150-250 MeV electron beams in radiation therapy

被引:126
作者
DesRosiers, C
Moskvin, V
Bielajew, AF
Papiez, L
机构
[1] Indiana Univ, Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
[2] Univ Michigan, Ann Arbor, MI 48109 USA
关键词
D O I
10.1088/0031-9155/45/7/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
High-energy electron beams in the range 150-250 MeV are studied to evaluate the feasibility for radiotherapy. Monte Carlo simulation results from the PENELOPE code are presented and used to determine lateral spread and penetration of these beams. It is shown that the penumbra is comparable to photon beams at depths less than 10 cm and the practical range (R-p) of these beams is greater than 40 cm. The depth dose distribution of electron beams compares favourably with photon beams. Effects caused by nuclear reactions are evaluated, including increased dose due to neutron production and induced radioactivity resulting in an increased relative biological effectiveness (RBE) factor of <1.03.
引用
收藏
页码:1781 / 1805
页数:25
相关论文
共 33 条
[1]  
ANDREO P, 1992, 1 RIASTUOPTP U OS PR, P1
[2]  
[Anonymous], 1969, INDUCED RADIOACTIVIT
[3]   3-DIMENSIONAL CONFORMAL RADIATION-THERAPY MAY IMPROVE THE THERAPEUTIC RATIO OF HIGH-DOSE RADIATION-THERAPY FOR LUNG-CANCER [J].
ARMSTRONG, JG ;
BURMAN, C ;
LEIBEL, S ;
FONTENLA, D ;
KUTCHER, G ;
ZELEFSKY, M ;
FUKS, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 26 (04) :685-689
[4]   PENELOPE - AN ALGORITHM FOR MONTE-CARLO SIMULATION OF THE PENETRATION AND ENERGY-LOSS OF ELECTRONS AND POSITRONS IN MATTER [J].
BARO, J ;
SEMPAU, J ;
FERNANDEZVAREA, JM ;
SALVAT, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1995, 100 (01) :31-46
[5]  
Berger M.J., 1963, METHODS COMPUT PHYS, P135
[6]   DESIGN PRINCIPLES AND CLINICAL POSSIBILITIES WITH A NEW GENERATION OF RADIATION-THERAPY EQUIPMENT - A REVIEW [J].
BRAHME, A .
ACTA ONCOLOGICA, 1987, 26 (06) :403-412
[7]   ELECTRON AND PHOTON BEAMS FROM A 50 MEV RACETRACK MICROTRON [J].
BRAHME, A ;
KRAEPELIEN, T ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1980, 19 (04) :305-319
[8]  
CEMBER H, 1989, INTRO HLTH PHYSICS, P150
[9]  
DAS I, 1999, MED PHYS, V26, P1122
[10]   Monte Carlo simulation of the inelastic scattering of electrons and positrons using optical-data models [J].
Fernandez-Varea, JM .
RADIATION PHYSICS AND CHEMISTRY, 1998, 53 (03) :235-245