Fisher Information-Based Evaluation of Image Quality for Time-of-Flight PET

被引:26
|
作者
Vunckx, Kathleen [1 ]
Zhou, Lin [1 ]
Matej, Samuel [2 ]
Defrise, Michel [3 ]
Nuyts, Johan [1 ]
机构
[1] Katholieke Univ Leuven, Dept Nucl Med, B-3000 Louvain, Belgium
[2] Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
[3] VU Brussels, Dept Nucl Med, B-1090 Brussels, Belgium
基金
美国国家卫生研究院;
关键词
Filtered backprojection (FBP); Fisher information; image quality; maximum likelihood expectation maximization (MLEM); time-of-flight information in positron emission tomography (TOF PET); EMISSION-TOMOGRAPHY; EM ALGORITHM; RECONSTRUCTION; LIKELIHOOD; RESOLUTION; VARIANCE; BENEFIT;
D O I
10.1109/TMI.2009.2029098
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The use of time-of-flight (TOF) information during positron emission tomography (PET) reconstruction has been found to improve the image quality. In this work we quantified this improvement using two existing methods: 1) a very simple analytical expression only valid for a central point in a large uniform disk source and 2) efficient analytical approximations for postfiltered maximum likelihood expectation maximization (MLEM) reconstruction with a fixed target resolution, predicting the image quality in a pixel or in a small region of interest based on the Fisher information matrix. Using this latter method the weighting function for filtered backprojection reconstruction of TOF PET data proposed by C. Watson can be derived. The image quality was investigated at different locations in various software phantoms. Simplified as well as realistic phantoms, measured both with TOF PET systems and with a conventional PET system, were simulated. Since the time resolution of the system is not always accurately known, the effect on the image quality of using an inaccurate kernel during reconstruction was also examined with the Fisher information-based method. First, we confirmed with this method that the variance improvement in the center of a large uniform disk source is proportional to the disk diameter and inversely proportional to the time resolution. Next, image quality improvement was observed in all pixels, but in eccentric and high-count regions the contrast-to-noise ratio (CNR) increased less than in central and low- or medium-count regions. Finally, the CNR was seen to decrease when the time resolution was inaccurately modeled (too narrow or too wide) during reconstruction. Although the maximum CNR is not very sensitive to the time resolution error, using an inaccurate TOF kernel tends to introduce artifacts in the reconstructed image.
引用
收藏
页码:311 / 321
页数:11
相关论文
共 50 条
  • [31] Dark Count Resilient Time Estimators for Time-of-Flight PET
    Lemaire, William
    Therrien, Audrey C.
    Pratte, Jean-Francois
    Fontaine, Rejean
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (01) : 24 - 29
  • [32] Focus on time-of-flight PET: the benefits of improved time resolution
    Maurizio Conti
    European Journal of Nuclear Medicine and Molecular Imaging, 2011, 38 : 1147 - 1157
  • [33] The assessment of time-of-flight on image quality and quantification with reduced administered activity and scan times in 18F-FDG PET
    Armstrong, Ian S.
    James, Jackie M.
    Williams, Heather A.
    Kelly, Matthew D.
    Matthews, Julian C.
    NUCLEAR MEDICINE COMMUNICATIONS, 2015, 36 (07) : 728 - 737
  • [35] Recent Advances and Future Advances in Time-of-Flight PET
    Moses, William W.
    NUCLEAR PHYSICS METHODS AND ACCELERATORS IN BIOLOGY AND MEDICINE, 2009, 1204 : 119 - 125
  • [36] LROC analysis of human detection performance in PET and time-of-flight PET
    Gifford, HC
    Wells, RG
    King, MA
    MEDICAL IMAGING 1999: IMAGE PERCEPTION AND PERFORMANCE, 1999, 3663 : 187 - 198
  • [37] Optimizing time-of-flight of PET/CT image quality via penalty (3 value in Bayesian penalized likelihood reconstruction algorithm
    Murat, H.
    Zulki, M. A. A.
    Said, M. A.
    Kechik, M. Awang
    Tahir, D.
    Karim, M. K. Abdul
    RADIOGRAPHY, 2025, 31 (01) : 343 - 349
  • [38] Performance Investigation of a Time-of-Flight PET/CT Scanner
    Jakoby, B. W.
    Bercier, Y.
    Conti, M.
    Casey, M.
    Gremillion, T.
    Hayden, C.
    Bendriem, B.
    Townsend, D. W.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3013 - +
  • [39] Imaging Performance of a LaBr3-Based Time-of-Flight PET Scanner
    Daube-Witherspoon, Margaret E.
    Surti, Suleman
    Perkins, Amy E.
    Kyba, C. M.
    Matej, Samuel
    Karp, J. S.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4736 - +
  • [40] Fully 3D list-mode time-of-flight PET image reconstruction on GPUs using CUDA
    Cui, Jing-yu
    Pratx, Guillem
    Prevrhal, Sven
    Levin, Craig S.
    MEDICAL PHYSICS, 2011, 38 (12) : 6775 - 6786