Modulator design for x-ray scatter correction using primary modulation: Material selection

被引:34
作者
Gao, Hewei [1 ]
Zhu, Lei [2 ,3 ]
Fahrig, Rebecca [1 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Nucl & Radiol Engn Program, Atlanta, GA 30332 USA
[3] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Med Phys Program, Atlanta, GA 30332 USA
关键词
primary modulator; K-edge; beam hardening; scatter correction; SIMULATION;
D O I
10.1118/1.3457472
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: An optimal material selection for primary modulator is proposed in order to minimize beam hardening of the modulator in x-ray cone-beam computed tomography (CBCT). Recently, a measurement-based scatter correction method using primary modulation has been developed and experimentally verified. In the practical implementation, beam hardening of the modulator blocker is a limiting factor because it causes inconsistency in the primary signal and therefore degrades the accuracy of scatter correction. Methods: This inconsistency can be purposely assigned to the effective transmission factor of the modulator whose variation as a function of object filtration represents the magnitude of beam hardening of the modulator. In this work, the authors show that the variation reaches a minimum when the K-edge of the modulator material is near the mean energy of the system spectrum. Accordingly, an optimal material selection can be carried out in three steps. First, estimate and evaluate the polychromatic spectrum for a given x-ray system including both source and detector; second, calculate the mean energy of the spectrum and decide the candidate materials whose K-edge energies are near the mean energy; third, select the optimal material from the candidates after considering both the magnitude of beam hardening and the physical and chemical properties. Results: A tabletop x-ray CBCT system operated at 120 kVp is used to validate the material selection method in both simulations and experiments, from which the optimal material for this x-ray system is then chosen. With the transmission factor initially being 0.905 and 0.818, simulations show that erbium provides the least amount of variation as a function of object filtrations (maximum variations are 2.2% and 4.3%, respectively, only one-third of that for copper). With different combinations of aluminum and copper filtrations (simulating a range of object thicknesses), measured overall variations are 2.5%, 1.0%, and 8.6% for 25.4 mu m of copper, erbium, and tungsten, respectively. With and without 300 mu m of copper in the beam, the measured variations for 25.4 mu m of copper, erbium, and tungsten, 1 mm of aluminum, as well as 406 mu m of copper, are 1.8%, 0.2%, 5.5%, 1.9%, and 7.5%, respectively. Conclusions: The spatial variation in the effective transmission factor of the modulator blocker due to beam hardening caused by the modulator itself reaches a minimum when the K-edge of the modulator material is near the mean energy of the spectrum. An optimal modulator material selection using the K-edge discontinuity is therefore proposed. 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3457472]
引用
收藏
页码:4029 / 4037
页数:9
相关论文
共 16 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[3]   BEAM HARDENING IN X-RAY RECONSTRUCTIVE TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) :390-398
[4]  
Cheng Ming-Jong, 2008, Journal of Taiwan Fisheries Research, V16, P67
[5]  
Elmsley J., 2001, NATURES BUILDING BLO
[6]   X-ray imaging with amorphous selenium: Optimal spectra for digital mammography [J].
Fahrig, R ;
Rowlands, JA ;
Yaffe, MJ .
MEDICAL PHYSICS, 1996, 23 (04) :557-567
[7]   OPTIMIZATION OF SPECTRAL SHAPE IN DIGITAL MAMMOGRAPHY - DEPENDENCE ON ANODE MATERIAL, BREAST THICKNESS, AND LESION TYPE [J].
FAHRIG, R ;
YAFFE, MJ .
MEDICAL PHYSICS, 1994, 21 (09) :1473-1481
[8]   Beam hardening correction for middle-energy industrial computerized tomography [J].
Gao, Hewei ;
Zhang, Li ;
Chen, Zhiqiang ;
Xing, Yuxiang ;
Li, Shuanglei .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) :2796-2807
[9]   Scatter correction method for x-ray CT using primary modulation: Phantom studies [J].
Gao, Hewei ;
Fahrig, Rebecca ;
Bennett, N. Robert ;
Sun, Mingshan ;
Star-Lack, Josh ;
Zhu, Lei .
MEDICAL PHYSICS, 2010, 37 (02) :934-946
[10]   A method for simultaneous correction of spectrum hardening artifacts in CT images containing both bone and iodine [J].
Joseph, PM ;
Ruth, C .
MEDICAL PHYSICS, 1997, 24 (10) :1629-1634