07) : 589 - 592
  • [6] Evaluation of a portable recording device (ApneaLink™) for case selection of obstructive sleep apnea
    Chen, Hui
    Lowe, Alan A.
    Bai, Yuxing
    Hamilton, Peter
    Fleetham, John A.
    Almeida, Fernanda R.
    [J]. SLEEP AND BREATHING, 2009, 13 (03) : 213 - 219
  • [7] Practice parameters for the use of portable monitoring devices in the investigation of suspected obstructive sleep apnea in adults
    Chesson, AL
    Berry, RB
    Pack, A
    [J]. SLEEP, 2003, 26 (07) : 907 - 913
  • [8] Collop NA, 2007, J CLIN SLEEP MED, V3, P737
  • [9] Type III home sleep testing versus pulse oximetry: is the respiratory disturbance index better than the oxygen desaturation index to predict the apnoea-hypopnoea index measured during laboratory polysomnography?
    Dawson, Arthur
    Loving, Richard T.
    Gordon, Robert M.
    Abel, Susan L.
    Loewy, Derek
    Kripke, Daniel F.
    Kline, Lawrence E.
    [J]. BMJ OPEN, 2015, 5 (06):
  • [10] Eckert Danny J, 2008, Proc Am Thorac Soc, V5, P144, DOI 10.1513/pats.200707-114MG